Next Article in Journal
Experimental Study on the Water Absorption, Compaction Effect, and Pull-Out Bearing Characteristics of Water-Absorbing and Compaction Anchoring Bolts
Previous Article in Journal
Performance Analysis of a Communication Failure and Repair Mechanism with Classified Primary Users in CRNs
 
 
Font Type:
Arial Georgia Verdana
Font Size:
Aa Aa Aa
Line Spacing:
Column Width:
Background:
Article

Design and Experimental Study of a Down-Drive Piezoelectric High-Frequency Fatigue Testing Machine

School of Mechanical and Aerospace Engineering, Jilin University, No. 5988, Renmin Avenue, Changchun 130025, China
*
Author to whom correspondence should be addressed.
Appl. Sci. 2024, 14(16), 6961; https://doi.org/10.3390/app14166961
Submission received: 7 July 2024 / Revised: 25 July 2024 / Accepted: 26 July 2024 / Published: 8 August 2024

Abstract

:
In order to solve the problem of large dynamic load and low loading accuracy prevailing in traditional fatigue testing machines, and to meet the fatigue test loading demand of alternating small-load microelectromechanical devices and bionic devices, a high-frequency resonant piezoelectric fatigue tester that can apply alternating small loads in the range of 0.1 N~100 N was designed and manufactured, with a piezoelectric oscillator as the driving source of the system. Firstly, the structure of this fatigue testing machine is designed, and a dynamic model is established for the fatigue testing machine to theoretically study the influence of the system mass and stiffness on the change in the resonance frequency of the fatigue testing machine. Then, numerical calculations and experiments are carried out to verify the effect of system mass and stiffness changes on the intrinsic frequency of the fatigue testing machine. Finally, the fatigue testing machine was used to carry out fatigue test on the dog-bone-shaped specimens, and the failure process of specimen fatigue was verified by analysing the load applied to the specimen by the fatigue testing machine, and at the same time, it was proved that the performance of the present fatigue testing machine complied with the requirements. It provides a new technical means to meet the fatigue test loading demand of small-load electromechanical devices and bionic devices.

1. Introduction

In recent years, with the development of microelectromechanical technology and biomimetic technology, products such as microactuators [1,2], microsensors [3], and bionic fish [4] have emerged. The development of technologies for testing the mechanical properties of these products under near-service conditions has become extremely urgent. Fatigue testing is an important method to study the mechanical properties of materials [5,6]. Due to the complexity of the fracture conditions of materials [7], fatigue testers have been designed to test them in bending [8,9], torsion [10], or tension [11,12]. Tensile fatigue studies of products are very helpful in recognizing the safety of steel sheet assemblies, and uniaxial test methods are still widely used compared to the complexity of biaxial bending tests. In the case where uniaxial pure tensile loading is considered, the specimen is clamped in the fixture of a fatigue testing machine, and cyclic tensile stresses at a fixed frequency generated by the fatigue testing system are applied to the specimen in order to fail the specimen and calculate the fatigue life.
The variable load applied to the specimen by the fatigue tester is essentially generating a vibratory load. In the current research, uniaxial fatigue tester drives are mainly categorized into three types, “electromagnetic drive” [13,14], “electro-hydraulic servo drive” [15,16], and “piezoelectric drive” [17]. Among them, the electromagnetically driven type is characterized by fast response and is suitable for routine material testing, and it is also the most commonly used fatigue test equipment; The electro-hydraulic servo-driven type has a larger output power and is suitable for the inspection of large, medium, and heavy loaded parts, but with a low operating frequency and a long testing time, the range of applications is more restricted. Piezoelectric materials are often used as high-speed and -precision actuators due to their fast response, lack of electromagnetic interference, high resolution, and high deformation accuracy [18]. As a result, researchers favour fatigue testing machines that use piezoelectric materials as the drive source. Piezoelectric fatigue testers can be categorized into three types: the first one is piezoelectric ceramic ultrasonic transducer-driven fatigue testers [19,20], the second is piezoelectric stack-driven fatigue testers [21], and the third one is piezoelectric vibrator-driven fatigue testers [22].
Zeng [17] designed a circular piezoelectric vibrator-driven fatigue testing machine, adjusting the frequency of the alternating voltage acting on the piezoelectric vibrator to induce the resonance of the whole system, and the inertial force generated by the mass of the elastic loader in the resonance state is reciprocated on the specimen, thus realizing fatigue testing. Li Kang [23] designed a piezoelectric-driven horizontal fatigue tester, which avoids the interference of the mass of the fixture in the vertical structure for static or dynamic loading. Ma Lei [24] proposed a new type of piezoelectric high-frequency micro load fatigue testing machine; the specimen is only subjected to alternating stress in the horizontal direction, avoiding the influence of other factors on the loading of the specimen, to ensure that the specimen fatigue test results are accurate. Most of the existing fatigue testing machines at home and abroad are designed for macro-scale devices, the dynamic loading is generally more than 500 N; the development of a more mature piezoelectric fatigue testing machine with a dynamic load greater than 100 N, and the development of piezoelectric fatigue testers for the range of less than 100 N, are still in the exploration stage.
Aiming at the problems of large motorised load and low loading accuracy of traditional fatigue test, this paper designs a high-frequency piezoelectric fatigue tester which can apply alternating small loads in the range of 0.1 N~100 N, establishes the kinetic model of its system, and carries out the theoretical and simulation analysis of the system, and finally fabricates the fatigue test prototype and conducts experimental tests on the dog-bone-shaped steel plate specimens. By analysing the load applied to the specimen by the fatigue testing machine, the failure process of specimen fatigue is verified, and at the same time, it is proved that the performance of this fatigue testing machine meets the requirements: to provide test methods for microelectromechanical devices and bionic devices to obtain the dynamic performance parameters under service conditions, and to provide technical means to reveal the related mechanical behaviour, deformation damage mechanism and performance evolution law under small loads.

