Next Article in Journal
An Orthodontic Treatment Case of a TMD Patient with Maxillary Posterior Intrusion Using TSADs
Previous Article in Journal
Comparison of Cartilage Conduction Hearing Aid, Bone Anchored Hearing Aid, and ADHEAR: Case Series of 6 Patients with Conductive and Mixed Hearing Loss
 
 
Font Type:
Arial Georgia Verdana
Font Size:
Aa Aa Aa
Line Spacing:
Column Width:
Background:
Article

Research on Compressed Sensing Spectrometry Based on Electro-Optical Transmittance Coding

1
School of Data Science and Technology, North University of China, Taiyuan 030051, China
2
Engineering and Technology Research Center of Shanxi Province for Opto-Electric Information and Instrument, Taiyuan 030051, China
3
School of Electrical and Control Engineering, North University of China, Taiyuan 030051, China
*
Authors to whom correspondence should be addressed.
Appl. Sci. 2022, 12(23), 12097; https://doi.org/10.3390/app122312097 (registering DOI)
Submission received: 4 November 2022 / Revised: 19 November 2022 / Accepted: 24 November 2022 / Published: 26 November 2022

Abstract

:
To develop fast and integrated spectrometry, we present a new method of compressed sensing spectrometry based on electro-optical transmittance coding. A LiNbO3 electro-optic modulator was applied to a directly compressed measurement code in the spectral dimension. Subsequently, a sequential forward floating selection algorithm was employed to select the coding measurement matrix, and sparsity adaptive matching pursuit was used as a solver algorithm. The principle was analyzed, and the measurement system was built for verification experiments. The experimental results reveal that the spectral reconstruction relative error is in the order of 10−2, and the full width at half maximum of the spectral measurement is as high as 1.2 nm. The spectral resolution can reach approximately 0.4 nm in the visible light band range (0.38–0.78 μm) with 1024 spectral channels. The compressed ratio of the compressed sensing spectrometry is up to 1:29. Furthermore, the spectral signal measurement time is only 0.25 ms. The study demonstrates a novel method, which exhibits high precision, speed, high compressed ratio and hyperspectral resolution.

1. Introduction

Spectroscopy reveals the essential characteristics of atoms, molecules, and molecular structures of substances [1]. It plays an important role in various fields, such as nondestructive testing, food safety testing, meteorological monitoring, agricultural general survey, geological resource exploration, and military target reconnaissance. With the development of spectrometry, the snapshot, miniaturized, and integrated methods and instruments have been gradually expanded [2,3,4,5]; however, they still face unprecedented challenges.
A conventional spectrometer is realized primarily through light splitting, filtering or interference. Computational spectrometry based on compressed sensing theory promotes the development of a new method of integrated and snapshot spectrum measurement. Brady et al. took the lead in using a two-dimensional coding template and dispersive optical element to study snapshot compressed sensing spectrometry [6,7]. This method uses both a binary-valued aperture to realize transmittance coding and a dispersive element to mix signals arriving at the focal plane of detector arrays. The spectrum is further inverted through the compressed sensing reconstruction algorithm. The target spectral data can be obtained after several measurements, and the spectrometry is fast and efficient. However, this method requires transformation from the spectral to the spatial dimension; consequently, it uses spatial coding to realize spectral coding, which undoubtedly increases the complexity of instrument architecture, cost and volume. Adrian et al. placed a liquid crystal variable retarder (LCVR) between two orthogonal polarizers, and the regulation of incident spectral transmittance was flexibly realized by controlling LCVR’s phase retardation [8]. Spectrum measurement with a high compressed ratio was demonstrated by coding the transmittance under different retardations. The method accomplished the coding in the spectral dimension, thereby producing a simpler and smaller system structure. Furthermore, compressed spectrometry is proposed using a Fabry Perot (FP) resonator, and the piezoelectric driver was used to adjust the distance between the two mirrors of the FP resonator to obtain different spectral transmittance regulations [9]. LCVR and FP resonators can realize direct coding in the spectral dimension, which significantly reduces the volume of the system. Nevertheless, the time responses of the methods are both in the order of ms. This is not conducive for applications that require ultrafast spectrum measurements, such as the fluorescence spectrum, explosion radiation spectrum etc. In addition, LCVR and FP are significantly affected by ambient temperature; additionally, the FP resonator is also sensitive to vibration interference. Further, the methods have poor environmental adaptability and limited working stability.
Recently, a new type of computational spectrometry combines recent progress in nanophotonics. Etalon arrays, etched thin films, and metasurfaces have been built on the focal plane of detector arrays for the development of micro spectrometers [10,11,12,13,14]. These methods are also coded in the spectral dimension, and their spectral resolutions are directly constrained by the number of coding arrays. The spectral resolution has exceeded 1 nm, but the fabrication of a large scale metasurface composed of hundreds different coding arrays is not only time-consuming but also costly.
In summary, fast, snapshot, and integrated spectrometry based on compressed sensing with transmittance coding in the spectral dimension must be studied. In this study, compressed sensing spectrometry based on electro-optic transmittance coding is proposed as an alternative to LCVR, FP and film coding templates. The sinusoidal driven transmittance coding and uncorrelated discrete compression measurement mechanism are explored, and a precision reconstruction algorithm is established. Additionally, a fast hyperspectral compressed measurement method is developed.

