Next Article in Journal
Antioxidant and Biological Activities of Mahajanaka Mango Pulp Extract in Murine Models
Previous Article in Journal
AI-Powered Intelligent Seaport Mobility: Enhancing Container Drayage Efficiency through Computer Vision and Deep Learning
 
 
Font Type:
Arial Georgia Verdana
Font Size:
Aa Aa Aa
Line Spacing:
Column Width:
Background:
Article

Application of Performance Test Method in Korea for LED Optical Medical Device Samples

1
Department of Culture Convergence Technology Interdisciplinary Programs, Changwon National University, 20 Changwondaehak-ro, Uichang-gu, Changwon-si 51140, Republic of Korea
2
Korea Institute of Industrial Technology, 66, Hoegi-ro, Dongdaemun-gu, Seoul 03787, Republic of Korea
3
Department of Mechatronics Convergence, Changwon National University, 20 Changwondaehak-ro, Uichang-gu, Changwon-si 51140, Republic of Korea
*
Author to whom correspondence should be addressed.
Appl. Sci. 2023, 13(22), 12215; https://doi.org/10.3390/app132212215
Submission received: 13 August 2023 / Revised: 30 October 2023 / Accepted: 30 October 2023 / Published: 10 November 2023

Abstract

:
This study obtained key performance items related to phototherapy and the standards of light-emitting diode (LED) phototherapy devices through the published literature to secure LED phototherapy device safety in South Korea. Based on these items and IEC 62471-1 for the photobiological safety of lamps and lamp systems, a performance test method and wavelength parameters, which are the main performance indicators of the LED phototherapy device, were derived. We conducted a performance test derived from this study using samples with the same performance as the sample that is being developed. The samples comprised a laser and IR LED, red LED, and blue LED. All samples were within ±5%, of the standard threshold. The 455 nm blue LED and 625 nm red LED were close to the set value with a difference of 1 to 2 nm; their wavelengths’ accuracy was within ±0.32%, ±0%, and ±0.219%, depending on the samples’ wavelengths, which was the same or similar to the error range of ±0%. As for laser and IR LED, the difference from the set value was as large as 2 to 7 nm, and their error ranges were close to ±1%, as shown by ±0.615%, ±0.828%, or ±0.591%, which were larger than the values for red and blue LEDs.

1. Introduction

The development of light-based technologies has facilitated the use of various types of phototherapy devices in the medical field. Representative phototherapy devices are laser surgical instruments and medical laser irradiators. The former are used for the removal, ablation, and destruction of areas in need of treatment using the light of a strong laser; the latter irradiate the skin, induce skin regeneration, and relieve pain [1]. These medical devices are used under the guidance of healthcare providers in medical institutions and can be hazardous because they use strong light. According to the Korean Ministry of Food and Drug Safety, medical devices are divided into four classes according to the hazard level [2]. Devices such as lasers are currently represented by Class 3, which is the grade with high risk. However, various types of personal phototherapy devices have recently emerged and are used without any instructions from healthcare providers. Among them, representative personal phototherapy devices include phototherapy devices using LED. LED phototherapy devices use a low-level phototherapeutic method known as low-level laser therapy (LLLT). The wavelengths primarily used in LLLT fall within the range of 600 nm to 1000 nm, which is close to the visible and near-infrared spectrum. These wavelengths can penetrate the surface of the skin and are absorbed by the tissues, enhancing cellular activity. They are employed for pain or inflammation relief, scar healing, and skin improvement [3,4]. LED phototherapy devices with LLLT are used for skin improvement and cosmetic purposes. They vary in shape depending on the site at which to be used and are of various types. Certain devices use only LEDs, while others are used in combination with a laser, and their boundary with medical laser irradiators is vague. Based on the report on personal phototherapy devices by the Korea Consumer Agency, there were 172 adverse events of LED masks, which were home phototherapy devices, from 2018 to August 2020. They comprise 134 cases of skin and subcutaneous tissue injuries, 6 cases of burns, 1 of heat sensation and dyspnoea, and 1 of bruising [5]. As various phototherapy devices continue to be developed, ensuring adequate safety is paramount. This study contributed to the industrial development of LED phototherapy devices and new phototherapy devices. We derived a performance test method to secure safety related to the performance of phototherapy devices using LEDs, from a Korean perspective.

