Optical Response of Expired EBT3 Film for Absorbed Dose Measurement in X-ray and Electron Beam Range
Abstract
:1. Introduction
2. Materials and Methods
2.1. Radiography X-rays Experiment Procedure
2.2. High Energy LINAC Procedure
3. Results
3.1. Measurement Absorbed Dose
3.2. Radiation Exposure Effect on Expired EBT3 Films
3.3. Visible Spectroscopy Analysis of Expired EBT3 Gafchromic Film
3.3.1. Spectrum of Absorbency of Expired EBT3 Film
3.3.2. Linear Regression Analysis
3.3.3. Multiple Linear Regression (MLR) Analysis
3.4. Net ROD Measurement
4. Conclusions
Author Contributions
Funding
Institutional Review Board Statement
Informed Consent Statement
Data Availability Statement
Conflicts of Interest
References
- Sankar, A.; Ayyangar, K.M.; Nehru, R.M.; Kurup, P.G.; Murali, V.; Enke, C.A.; Velmurugan, J. Comparison of Kodak EDR2 and Gafchromic EBT film for intensity-modulated radiation therapy dose distribution verification. Med. Dosim. 2006, 31, 273–282. [Google Scholar] [CrossRef] [PubMed]
- Devic, S.; Seuntjens, J.; Sham, E.; Podgorsak, E.B.; Schmidtlein, C.R.; Kirov, A.S.; Soares, C.G. Precise radiochromic film dosimetry using a flat-bed document scanner. Med. Phys. 2005, 32, 2245–2253. [Google Scholar] [CrossRef] [PubMed]
- Fotina, I.; Winkler, P.; Künzler, T.; Reiterer, J.; Simmat, I.; Georg, D. Advanced kernel methods vs. Monte Carlo-based dose calculation for high energy photon beams. Radiother. Oncol. 2009, 93, 645–653. [Google Scholar] [CrossRef] [PubMed]
- Polednik, M.; Wolff, D.; Steil, V.; Delana, A.; Wenz, F.; Menegotti, L.; Trento, F.S. Optimization of the Gafchromic TM EBT protocol for IMRT QA. Z. Med. Phys. 2009, 19, 29–37. [Google Scholar]
- Fiandra, C.; Ricardi, U.; Ragona, R.; Anglesio, S.M.; Giglioli, F.R.; Calamia, E.; Lucio, F. Clinical use of EBT model Gafchromic™ film in radiotherapy. Med. Phys. 2006, 33, 4314–4319. [Google Scholar] [CrossRef] [PubMed]
- Darafsheh, A.; Hao, Y.; Maraghechi, B.; Cammin, J.; Reynoso, F.J.; Khan, R. Influence of 0.35 T magnetic field on the response of EBT3 and EBT-XD radiochromic films. Med. Phys. 2020, 47, 4543–4552. [Google Scholar] [CrossRef]
- Dreindl, R.; Georg, D.; Stock, M. Radiochromic film dosimetry: Considerations on precision and accuracy for EBT2 and EBT3 type films. Z. Med. Phys. 2014, 24, 153–163. [Google Scholar] [CrossRef]
- Farah, N.; Francis, Z.; Abboud, M. Analysis of the EBT3 Gafchromic film irradiated with 6 MV photons and 6 MeV electrons using reflective mode scanners. Phys. Medica. 2014, 30, 708–712. [Google Scholar] [CrossRef]
- Butson, M.J.; Yu, P.K.N.; Metcalfe, P. Effects of read-out light sources and ambient light on radiochromic film. Phys. Med. Biol. 1998, 43, 2407–2412. [Google Scholar] [CrossRef]
- Devic, S.; Tomic, N.; Pang, Z.; Seuntjens, J.; Podgorsak, E.B.; Soares, C.G. Absorption spectroscopy of EBT model film. Med. Phys. 2007, 34, 112–118. [Google Scholar] [CrossRef]
- Almond, P.R.; Biggs, P.J.; Coursey, B.M.; Hanson, W.F.; Huq, M.S.; Nath, R.; Rogers, D.W. AAPM’s TG-51 protocol for clinical reference dosimetry of high-energy photon and electron beams. Med. Phys. 1999, 26, 1847–1870. [Google Scholar] [CrossRef] [PubMed] [Green Version]
- Sukhikh, E.; Sukhikh, L.; Malikov, E. Polimer Gafchromic EBT3 films in clinical dosimetry. Adv. Mater. Res. 2015, 1084, 572–576. [Google Scholar] [CrossRef]