2. Principle

Coding compression measurements are necessary processes in compressed sensing spectrometry. Compared with transmittance control devices, such as LCVR, FP or films, this study proposes the transverse electro-optic modulation of LiNbO3 to realize transmittance control coding. The basic principle is shown in Figure 1. A X-cut LiNbO3 electro-optic crystal is the key component of the coding modulator. Herein, incident light passes along the optical axis (z-axis), and the driving electric field is applied along the x-axis, which is perpendicular to the light direction. The electro-optic device is placed between two polarizers with orthogonal transmittance axes. The intensity of incident light is regulated, which presents transmittance regulation, and realizes intensity coding in the spectral dimension.
In Figure 1, two transverse refractive index axes of the LiNbO3 electro-optic crystal are formed at 45° around the X axis under the action of an electric field, and the value of the crystal’s refractive index changes according to the applied electric intensity. When the incident light passes through the electro-optical coding modulator between the polarizers, the optical transmittance can be described as [15]
T ( λ , V ) = I ( λ , V ) / I 0 ( λ ) sin 2 ( δ ( λ , V ) / 2 ) ,
where, λ is the wavelength of the incident light; V is the voltage of the external electric field applied to the electro-optical transmittance coding modulator; I 0 ( λ ) and I ( λ , V ) are the light intensity before and after passing through the modulator, respectively; δ ( λ , V ) is the phase retardation of two polarization components in the LiNbO3 electro-optic crystal, and it changes with the change of the refractive index, which can be expressed as [16]
δ ( λ , V ) = 2 π λ ( n x n y ) L z = 2 π λ n 0 3 γ 22 L z L x V .
where, L x and L z are the length of the crystal in the x-axis and z-axis, respectively. The electric field applied in the x-direction is directly proportional to the voltage V. According to Equations (1) and (2), when the size of the LiNbO3 crystal is optimally designed, the spectral transmittance can be determined by the external electric field voltage V only. Therefore, the transmittance can be regulated by the modulator with different external electric field voltages. Multiple and different transmittance peaks can appear in the entire spectral range, and different external electric field voltages can realize the separation of transmittance peak wavelengths and the least overlap. This condition is conducive to the realization of compressed sensing spectrum measurements. As shown in Figure 1, the detector detected the light intensity each time, which is regulated by the electro-optic modulator. Each circumstance constitutes a special coding measurement in the process, and it can be described as
Y j = T ( λ , V j ) X ( λ ) d λ ,