2. Materials and Methods

2.1. Determination of Performance Test Items Related to LED Phototherapy Devices

The goal of this study was to derive the performance test method to preserve the stability of LED optical medical devices and to obtain suitable test materials for these devices by investigating performance issues and conducting related performance tests. We proceeded with reviewing previous research and standards related to optical medical devices such as the International Organization for Standardization (ISO) and the International Electrotechnical Commission (IEC). For the literature review, a literature search medium was used, and the search was conducted with “LLLT,” “PBM(PBM, Photobiomodulation),” and “LED” as search keywords. Based on the reviewed literature, performance test items and methods were derived by examining key performance issues related to optical medical device therapy and LED optical medical device standards.

2.2. Performance Test Method for the Phototherapy Device Sample

We examined the safety of the LED phototherapy devices by applying the performance test items derived from this study to the LED medical device developed by T company, shown in Figure 1. The company’s LED phototherapy device samples were developed for the treatment of hair loss and scalp improvement and were shaped like a helmet. The LED phototherapy device comprises 54 lasers and 126 tricolor LEDs. Its performance parameters are 10 mW for the output power of the laser, 3 mW for the intended output power, and 650 nm for wavelength. Each LED has an output power of 3 mW, with wavelengths of 625 nm for the red LED, 455 nm for the blue LED, and 845 nm for the infrared (IR). The key performance test items and methods of optical medical devices are obtained and used in the model to protect the validity of the test.

3. Results

3.1. LED Optical Medical Device Literature Investigation

As a result of the literature review, three volumes were reviewed for LLLT, two volumes for LED, two volumes for PBM optical medical device-related literature, and five volumes for related standards. In addition, two books related to Korean medical devices and phototherapy were mentioned. By reviewing the research literature, the main performance parameters according to the treatment using LED light medical devices, standards related to LED light medical devices, and performance test parameters and methods were obtained.

3.1.1. Key Performance Points for Treatment Using LED Optical Medical Devices

Phototherapy is a medical technique that has been used for thousands of years to treat various diseases. An example is sun therapy by Hippocrates, the father of Western medicine [6]. In modern times, various treatments using light have evolved. A common treatment that uses LEDs is photobiomodulation (PBM). This was previously called low-level laser therapy (LLLT) and was later changed to low-level light therapy by Kendrick C. Smith due to the advent of the next-generation LED with higher photon intensity [7]. After the change, the name was again changed to PBM due to uncertainty about the exact meaning of low power [8]. PBM was proposed by Andre Meister (1903–1984), a medical expert who researched and published the biological effects of low-power lasers in 1967 [9]. According to the literature, PBM is a treatment that uses a low-power laser or LED in the 1 m to 50 mW range to induce tissue regeneration and relieve inflammation. Red or near-infrared light with a wavelength of 600 nm to 1000 nm is used, and unlike other laser treatments that destroy tissue, PBM is known for its photochemical effect similar to plant photosynthesis [3,10]. The mechanism of PBM is not yet clearly elucidated; however, when tissue is irradiated with low-output IR light, it is absorbed by a chromophore that penetrates the skin surface and absorbs light in a specific wavelength band. In the body, at that time [11,12], the absorbed light energy is converted into chemical energy that increases the synthesis of adenosine triphosphate (ATP), which is the energy source of the cells in the cell tissue, resulting in cell regeneration and rehabilitation as well as pain and inflammation relief [13,14]. According to the literature related to LED-LLLT, the main performance issues for LED light sources used in PPM include wavelength, power density, and luminance. In terms of wavelength, the first law of light biology, the Grothes–Draper law, states that when the light of the appropriate wavelength is applied to the chromophore, the chromophore has no absorption or reaction. Power density refers to the output used per unit area, and the unit is W·cm−2. When the power density is insufficient, light absorption does not occur in the tissue, and when the power density is high, it is converted to excessive heat for the tissue, which could be harmful. Fluence (energy density) or “dose” refers to the amount of energy applied per unit area. The energy density is multiplied by the radiation time, and the unit is J·cm−2. If the power density is too low, the irradiation time must be extended to achieve the best power density, which is why the desired results cannot be obtained [7,12].