2. Materials and Methods

2.1. Structural Design

The structure of the piezoelectric fatigue testing machine is shown in Figure 1, which mainly consists of a piezoelectric vibrator, preset load adjusting mechanism, force detecting mechanism, clamping mechanism, specimen, dynamic load adjusting mechanism, and fixing mechanism, as shown in Figure 1a. The piezoelectric vibrator provides the exciting force for the system and consists of a circular substrate and two circular piezoelectric certificates. Among them, the upper and lower circular piezoelectric ceramics are polarized in opposite directions, and under the action of external alternating voltage, the upper and lower circular piezoelectric ceramics will vibrate due to the inverse piezoelectric effect, as shown in Figure 1b. The dynamic load adjustment mechanism is set up to adjust the resonance frequency and force of the resonance system, which mainly consists of a vibration transmitting plate, flexibility loop, lower support plate, vibration transmitting rod, driving and counterweight parts, and each part is connected together by the vibration transmitting rod, as shown in Figure 1c. While the core part of the dynamic load adjustment mechanism is the drive and counterweight part, the piezoelectric actuator is mounted in the lower part of the fatigue tester and transmits the excitation force to the elastic ring through a connecting shaft with a diameter of 5 mm. The piezoelectric actuator consists of a double-layered piezoelectric vibrator with a sandwich, in which the piezoelectric ceramic and metal are bonded together with epoxy resin. The inverse piezoelectric effect of the piezoelectric vibrator generates alternating excitation forces for the actuator, which are amplified by two elastic rings. In addition, in order to vary the driving frequency, two identical semi-cylindrical steel mass blocks were mounted on the upper plate of the piezoelectric actuator and secured with bolts of 3 mm in diameter, as shown in Figure 1d.