X ( λ ) is the spectral power distribution of the incident light. To perform compressed spectrum measurement analysis and inversion reconstruction, the light intensities detected by the detector are recorded in a series of discrete values. The measurement process is described in a matrix form as follows.
y = Φ x ,
The spectral signal is represented by x, and the number of spectral channels is set to N such that x meets x R N × 1 . Multiple measured spectral signals are represented by y, and the number of measurements is set to M such that y meets y R M × 1 . For compressed sensing measurements, M is much less than N. The point is that the coding measurement matrix Φ ( Φ R M × N ) must be optimally designed to ensure that it is irrelevant as far as possible. By combining Equations (1) and (2), the design and implementation of Φ can be achieved by coding the spectral transmittance via the external electric field voltage. Φ composed of different spectral transmittance coding can be further described as
Φ = [ T ( λ 1 , V 1 ) T ( λ N , V 1 ) T ( λ i , V j ) T ( λ 1 , V M ) T ( λ N , V M ) ] = [ sin 2 ( δ ( λ 1 , V 1 ) / 2 ) sin 2 ( δ ( λ N , V 1 ) / 2 ) sin 2 ( δ ( λ i , V j ) / 2 ) sin 2 ( δ ( λ 1 , V M ) / 2 ) sin 2 ( δ ( λ 1 , V M ) / 2 ) ] .
For the N spectrum channels, M external electric field voltages are used to obtain the compressed sensing measurement data, where M << N, as shown in Figure 2.
The measurement matrix in Equation (5) is particularly similar to the partial Fourier matrix in the typical compressed sensing measurement matrix. Each element of Φ is T ( λ i , V j ) = sin 2 ( δ ( λ i , V j ) / 2 ) = 1 / 2 1 / 2 cos ( β ( V j ) 2 π / λ i ) , where β ( V j ) is determined by the optical path difference induced by the LiNbO3 electro-optic crystal. This parameter can be randomly selected after the transmittance calibration experiment is carried out. From the calibration results, the most uncorrelated transmittance under M external electric field voltages is selected, and M highly uncorrelated compressed sensing measurements can be realized.
Most of the spectral signals are sparse. The sparsity for the N channels spectral signals can be expressed as x = Ψ α . α is the K-sparse vector of spectral signals (including K << N non-zero elements), and Ψ is the sparse matrix. Equation (4) can be rewritten as
y = Φ x = Φ Ψ α = A α ,
where A is the sensing matrix, this is, A = Φ Ψ . For continuous spectral signals, mathematical transformations such as FFT, DCT and the wavelet basis or learning dictionary can be applied to the sparsity of spectral signals [7]. Thus, the sparse matrix Ψ , combined with the coding measurement matrix Φ , can be used to determine the sensing matrix A. Therefore, with the compressed sensing measurement signals y and after solving the inverse problem of Equation (6) and the optimal spectral sparse signal, α ^ can be solved.
α ^ = arg min α { y A α 2 + r α 1 } .
where, r is a regularization parameter. The spectrum signal can be recovered from the sparse signal x ^ = Ψ α ^ .