3.1.2. Items Related to Standards for LED Phototherapy Devices

To derive LED phototherapy device test parameters, the relevant standards were investigated by referring to the international standards, ISO and IEC, as well as Korean standards. Currently, there are two main categories of common standards for medical devices in South Korea. The classification consists of standards applied comprehensively and standards applied separately by medical device items. Additionally, it comprises medical devices that use electricity as a whole and those that do not use electricity [15]. Comprehensive standards are common standards for electromechanical safety, electromagnetic waves and safety, and biological safety. The common standards for electromechanical safety are those applicable to the basic safety and essential performance of medical electrical appliances and medical electrical systems [15]. In the common standards on electromechanical safety, standards related to phototherapy devices are IEC 60601-1, IEC 60601-2-22, IEC 60825-1, IEC 60825-8, and IEC 62471-2. IEC 60601-1 is a set of standards that ensure the safety of medical electrical equipment, cover basic safety and essential performance requirements of medical electrical equipment, and prevent electrical, mechanical, or functional errors that present unacceptable hazards to patients and operators [16]. IEC 60601-2-22 is an individual standard of IEC 60601-1 applied to basic safety and essential performance of surgical, therapeutic, diagnostic, cosmetic, and animal laser instruments for use in humans or animals. IEC 60601-2-22 is a standard applicable to phototherapy devices corresponding to classes 3B to 4 and applied to the safety of laser output power. For example, on the 201.12.1.101 laser output power display, the actual laser output power measured in the operating area should not exceed ±20% of the set value [17]. IEC 60825-1 is a standard applied to the safety of laser products that emit light with a wavelength range of 180 nm to 1 mm. Per this standard, classes are divided into 1, 1M, 1C, 2, 2M, 3R, 3B, and 4, as described in Table 1 [18].
These ratings are classified as per the accessible emission limit (AEL) according to the maximum permissible exposure (MPE) in Annex A of IEC 60825-1, as shown in Table 2. The MPE provided in IEC 60825-1 presents additional information that can assist the manufacturer in assessing safety items for the intended use of the user. The MPE contained in the standard has been adopted from exposure limits published by the International Commission on Non-Ionized Radiation Protection, and MPE values have been used as the basis for the safe design of products and for providing information to users [18]. IEC 60825-8 provides guidance to employers, competent authorities, laser safety managers, and others on the safe use of laser and laser equipment classified as 3B and 4. An investigation identified potential side effects according to wavelengths in Appendix A of IEC 60825-1, which are shown in Table 2 [19].

3.1.3. Items for Standards Related to LED Phototherapy Devices

The main performance requirements and specifications related to performance testing of optical medical devices were grasped through the literature. It has been confirmed that wavelength, power density, and fluorescence are key performance parameters for LED optical medical devices. If a wavelength that does not match the chromophore of the target to be treated is used, suitable results cannot be obtained; in the case of fluorescence that does not match, the desired result cannot be obtained, and thermal damage can be expected. Therefore here, the key performance test parameters obtained were wavelength, power, and power density, as shown in Table 2. For assessing wavelength, we used the wavelength accuracy test method according to IEC 62471-1 Annex B; the accurate calculation formula was used to obtain the measured value and the actual value as a percentage, and the quality of the test ±□% within the error range was determined. According to this test standard, the safety testing criteria for Korean medical devices adhere to the range specified by the manufacturer. In this paper, it is indicated as ±□% [20]. Power density is the output per unit area, and an accuracy test for output and a test for illumination were obtained. Output accuracy test is in accordance with IEC 62471-1 Annex C. As a test method, the measured and the actual values were calculated as a percentage using a power meter [21]. The quality of the test should be ±□% with an error margin of ±□%. The test for illumination is IEC 62471-1. 5, wherein we followed the measurements of the lamp and the lamp device and measured it using an illuminance meter. The test standard is IEC 62471-1. 4. 4. The test method that follows the hazard exposure limits is IEC 62471-1.5. The test method was obtained according to the measurement of lamps and lighting fixtures and IEC 62471-1 4. Exposure limit standards were followed. Energy density, one of the key performance factors, is the product of energy density and irradiation time, and the test item of energy density was obtained. However, irradiation time was excluded because it was difficult to obtain. The test parameters can vary depending on the LED optical medical device.