- León-Marroquín, E.Y.; Mulrow, D.J.; Khan, R.; Darafsheh, A. Spectral analysis of the EBT3 radiochromic films for clinical photon and electron beams. Med. Phys. 2019, 46, 973–982. [Google Scholar] [CrossRef] [PubMed]
- León-Marroquín, E.Y.; Mulrow, D.; Darafsheh, A.; Khan, R. Response characterization of EBT-XD radiochromic films in megavoltage photon and electron beams. Med. Phys. 2019, 46, 4246–4256. [Google Scholar] [CrossRef]
- Darafsheh, A.; Zhao, T.; Khan, R. Spectroscopic analysis of EBT-XD radiochromic films irradiated with proton and photon therapy beams. Phys. Med. Biol. 2020, 65, 205002. [Google Scholar] [CrossRef]
- Ahmad Shah, W.I.S.; Omar, A.F. Spectroscopy and Light Emitting Diodes Based System in Characterizing External Beam Therapy 3 Films for Solar Ultraviolet Measurement. Photonic Sens. 2020, 10, 34–44. [Google Scholar] [CrossRef] [Green Version]
- Osman, U.S.; Omar, A.F. Visible Spectroscopy in EBT3 Solar Ultraviolet Dosimeter; Universiti Sains Malaysia (USM): Penang, Malaysia, 2021. [Google Scholar]
- Williams, M.; Metcalfe, P. Radiochromic film dosimetry and its applications in radiotherapy. AIP Conf. Proc. 2011, 1345, 75–99. [Google Scholar]
- Callens, M.B.; Crijns, W.; Depuydt, T.; Haustermans, K.; Maes, F.; D’Agostino, E.; Wevers, M.; Pfeiffer, H.; Abeele, K.V.D. Modeling the dose dependence of the vis-absorption spectrum of EBT3 GafChromic™ films. Med. Phys. 2017, 44, 2532–2543. [Google Scholar] [CrossRef] [Green Version]
- Mirza, J.A.; Park, H.; Ye, S.-J. Use of radiochromic film as a high-spatial resolution dosimeter by Raman spectroscopy. Med. Phys. 2016, 43, 4520–4528. [Google Scholar] [CrossRef]
- Lewis, D.; Micke, A.; Yu, X.; Chan, M.F. An efficient protocol for radiochromic film dosimetry combining calibration and measurement in a single scan. Med. Phys. 2012, 39, 6339–6350. [Google Scholar] [CrossRef]
- Aydarous, A.; Al-Omary, E.A.; El Ghazaly, M. Characterization of Gafchromic EBT3 films for ultraviolet radiation dosimetry. Radiat. Eff. Defects Solids 2014, 169, 249–255. [Google Scholar] [CrossRef]
- Butson, E.T.; Cheung, T.; Yu, P.K.; Butson, M.J. Measuring solar UV radiation with EBT radiochromic film. Phys. Med. Biol. 2010, 55, N487–N493. [Google Scholar] [CrossRef] [PubMed]
- Chan, P.M.; Ng, C.Y.P.; Beni, M.S.; Law, S.S.; Yu, K.N. Auto-development issue in quality assurance of biological X-ray irradiator using Gafchromic EBT3 film. Results Phys. 2017, 7, 444–445. [Google Scholar] [CrossRef]
- Chiu-Tsao, S.; Massillon-Jl, G.; Domingo-Muñoz, I.; Chan, M. SU-E-T-96: Energy Dependence of the New GafChromic-EBT3 Film’s Dose Response-Curve. Med. Phys. 2012, 39, 3724. [Google Scholar] [CrossRef] [PubMed]
- Casanova Borca, V.; Pasquino, M.; Russo, G.; Grosso, P.; Cante, D.; Sciacero, P.; Girelli, G.; la Porta, M.R.; Tofani, S. Dosimetric characterization and use of GAFCHROMIC EBT3 film for IMRT dose verification. J. Appl. Clin. Med. Phys. 2013, 14, 4111. [Google Scholar]
Expired EBT3 Samples | Absorbed Dose (Gy) |
---|---|
1 | 0.62 |
2 | 1.12 |
3 | 2.39 |
4 | 2.90 |
5 | 3.72 |
Radiation Source | Prescribed Dose (MU) | Absorbed Dose (Gy) |
---|---|---|
X-rays: Energy: 6 MV dmax: 1.5 cm Field size: 10 cm × 10 cm SSD calibration | 100 | 1.00 |
200 | 2.00 | |
300 | 3.00 | |
400 | 4.00 | |
500 | 5.00 | |
Electron beam: Energy: 6 MeV dmax 1.3 cm Field size: 10 cm × 10 cm SSD calibration | 100 | 1.06 |
200 | 2.12 | |
300 | 3.20 | |