3. Experiment

According to Figure 1, a compressed sensing spectrometry system is built, as shown in Figure 3. The size of the LiNbO3 electro-optic crystal (LX, Ly, LZ) is selected as 2 mm × 4 mm × 40 mm, the light passing surface is the xoy plane, and the size is 2 mm × 4 mm. The distance the light travelled is 40 mm. The LiNbO3 electro-optic crystal (provided by Shanghai Institute of Optics and Fine Mechanics, Chinese Academy of Sciences) was grown by the Czochralski method. Its absorption loss is less than 0.1%. The dimensional tolerance of the crystal is ±0.1 mm. To uniformly apply the driving voltage to the LiNbO3 electro-optic crystal plate, a 30 μm thick aluminum film was plated on the two main sections of the crystal plate—yoz plane. The aluminum film serves as the electrode layer and can evenly distribute the applied voltage on the electrode surface and form a uniform electric field in the LiNbO3 electro-optic crystal plate. The electric field is provided by a high voltage driving circuit made by ourselves, and the driving circuit is composed primarily of an FPGA and an LC circuit. The FPGA provides a 1 kHz frequency of sinusoidal AC driving control voltage, and the magnitude of the driving voltage can be more than 2000 V resonantly amplified by the LC circuit. The voltage will choose all the values in the range of 0–2000 v within 0.25 ms in 1/4 cycles.
The polarizers are uncoated calcite Gran Taylor polarizers, the light transmission range is 0.35~2.3 μm, and the extinction ratio is up to 105:1. The light transmission aperture is 5 mm × 5 mm, which not only meets the light transmission requirements, but also reduces the volume of the spectrum measurement optical system as much as possible. The silicon-based amplified photodetector PDA10 A2 produced by Thorlabs company, with a wavelength response range of 0.2~1.1 μm and a measurement bandwidth response of DC-150 MHz was selected for photoelectric detection. Both collimating lens f1 and converging lens f2 are K9 glass flat convex lenses, with a focal length f = 6 mm and diameter D = 6 mm. Optical input and output interfaces are equipped with an optical fiber FC/PC adapter. Throughout the entire spectrometry system, the light to be measured is coupled by optical fiber and collimated into parallel light via the collimating lens f1; subsequently, the incident light is collimated into the LiNbO3 electro-optical transmittance coding modulator composed of polarizer1, the LiNbO3 electro-optic modulator, and polarizer2. The encoded optical signal is concentrated and emitted through converging lens f2, and finally detected by the photodetector. The dimensions of the entire spectrometry system are only 30 mm × 30 mm × 90 mm.
To verify the feasibility of this scheme, the system was selected to be operated in the visible light range of 0.38–0.78 μm. After the optical transmittance coding was calibrated, the compressed sensing spectrum measurement experiment was performed.

3.1. Transmittance Coding Experiment

A halogen tungsten lamp was used as the light source, and the LiNbO3 electro-optic modulator was driven by a DC high voltage to calibrate the transmittance. In the calibration experiment, the voltage increases from 0 to 2000 V in step of 1 V. The partial spectral transmittance calibration results were selected and recorded, as shown in Figure 4.
The calibration results in Figure 4 reveal that the transmittance can be flexibly adjusted according to the driving voltage, and the number of peaks in the spectral transmittance increases when increasing the driving voltage. Multiple and different transmittance peaks were achieved. This provides sufficient elements to construct the coding measurement matrix Φ . The sequential forward floating selection (SFFS) algorithm was employed to select the most irrelevant transmittance from the values of spectral transmittance regulation to form the coding measurement matrix Φ and realize spectral transmittance coding [17]. The correlation coefficient ( μ ( Φ ) ) is regarded as the characteristic criterion (J) for selecting the elements of the matrix Φ , and the criterion (J) is J = 1 / μ ( Φ ) . The spectral transmittance with measurement times from 2 to 200 form the electro-optical transmittance coding measurement matrix Φ . Their characteristic criterion (J) was obtained and is shown in Figure 5.
Note that M external electric field voltages must be measured to realize an uncorrelated coding measurement matrix Φ to meet the non-correlation as much as possible. As shown in Figure 5, before 30 measurements, the correlation coefficient of the matrix Φ changes considerably with an increase in the number of measurements. When the number of measurements increases from 30 to 50, the change is gradual. In addition, the correlation of the measurement matrix changes negligibly, for more than 50 measurements.