3.2. LED Optical Medical Device Literature Investigation

Among the test items obtained from Section 3.1.3, the wavelength accuracy test was applied to LEDs built into the T company model, and the same test was performed on the laser built into the model. To measure the wavelength, as shown in Figure 2 below, we wanted to measure the main performance factors of the two models of the T company, the laser wavelength, and the LED wavelength, using the spectrometer of the T company. The two models have the same performance and we wanted to compare them by measuring the wavelengths in each mode.
As for the measurement method, the light source to be measured was aligned with the spectrometer’s wavelength meter sensor, as shown in Figure 3. The wavelength measured was then checked by the monitoring function in the software on the computer connected to the spectrometer.

3.3. LED Optical Medical Device Literature Investigation

The two samples from the T company were numbered 1 and 2, and the laser and LED were measured for each mode of the samples. The set values for each mode of samples were the same for the laser and IR LED, and the set values for the laser and the IR LED were 650 nm and 845 nm, respectively. The set values for the red LED in H (Hair) mode and the blue LED in S (Scalp) mode were 625 nm and 455 nm, respectively. The measured values for the two samples are shown in Figure 4.
The laser wavelengths per mode obtained from the T company’s samples measured with the spectrometer are shown in Figure 5. The graph shows a maximum peak value of 654 nm. When the measured wavelength band was visually checked, the graphs overlapped because the measured laser wavelength bands were similar to the set value of 650 nm. However, as it was difficult to accurately differentiate the superimposed graph with the naked eye, the graph was adjusted to distinguish the laser wavelength band measured, as shown in Figure 6. Using the adjusted graph, the measured wavelength for each mode of the two samples and the accuracy of the laser wavelengths were checked.
Laser measurements of T company’s samples revealed 654 nm in H mode, 654 nm in S mode (Sample 1), 652 nm in H mode, and 653 nm in S mode (Sample 2). They differed by 2 to 4 nm from the set value of 650 nm in the laser. As each of the measured lasers was the same for each of the samples, Sample 1 indicated the same value. However, Sample 2, though it was the same object, demonstrated a difference of 1 nm depending on the mode. When each measured value was applied to the wavelength accuracy equation, Sample 1 showed an error range of ±0.615% because the values per mode were the same. As for Sample 2, the error range was ±0.31% in H mode and ±0.461% in S mode. The laser wavelengths of the two samples were within the error range of ±5%, which was the test standard set by the manufacturer. For both samples, LED measurements with the spectrometer revealed the maximum peak values of 627 nm, 454 nm, and 838 nm, which were visible to the eye. When the graph was checked, both samples were found to be similarly matched and overlapped with the laser graph measured earlier. However, a difference of 1 to 3 nm was observed compared with the LED set values of 625 nm for red, 455 nm for blue, and 845 nm for IR. The graphs were adjusted, as shown in Figure 7, Figure 8 and Figure 9 for precise determination. The respective graphs confirmed LED wavelengths by each mode and the accuracy of wavelengths for the two samples.
The IR LED measured value was 838 nm in both H and S modes for Sample 1. In Sample 2, the values for H and S modes were 840 nm and 839 nm, respectively. For each wavelength, there were differences of 7 nm in Sample 1 and 5 to 6 nm in Sample 2 from the set value of 845 nm. The wavelength accuracy of Sample 1 was within ±0.828% because all measured values were the same. The wavelength accuracy of Sample 2 was ±0.591% in H mode and ±0.71% in S mode. Both samples were within the test standard of ±5%. In the case of IR LED and laser, the same object was applied in each mode. The measured values according to the mode were distinctly different from the set value.
The red LED in T company’s samples was used only in H mode. The measured values in Samples 1 and 2 were 627 nm with a difference of 2 nm from the set value of 625 nm. The accuracy of the red wavelength was ±0.32% for both samples, thus indicating that both samples had an error range within ±5%. Unlike the laser and IR LED measured earlier, the red LEDs had the same values for Samples 1 and 2, and the difference from the set value was small.
The blue LED of T company’s samples was used only in S mode. The measured values were 455 nm in Sample 1 and 454 nm in Sample 2. In Sample 1, the measured value was the same as the set value of 455 nm, whereas Sample 2 had a difference of 1 nm. Regarding the accuracy of the wavelength, Sample 1 had an error range of ±0% because the measured value was equal to the set value, whereas Sample 2 had ±0.219% in the error range. Thus, the values were within the test standard of ±5% in the error range. In this study, we tested the wavelength performance of both the laser and the LED used in the samples of the T company, from which we found that all measured values were within the reference values. The measured value of the blue LED was closer to the set value than the data measured earlier, which was found stable, as shown in Figure 9. The values of the wavelengths measured in this study and their accuracy are shown in Table 3. In conclusion, all measured values were within ±5%, the test standard. However, the laser and IR LED, which are used regardless of mode, had larger differences between the measured values and the set values among the measurements.