400 | 4.28 | |
500 | 5.34 |
Energy and Source of Radiation | 90-kVp X-Rays | 6-MV Photon Beam | 6-MeV Electron Beam | |||
---|---|---|---|---|---|---|
Wavelength (nm) | R2 (%) | RMSE (Gy) | R2 (%) | RMSE (Gy) | R2 (%) | RMSE (Gy) |
521.00, 532.13, 600.28 | 95.2 | 0.496 | 99.1 | 0.280 | 98.4 | 0.406 |
541.48, 561.11, 600.28 | 98.0 | 0.318 | 99.6 | 0.190 | 99.7 | 0.184 |
561.11, 600.28, 640.18 | 96.5 | 0.425 | 99.9 | 0.113 | 99.9 | 0.087 |
600.28, 624.50, 640.18 | 97.3 | 0.371 | 99.7 | 0.167 | 98.0 | 0.446 |
600.28, 654.10, 650.79 | 99.6 | 0.147 | 99.3 | 0.251 | 98.8 | 0.350 |
632.52, 641.84, 678.11 | 100.0 | 0.018 | 99.4 | 0.234 | 97.6 | 0.488 |
635.85, 646.82, 664.00 | 99.4 | 0.174 | 99.1 | 0.277 | 97.3 | 0.521 |
650.79, 654.10, 660.37 | 98.0 | 0.324 | 99.0 | 0.289 | 99.8 | 0.144 |
655.09, 675.81, 678.11 | 99.8 | 0.101 | 98.8 | 0.324 | 97.7 | 0.477 |
Wavelength (nm) | R2 (%) | RMSE (Gy) |
---|---|---|
541.48, 561.11, 600.28 | 89.3 | 0.585 |
650.79, 654.10, 660.37 | 85.2 | 0.688 |
600.28, 650.79, 654.10 | 85.4 | 0.682 |
Wavelength (nm) | R2 (%) | RMSE (Gy) |
---|---|---|
541.48, 561.11, 600.28 | 97.2 | 0.350 |
650.79, 654.10, 660.37 | 96.9 | 0.364 |
600.28, 650.79, 654.10 | 96.4 | 0.395 |
Source of Radiation | Dose (Gy) | Net ROD | RMSE (Gy) | LOD | LOQ |
---|---|---|---|---|---|
90-kVp X-rays (Radiography) | 0.62 | 0.038 | 0.038 | 1.044 | 3.163 |
1.12 | 0.047 | ||||
2.39 | 0.086 | ||||
2.90 | 0.118 | ||||
3.75 | 0.147 | ||||
6-MV photon beam (LINAC) | 1.00 | 0.086 | 0.113 | 1.057 | 3.202 |
2.00 | 0.156 | ||||
3.00 | 0.220 | ||||
4.00 | 0.285 | ||||
5.00 | 0.330 | ||||
6-MeV electron beam (LINAC) | 1.06 | 0.072 | 0.115 | 2.003 | 6.070 |
2.12 | 0.185 | ||||
3.20 | 0.225 | ||||
4.28 | 0.280 | ||||
5.34 | 0.331 |
Absorbed Dose (Gy) | Percentage Difference of Net ROD (%) | |
---|---|---|
90-kVp and 6-MV X-rays | 6-MV X-rays and 6-MeV Electron Beam | |
1.00 | 67.1 | 8.5 |
2.00 | 64.1 | 10.2 |
3.00 | 60.5 | 8.1 |
4.00 | 58.9 | 7.3 |
5.00 | 52.5 | 1.0 |
Publisher’s Note: MDPI stays neutral with regard to jurisdictional claims in published maps and institutional affiliations. |
© 2022 by the authors. Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (https://creativecommons.org/licenses/by/4.0/).
Share and Cite
Musa, M.; Abu Mhanna, H.Y.; Omar, A.F.; Radzi, Y.M.; Akhdar, H.F.; Ishak, N.H.; Al Ewaidat, H. Optical Response of Expired EBT3 Film for Absorbed Dose Measurement in X-ray and Electron Beam Range. Appl. Sci. 2022, 12, 3974. https://doi.org/10.3390/app12083974
Musa M, Abu Mhanna HY, Omar AF, Radzi YM, Akhdar HF, Ishak NH, Al Ewaidat H. Optical Response of Expired EBT3 Film for Absorbed Dose Measurement in X-ray and Electron Beam Range. Applied Sciences. 2022; 12(8):3974. https://doi.org/10.3390/app12083974
Chicago/Turabian StyleMusa, Maisarah, Hamad Yahia Abu Mhanna, Ahmad Fairuz Omar, Yasmin Md Radzi, Hanan Fawaz Akhdar, Nor Hafizah Ishak, and Haytham Al Ewaidat. 2022. "Optical Response of Expired EBT3 Film for Absorbed Dose Measurement in X-ray and Electron Beam Range" Applied Sciences 12, no. 8: 3974. https://doi.org/10.3390/app12083974
APA StyleMusa, M., Abu Mhanna, H. Y., Omar, A. F., Radzi, Y. M., Akhdar, H. F., Ishak, N. H., & Al Ewaidat, H. (2022). Optical Response of Expired EBT3 Film for Absorbed Dose Measurement in X-ray and Electron Beam Range. Applied Sciences, 12(8), 3974. https://doi.org/10.3390/app12083974