3.2. Compressed Sensing Spectrum Measurement Experiment

The built compressed sensing spectrometry system was used to measure the blue, green, and red LED light sources. The FPGA provides a sinusoidal AC driving control voltage source, and simultaneously collects the driving voltage signal and measured spectrum signals. For the blue LED light sources, as shown in Figure 6, the non-correlation measurement with transmittance coding is adopted. The measurement data collected by the detector under different driving voltages are shown in Figure 6b. To solve the minimization problem in Equation (7), a sparsity adaptive matching pursuit (SAMP) solver was used [18]. Further, the sparse matrix Ψ was set as wavelet transform-db4, which exhibits good compressibility and reconstruction performance. The number of spectral bands are N = 1024. The spectrum signal recovered using M = 35 electro-optical transmittance coding compressed sensing measurements is shown in Figure 6c.
Figure 6 shows that the measured spectral signals are different under the transmittance coding realized by different driving voltages, and the driving voltage of non-correlated measurements obtained using the SFFS algorithm is not selected at equal intervals. The original spectrum of the blue LED was also measured via a commercial spectrometer, Ocean Optics HR4000, a high-resolution optical fiber spectrometer with a resolution of 0.2 nm. The experimental results of the measurement methods agree well. Additionally, the reconstruction spectrum was also obtained using different numbers of measurements from 20 to 100, and the spectral reconstruction accuracy can be calculated using the reconstruction relative error defined as
σ = x ^ x 2 x 2 × 100 % .
According to Figure 6d, the relative error does not exceed 3.69% when the measurements number reaches 35. The spectral measurements and reconstruction results of green LED and red LED light sources were recorded concurrently for M = 35 measurements, as shown in Figure 7.
The spectral measurement reconstruction results of LED light sources reveal that the compressed spectral measurement results obtained using our method are in good agreement with the measurement results of the commercial spectrometer HR4000. The reconstruction relative errors of blue, green, and red LED according to Equation (8) are recorded in Table 1.
The experimental results of the LED light sources verify that compressed spectrum measurement with the electro-optical transmittance coding method can achieve high-precision spectrum reconstruction and measurements. Moreover, a narrow linewidth semiconductor laser with a center wavelength of 660 nm and line width of 0.06 nm is used to test the resolution of the method. In the experiment, the number of measurements is also M = 35 and the number of reconstructed spectral data is N = 1024. The experimental results are shown in Figure 8.
As shown in Figure 8, the relative error of laser spectral reconstruction is 7.20%. The full width at half maximum (FWHM) of the laser spectrum measured by the compressed sensing spectrum measurement system is 1.2 nm, which is slightly less than 0.8 nm of HR4000 measurements. This similarity further demonstrates that the FWHM of the spectrum measurement system designed in this paper is small. The system has a high spectral resolution, and the theoretical spectral resolution can reach approximately 0.4 nm for a wide spectral range (0.38–0.78 μm) with 1024 spectral channels. Further, the spectrum of the lighting fluorescent lamp in the laboratory was measured. The experimental results are shown in the Figure 9.
The experimental result exhibits good agreement with that of HR4000, and the relative error of spectral reconstruction is 7.65%, which further proves that the proposed spectral measurement system scheme produces high precision measurements and spectrum reconstructions.