4. Conclusions

This study investigated parameters related to the performance of LED phototherapy devices to ensure their safety. Although LLLT using LED phototherapy devices, which have been recently examined, has no reported principle, it was possible to identify wavelength as the main performance parameter, which is the same for other phototherapy devices. The standard of phototherapy devices equipped with both LED and lasers, such as T company’s phototherapy device, needs to be explored in combination with LED and laser standards. The investigation of standards was able to identify the classes of medical laser devices in IEC 60825-1 and their hazard depending on the wavelength in IEC 60825-8, the appendix of IEC 60825-1. However, certain laser standards were found to be applied to 3B or higher classes of a laser. As the samples of T company’s device used in this study were used in combination with LED and low-level laser, it was difficult to investigate the standards because of the differentiation between the LED and laser. For LED phototherapy devices, IEC 62471-1, which is the same standard used for lamp-based medical devices, was applied. Referring to IEC 62471-1 and the relevant literature and guidelines, this study derived the performance test method and test parameters for wavelength, the main performance item, and applied them to samples of the T company’s device. The derived performance evaluation item was the accuracy test of the wavelength according to IEC 62471-1. The test was to determine whether the LED and laser wavelengths were stable. The test standard was within ±5% in the error range corresponding to the test standard of T company. The test method checked whether the measured wavelength shown through the spectrometer was within the error range through the percentage. According to the measurement results, the light sources of the samples complied with the test standard set by the manufacturer. The measured wavelength for samples differed from the set value by 1 to 2 nm in red and blue LEDs, and the accuracy error range of the wavelength was close to 0%. However, laser and IR LEDs, unlike red and blue LEDs, had a difference as large as 2 to 7 nm between the set value and the measured value though they were used regardless of the mode. Moreover, the error range was also shown to be close to 1%, unlike those of the red and blue LEDs. To make it closer to the set value, it is necessary to study the wavelength of phototherapy devices whose accuracy is important for safety. We believe that a study on the wavelength of the phototherapy device similar to the set value would contribute to the safety of LED phototherapy devices and the industrial advancement of phototherapy devices similar to LED phototherapy devices.

Author Contributions

Conceptualization, J.-Y.L.; methodology, Y.-S.K.; verification, S.-G.P.; formal analysis, J.-Y.L.; survey, J.-Y.L.; resources, Y.-S.K.; data curation, J.-Y.L.; writing—original draft, J.-Y.L.; writing—review and editing, J.-Y.L.; visualization, J.-Y.L.; director, S.-G.P.; project management, S.-G.P.; funding, S.-G.P. All authors have read and agreed to the published version of the manuscript.

Funding

This research received no external funding.

Institutional Review Board Statement

Not applicable.

Informed Consent Statement

Not applicable.

Data Availability Statement

The data presented in this study are available in the article.

Conflicts of Interest

The authors declare no conflict of interest.

References

  1. Heo, C.-H. Development of a Safety Evaluation Guideline for Low Level Lasers; Korea Ministry of Food and Drug Safety, Seoul National University Bundang Hospital: Seongnam-si, Republic of Korea, 2007; Volume 3, pp. 154–196. [Google Scholar]
  2. Park, S.-L. Guidelines for the Application of Suitability Criteria within the Good Manufacturing Practice (GMP) Standards for Medical Devices; Korean Ministry of Food and Drug Safety: Cheongju, Republic of Korea, 2022.
  3. Avci, P.; Gupta, A.; Sadasivam, M.; Vecchio, D.; Pam, Z.; Pam, N.; Hamblin, M.R. Low-level laser (light) therapy (LLLT) in skin: Stimulating, healing, restoring. Semin. Cutan. Med. Surg. 2013, 32, 41–52. [Google Scholar] [PubMed]
  4. Mussttaf, R.A.; Jenkins, D.F.L.; Jha, A.N. Assessing the impact of low level laser therapy (LLLT) on biological systems: A review. Int. J. Radiat. Biol. 2019, 95, 120–143. [Google Scholar] [CrossRef] [PubMed]
  5. Lee, M.-K. LED Mask: 172 Cases of Burns and Other Side Effects… Need for Enhanced Safety Measures. Hankyung. 7 October 2020. Available online: https://www.hankyung.com/article/2020100750797 (accessed on 20 September 2023).
  6. Litscher, G. History of Laser Acupuncture: A Narrative Review of Scientific Literature. Med Acupunct. 2020, 32, 201–208. [Google Scholar] [CrossRef] [PubMed]
  7. Kim, W.-S.; Calderhead, R.G. Is Light-Emitting Diode Phototherapy (Led-Lllt) Really Effective? Laser Ther. 2011, 20, 205–215. [Google Scholar] [CrossRef] [PubMed]
  8. Hamblin, M.R. Shining light on the head: Photobiomodulation for brain disorders. BBA Clin. 2016, 6, 113–124. [Google Scholar] [CrossRef]
  9. Mester, A.; Mester, A. The History of Photobiomodulation: Endre Mester (1903–1984). Photomed. Laser Surg. 2017, 35, 393–394. [Google Scholar] [CrossRef] [PubMed]
  10. Huang, Y.-Y.; Chen, A.C.-H.; Carroll, J.D.; Hamblin, M.R. Biphasic Dose Response in Low Level Light Therapy. Dose-Response 2009, 7, 358–383. [Google Scholar] [CrossRef]
  11. Hamblin, M.R. Photobiomodulation or low-level laser therapy. J. Biophotonics 2016, 9, 1122–1124. [Google Scholar] [CrossRef] [PubMed]
  12. Jeong, J.-Y. Clinical Skin Care; MD World: Seoul, Republic of Korea, 2010. [Google Scholar]
  13. Chung, H.; Dai, T.; Sharma, S.K.; Huang, Y.Y.; Carroll, J.D.; Hamblin, M.R. The Nuts and Bolts of Low-level Laser (Light) Therapy. Ann. Biomed. Eng. 2012, 40, 516–533. [Google Scholar] [CrossRef]
  14. Kim, J.T.; Bae, S.B.; Yoon, D.H. LED Therapy Device for Skin Disease Treatment; Korea Electronics and Telecommunications Research Institute: Daejeon, Republic of Korea, 2010. [Google Scholar] [CrossRef]
  15. Good Books Chaoreum. National Institute of Medical Device Safety Information. Medical Device Regulatory Science (RA) Expert Volume 1: Pre-Market Approval; Good Books Chaoreum: Gwangju Metropolitan City, Republic of Korea, 2020. [Google Scholar]
  16. IEC 60601-1; Medical Electrical Equipment—Part 1: General Requirements for Basic Safety and Essential Performance. IEC: Geneva, Switzerland, 2016.
  17. IEC 60601-2-22; Medical Electrical Equipment—Part 2–22: Particular Requirements for the Basic Safety and Essential Perfor-mance of Surgical, Cosmetic, Therapeutic and Diagnostic Laser Equipment. IEC: Geneva, Switzerland, 2019.