4. Discussion

The experimental results of LED light sources demonstrate that the reconstruction relative error does not decrease for M = 35 measurements. The compressed sensing spectrometry method was manipulated in the visible light band of 0.38~0.78 μm for spectral N = 1024 channels; thus, the compressed ratio is M/N = 1:29, which indicates that the spectrometry achieves a high compressed ratio measurement. The compressed spectral measurement results exhibit good agreement with the results of the commercial spectrometer HR4000. The reconstruction relative error of LED light source is less than 3.79%, whereas the reconstruction relative errors of laser and fluorescent lamp spectrums are slightly more than double. This difference is primarily because laser and fluorescent lamp spectrums have narrower spectral bands compared with the LED light spectrum. Nevertheless, the relative errors are maintained in the order of 10−2; thus, their reconstruction accuracy is high.
The FWHM of the spectral measurement is as high as 1.2 nm, which is slightly more than 0.8 nm of the HR4000 measurement. Furthermore, the spectral resolution can reach approximately 0.4 nm in the visible light band range (0.38~0.78 μm) with 1024 spectral channels. For comparison, performance of several reported methods were listed in Table 2. Hence, the proposed scheme can achieve more hyperspectral resolution measurements with less measurements. In addition, there are no moving elements in our optical system, and the LiNbO3 electro-optic crystal has good temperature stability, thus the spectrometry method is not affected by the temperature change and vibration interference.
The LiNbO3 electro-optical transmittance coding modulator has a wide allowable light transmission range of 0.35~5 μm, the detection spectral range can be prolonged to near infrared or shortwave infrared regions. Moreover, the measurement of the compressed spectral signal is completed within 1/4 cycles of the driving voltage of the electro-optic modulator, and the signal measurement time does not exceed 0.25 ms, which is better than the minimum integration time of 4 ms of HR4000. Thus, the proposed method realizes fast and wide spectral range hyperspectral resolution measurements. The size of the LiNbO3 electro-optic crystal (LX, Ly, LZ) was selected as 2 mm × 4 mm × 40 mm in this paper. To ensure sufficient beam aperture and the uniformity of the driving electric field, it is best to set the light passing area (xoy plane) to 2 mm × 4 mm. It is appropriate to set the length of crystal (LZ) to 20 mm to 60 mm. Significantly, the shorter the length is, the easier it is to align the optical path. While a higher driving voltage needs to be provided to obtain the irrelevant transmittance elements for the coding measurement matrix Φ according to Equations (2) and (5).

5. Conclusions

In summary, a new method of compressed sensing spectrometry based on electro-optical transmittance coding is presented. A LiNbO3 electro-optic modulator driven by sinusoidal AC control voltage was taken as the core to build the transmittance coder. The coding compression measurements were directly accomplished in the spectral dimension. The principle of compressed sensing spectrometry was established. The measurement system was built, and the transmittance coding and compressed sensing spectrum measurement experiments were carried out. The experimental results reveal that the spectral reconstruction relative error is in the order of 10−2, the FWHM value of the spectral measurement is as high as 1.2 nm, and the spectral resolution can reach approximately 0.4 nm. The method of compressed sensing spectrometry has a high compressed ratio (M/N = 1:29), and the spectral signal measurement time does not exceed 0.25 ms. The proposed method demonstrates high precise, high-speed, high compressed ratio, and integrated and wide spectral range hyperspectral measurements.

Author Contributions

Methodology, S.W.; software, S.W. and K.L.; formal analysis, K.L. and Z.W.; validation, S.W.; investigation, K.L. and Z.W.; resources, X.H. and Z.W.; data curation, S.W.; writing—original draft preparation, S.W. and K.L.; writing—review and editing, S.W.; supervision, X.H. and Z.W. All authors have read and agreed to the published version of the manuscript.

Funding

This research was funded by National Natural Science Foundation of China: 62205309; National Natural Science Foundation of China: 62205310.

Institutional Review Board Statement

Not applicable.

Informed Consent Statement

Not applicable.

Data Availability Statement

Not applicable.

Conflicts of Interest

The authors declare no conflict of interest.