  18. IEC 60825-1; Safety of Laser Products—Part 1: Equipment Classification, Requirements and User’s Guide. IEC: Geneva, Switzerland, 2014.
  19. IEC 60825-8; Safety of Laser Products—Part 8: Guidelines for Safe Use of Medical Laser Equipment. IEC: Geneva, Switzerland, 2022.
  20. Korea Ministry of Food and Drug Safety Notice No. 2020-29. Common Standards and Specifications for Electrical and Mechanical Safety of Medical Devices. Effective Date: 1 May 2020. Available online: https://www.law.go.kr/%ED%96%89%EC%A0%95%EA%B7%9C%EC%B9%99/%EC%9D%98%EB%A3%8C%EA%B8%B0%EA%B8%B0%EC%9D%98%20%EC%A0%84%EA%B8%B0%C2%B7%EA%B8%B0%EA%B3%84%EC%A0%81%20%EC%95%88%EC%A0%84%EC%97%90%20%EA%B4%80%ED%95%9C%20%EA%B3%B5%ED%86%B5%EA%B8%B0%EC%A4%80%EA%B7%9C%EA%B2%A9 (accessed on 20 September 2023).
  21. IEC 62471-1; Photobiological Safety of Lamps and Lamp Systems. IEC: Geneva, Switzerland, 2006.
Figure 1. LED phototherapy device sample.
Figure 1. LED phototherapy device sample.
Applsci 13 12215 g001
Figure 2. Wavelength measurement for the light source.
Figure 2. Wavelength measurement for the light source.
Applsci 13 12215 g002
Figure 3. Wavelength performance test devices (left: a spectrometer; right: the sensor of the spectrometer).
Figure 3. Wavelength performance test devices (left: a spectrometer; right: the sensor of the spectrometer).
Applsci 13 12215 g003
Figure 4. Results of laser wavelength measurements by mode obtained from T company’s samples.
Figure 4. Results of laser wavelength measurements by mode obtained from T company’s samples.
Applsci 13 12215 g004
Figure 5. Enhanced graph of laser wavelengths measured per mode in T company’s samples: Sample 1 (the black line represents H mode, and the blue line represents S mode), and Sample 2 (the red line corresponds to H mode, and the green line corresponds to S mode).
Figure 5. Enhanced graph of laser wavelengths measured per mode in T company’s samples: Sample 1 (the black line represents H mode, and the blue line represents S mode), and Sample 2 (the red line corresponds to H mode, and the green line corresponds to S mode).
Applsci 13 12215 g005
Figure 6. LED wavelength graph measured in sample modes of T company’s device.
Figure 6. LED wavelength graph measured in sample modes of T company’s device.
Applsci 13 12215 g006
Figure 7. IR LED wavelength graph measured in the samples of T company’s device, Sample 1 (the black line represents H mode, and the blue line represents S mode), and Sample 2 (the red line corresponds to H mode, and the green line corresponds to S mode).
Figure 7. IR LED wavelength graph measured in the samples of T company’s device, Sample 1 (the black line represents H mode, and the blue line represents S mode), and Sample 2 (the red line corresponds to H mode, and the green line corresponds to S mode).
Applsci 13 12215 g007
Figure 8. Red LED wavelength graph measured in the samples of T company’s device, Sample 1 (the black line) and Sample 2 (the red line).
Figure 8. Red LED wavelength graph measured in the samples of T company’s device, Sample 1 (the black line) and Sample 2 (the red line).
Applsci 13 12215 g008
Figure 9. Blue LED wavelength graph measured in the samples of T company’s device. Sample 1 (the black line) and Sample 2 (the red line).
Figure 9. Blue LED wavelength graph measured in the samples of T company’s device. Sample 1 (the black line) and Sample 2 (the red line).
Applsci 13 12215 g009
Table 1. Laser classes according to IEC 60825-1 [18].
Table 1. Laser classes according to IEC 60825-1 [18].
ClassesExplanationWavelengthOutput
(mW)
1It is not dangerous to direct the laser beam at the human body.Visible