References

  1. Zahir, S.A.D.M.; Omar, A.F.; Jamlos, M.F.; Azmi, M.A.M.; Muncan, J. A review of visible and near-infrared (Vis-NIR) spectroscopy application in plant stress detection. Sens. Actuat A-Phys. 2022, 338, 113468. [Google Scholar] [CrossRef]
  2. Kita, D.M.; Miranda, B.; Favela, D.; Bono, D.; Michon, J.; Lin, H.; Gu, T.; Hu, J. High-performance and scalable on-chip digital Fourier transform spectroscopy. Nat. Commun. 2018, 9, 4405. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  3. Wolffenbuttel, R.F. State-of-the-Art in Integrated Optical Microspectrometers. IEEE Trans. Instrum. Meas. 2004, 53, 197–202. [Google Scholar] [CrossRef] [Green Version]
  4. Bao, J.; Bawendi, M. A colloidal quantum dot spectrometer. Nature 2015, 523, 67–70. [Google Scholar] [CrossRef] [PubMed]
  5. Zhu, Y.; Lei, X.; Wang, K.X.; Yu, Z. Compact CMOS spectral sensor for the visible spectrum. Photonics Res. 2019, 7, 961. [Google Scholar] [CrossRef]
  6. Gehm, M.E.; John, R.; Brady, D.J. Single-shot compressive spectral imaging with a dual-disperser architecture. Opt. Express 2007, 15, 14013–14027. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  7. Wagadarikar, A.; John, R.; Willett, R. Single disperser design for coded aperture snapshot spectral imaging. Appl. Opt. 2008, 47, B44–B51. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  8. August, Y.; Stern, A. Compressive sensing spectrometry based on liquid crystal devices. Opt. Lett. 2013, 38, 4996–4999. [Google Scholar] [CrossRef] [Green Version]
  9. Oiknine, Y.; August, I.; Blumberg, D.G. Compressive sensing resonator spectroscopy. Opt. Lett. 2017, 42, 25–28. [Google Scholar] [CrossRef]
  10. Oiknine, Y.; August, I.; Blumberg, D.G. NIR hyperspectral compressive imager based on a modified Fabry–Perot resonator. J. Opt. 2018, 20, 044011. [Google Scholar] [CrossRef]
  11. Huang, E.; Ma, Q.; Liu, Z. Etalon Array Reconstructive Spectrometry. Sci. Rep. 2017, 7, 40693. [Google Scholar] [CrossRef] [Green Version]
  12. Cheolsun, K.; Woong, B.L.; Soo, K.; Yong, T.L.; Heung, N.L. Fabrication of 2D thin-film filter-array for compressive sensing spectroscopy. Opt. Laser Eng. 2019, 115, 53–58. [Google Scholar]
  13. Yang, J.; Cui, K.; Cai, X.; Xiong, J.; Zhu, H.; Rao, S. Ultraspectral imaging based on metasurfaces with freeform shaped meta-atoms. Laser Photonics Rev. 2022, 16, 2100663. [Google Scholar] [CrossRef]
  14. Gao, L.; Qu, Y.; Wang, L.; Yu, Z. Computational spectrometers enabled by nanophotonics and deep learning. Nanophotonics 2022, 11, 2507–2529. [Google Scholar] [CrossRef]
  15. Li, K.; Wang, S.; Han, X.; Wang, Z. Dispersion Measurement of Electro-Optic Coefficient γ22 of Lithium Niobate Based on Photoelastic Modulation. Appl. Sci. 2020, 10, 395. [Google Scholar] [CrossRef] [Green Version]
  16. Wang, S.; Wang, Z.; Han, X.; Li, K. Measurement method of electro-optic coefficients using photoelastic modulation. Appl. Opt. 2019, 58, 4271–4276. [Google Scholar] [CrossRef] [PubMed]
  17. Pudil, P.; Novovičová, J.; Kittler, J. Floating search methods in feature selection. Pattern Recognit. Lett. 1994, 15, 1119–1125. [Google Scholar] [CrossRef]
  18. Do, T.T.; Gan, L.; Nguyen, N. Sparsity adaptive matching pursuit algorithm for practical compressed sensing. In Proceedings of the 2008 42nd Asilomar Conference on Signals, Systems and Computers, Pacific Grove, CA, USA, 26–29 October 2008; pp. 581–587. [Google Scholar]
Figure 1. Schematic of spectrometry using a LiNbO3 electro-optical transmittance coding modulator.
Figure 1. Schematic of spectrometry using a LiNbO3 electro-optical transmittance coding modulator.
Applsci 12 12097 g001
Figure 2. Spectrum compressed sensing process.
Figure 2. Spectrum compressed sensing process.
Applsci 12 12097 g002