Invisible
-
1MIt is dangerous to see or direct laser beams with optical devices.
(However, products with access emission limits of Class 1)
1CProducts that directly use lasers for treatment or diagnosis of the skin or internal tissue. Eye exposure can be controlled by engineering methods, even if the emitted laser is rated 3R, 3B, and 4th.
2When laser beams are irradiated in the eyes, they can be protected with a blink of 0.25 s.
(However, products with access emission limits of 0.25 s or less in pulse width are rated 1.)
Visible<1
2MIt is dangerous if laser beams are directly used with optical devices.
(However, the restriction on access emission is a Class 2 product.)
3RIt is dangerous if laser beams are directly used on the eyes.Visible
Invisible
1~5
3BIt is dangerous if laser beams are directly used on the human body.Visible
Invisible
5~500
4It is dangerous if laser beams are directly used on the human body.Visible
Invisible
>500
Table 2. Derivation of performance test items for LED optical medical devices.
Table 2. Derivation of performance test items for LED optical medical devices.
Test ItemTest StandardTest Method
Wavelength accuracy testWithin the
error range ±□%
Measure the wavelength of the LED light source using the photodetector. Check whether the measured wavelength value is suitable in terms of the test standard after calculating by applying the measured and the actual values to the equation below:
A c c u r a c y ( % ) =   ( A c t u a l v a l u e M e a s u r e d v a l u e ) / A c t u a l v a l u e ×   100
Output accuracy testWithin the
error range ±□%
Measure the output of the LED light source using a power meter. After calculation, substitute the measured output value and the actual value in the formula below to check if it meets the test standards.
A c c u r a c y ( % ) =   ( A c t u a l v a l u e M e a s u r e d v a l u e ) / A c t u a l v a l u e ×   100
Illuminance measurementIEC 62471-1.4. Follow the exposure limit standards for harm.In accordance with IEC 62471-1 5. Measurement of lamps and lamp devices. Measure the illuminance of the LED light source using a luminance meter.
Table 3. Measured values by mode in the samples of T company’s device and the accuracy of wavelength.
Table 3. Measured values by mode in the samples of T company’s device and the accuracy of wavelength.
WavelengthLASER
Set Value:
650 nm
IR LED
Set Value:
845 nm
RED LED
Set Value:
625 nm
BLUE LED
Set Value:
455 nm
ModeSample 1Sample 2Sample 1Sample 2Sample 1Sample 2Sample 1Sample 2
H Mode654 nm652 nm838 nm840 nm627 nm627 nm--
S Mode654 nm653 nm838 nm839 nm--455 nm454 nm
Accuracy of wavelengthH±0.615%±0.31%±0.828%±0.59%±0.32%±0.32%--
S±0.615%±0.461%±0.828%±0.71%--±0%±0.219%
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content.

Share and Cite

MDPI and ACS Style

Lee, J.-Y.; Kwon, Y.-S.; Park, S.-G. Application of Performance Test Method in Korea for LED Optical Medical Device Samples. Appl. Sci. 2023, 13, 12215. https://doi.org/10.3390/app132212215

AMA Style

Lee J-Y, Kwon Y-S, Park S-G. Application of Performance Test Method in Korea for LED Optical Medical Device Samples. Applied Sciences. 2023; 13(22):12215. https://doi.org/10.3390/app132212215

Chicago/Turabian Style

Lee, Jae-Young, Yi-Suk Kwon, and Sang-Geon Park. 2023. "Application of Performance Test Method in Korea for LED Optical Medical Device Samples" Applied Sciences 13, no. 22: 12215. https://doi.org/10.3390/app132212215

Note that from the first issue of 2016, this journal uses article numbers instead of page numbers. See further details here.

Article Metrics

Back to TopTop