Figure 3. Compressed sensing spectrometry system: (a) Appearance; (b) Internal structure.
Figure 3. Compressed sensing spectrometry system: (a) Appearance; (b) Internal structure.
Applsci 12 12097 g003
Figure 4. Calibration of spectral transmittance coding under different driving voltages.
Figure 4. Calibration of spectral transmittance coding under different driving voltages.
Applsci 12 12097 g004
Figure 5. Relationship between the characteristic criterion and the number of transmittance coding.
Figure 5. Relationship between the characteristic criterion and the number of transmittance coding.
Applsci 12 12097 g005
Figure 6. Blue LED light source spectrum measurement: (a) Blue LED; (b) Compressed measurement signal; (c) Reconstructed spectrum signal; (d) Relative error with different number of measurements from 20 to 100.
Figure 6. Blue LED light source spectrum measurement: (a) Blue LED; (b) Compressed measurement signal; (c) Reconstructed spectrum signal; (d) Relative error with different number of measurements from 20 to 100.
Applsci 12 12097 g006
Figure 7. LED light source spectrum measurement: (a) Green LED light source; (b) Green LED measurement signal; (c) Red LED light source; (d) Red LED measurement signal; (e) Green LED spectrum reconstruction result; (f) Red LED spectrum reconstruction result.
Figure 7. LED light source spectrum measurement: (a) Green LED light source; (b) Green LED measurement signal; (c) Red LED light source; (d) Red LED measurement signal; (e) Green LED spectrum reconstruction result; (f) Red LED spectrum reconstruction result.
Applsci 12 12097 g007
Figure 8. Laser spectrum measurement experiment: (a) Laser light spot; (b) Results of compressed sensing spectrum measurements.
Figure 8. Laser spectrum measurement experiment: (a) Laser light spot; (b) Results of compressed sensing spectrum measurements.
Applsci 12 12097 g008
Figure 9. Fluorescent lamp spectrum measurement experiment: (a) Laboratory lighting fluorescent lamp source; (b) Compressed sensing spectrum measurement result.
Figure 9. Fluorescent lamp spectrum measurement experiment: (a) Laboratory lighting fluorescent lamp source; (b) Compressed sensing spectrum measurement result.
Applsci 12 12097 g009
Table 1. Reconstruction relative errors of LED light sources.
Table 1. Reconstruction relative errors of LED light sources.
Light SourcesReconstruction Relative Errors (%)
Blue LED3.69
Green LED3.58
Red LED3.79
Table 2. Comparison with reported spectrometry methods.
Table 2. Comparison with reported spectrometry methods.
MethodCompressed Ratio
M/N
Spectral ResolutionRef.
Aperture transmittance coding-10 nm[6]
LCVR coding1:100.4 nm[8]
FP coding1:131 nm[9,10]
Thin film coding1:141.4 nm[12]
Metasurface coding1:240.5 nm[13]
Electro-optic transmittance coding1:290.4 nmPresent work
Grating-0.2 nmOcean Optics HR4000
Publisher’s Note: MDPI stays neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Share and Cite

MDPI and ACS Style

Wang, S.; Li, K.; Han, X.; Wang, Z. Research on Compressed Sensing Spectrometry Based on Electro-Optical Transmittance Coding. Appl. Sci. 2022, 12, 12097. https://doi.org/10.3390/app122312097

AMA Style

Wang S, Li K, Han X, Wang Z. Research on Compressed Sensing Spectrometry Based on Electro-Optical Transmittance Coding. Applied Sciences. 2022; 12(23):12097. https://doi.org/10.3390/app122312097

Chicago/Turabian Style

Wang, Shuang, Kewu Li, Xie Han, and Zhibin Wang. 2022. "Research on Compressed Sensing Spectrometry Based on Electro-Optical Transmittance Coding" Applied Sciences 12, no. 23: 12097. https://doi.org/10.3390/app122312097

Note that from the first issue of 2016, this journal uses article numbers instead of page numbers. See further details here.

Article Metrics

Back to TopTop