2.2. Working Principle

Although fatigue testing machines have numerous successful applications, they can be traditionally categorized into three types based on the way energy is utilized when loading a specimen. The first type is the excitation energy recovery test system [25], where the specimen is loaded on a rotating eccentric wheel, where the released potential energy from the previous loading cycle is transferred as kinetic energy for the next cycle. The second type is the unincentivized energy recovery test system [26], with constant pressure and rate test loads induced by an electrohydraulic servo control system. Therefore, whether the test system is loaded or not, it still consumes a fixed amount of excitation energy. In contrast to a recycled test system, there is no energy to reuse here, and in addition, continuous testing generates waste heat, which requires additional energy for cooling. The last type is the resonant test system [27], where the driving frequency of the tester is designed to be close to the intrinsic frequency of the test system, which is the intrinsic vibration frequency of the system itself, and whose value is independent of the initial conditions and the magnitude of the external excitation to which it is subjected. In this design, a frequency is applied to the system to excite the resonance of a linear system or the subharmonic oscillations of a nonlinear system, and by applying a small excitation force, the amplitude can be increased so that the energy consumption of the tester is mainly focused on the resonance active damping. This fatigue tester consumes less energy than the first or second previously mentioned test systems.
In this paper, a small-load fatigue tester based on a Type III resonance test system is designed to use a piezoelectric vibrator as the driving force source of the system, which generates vibrations under the action of an applied alternating electric voltage. Adjust the frequency of the alternating voltage applied to the piezoelectric vibrator, prompting the vibration system consisting of the base, guide rod, preset load adjustment mechanism, force detection mechanism, clamping mechanism, specimen, dynamic load adjustment mechanism, and fixing mechanism to resonate. At this time, the inertial force generated by the counterweight mass block in the resonance state reciprocates on the specimen. In addition to this, the inertia force generated by the mass of the counterweight disk, the piezoelectric vibrator, and the lower disk in the status of resonance also acts on the specimen at the same time, thus realizing fatigue testing.
The fatigue testing machine needs to determine the force applied to the specimen in real time during fatigue testing, so it is necessary to build a force testing system as shown in Figure 2, which consists of transducers, boosters, 24VDC power supply, acquisition card, and acquisition software (V190521).
The transducer works on the principle of resistance strain-type, when subjected to external forces, the deformation of its internal elastomer, causing changes in the resistance value of the resistance strain gauges, the transducer’s built-in circuitry will be converted into a change in resistance to voltage changes and output to the booster, which will be converted into a standard voltage signal. The acquisition card is powered by USB and acquires the voltage signal output from the booster and transfers it to the computer via the USB 2.0 bus. The acquisition software displays and records the loads acting on the sensors in real time, as shown in Figure 3 for the interface diagram of the acquisition software.

2.3. Material Parameters

2.3.1. Piezoelectric Vibrator

The PZT-8 is often used as a Lanzmann sensor because of its fast response characteristics. In this study, piezoelectric vibrator ceramics were chosen for similar reasons. Figure 4 shows the cross section of the piezoelectric vibrator considering the thickness of the piezoelectric ceramic (Tc) and the thickness of the spring steel substrate (Tm). Table 1 shows its material parameters. Table 2 shows its matrix parameter.

2.3.2. Mass Block

A semi-cylindrical steel mass block is shown in Figure 5, which is made of 45# steel with thickness dimensions including 10 mm, 20 mm, and 30 mm, and the appropriate thickness is selected according to the need.

2.3.3. Elastic Ring

As shown in Figure 6, the elastic ring is divided into three different zones according to its thickness. The middle part is thicker and is used to rigidly connect the other components of the fatigue tester. The left and right ends have thickened arcs designed to prevent damage due to stress concentration on both sides of the elastic ring during operation. In addition to these areas, the part connecting the middle and ends of the elastic ring forms the third part, which is thinner. The stiffness of the elastic ring increases as the thickness of this part increases. By changing the thickness of the elastic ring, the stiffness of the fatigue tester can be adjusted, thus changing the intrinsic frequency of the tester.
Since the output force of the piezoelectric vibrator is relatively small, the size Tb at the elasticity of the elastic ring should not be too large. Take Tb min = 0.1 mm, Tb max = 1 mm, with 0.1 mm spacing, and the elastic ring material is 65 Mn spring steel.

2.3.4. Dog-Bone Steel Specimens

As shown in Figure 7 for the dog-bone-shaped steel specimen, since Q235b is a commonly used engineering material and its mechanical properties are very typical, the range of stresses applied to the specimen by the fatigue testing machine and the fatigue performance under alternating stresses are investigated using Q235b as an example. The chemical properties are shown in Table 3. Table 4 shows its mechanical properties.

3. System Dynamics Modelling and Simulation Analysis

3.1. Structural Design

A fatigue tester is a typical example of a resonant system. In order to simplify the calculation process, the test machine was discretized into several centralized mass cells. According to the actual working conditions, select the parts that transmit more force and displacement as the centralized mass unit, ensure that the stiffness within each unit is consistent or similar, and avoid the appearance of relatively moving parts within the same unit. Therefore, when calculating the vibration characteristics of a fatigue testing machine in terms of a mass–spring–damping system, this method not only ensures the accuracy of the calculation but also reduces the cost of the calculation. Table 5 describes the components of the centralized mass element subsystem. Figure 8 shows the spring-mass-damping system model of the fatigue testing machine.
Based on the above relationship for each mass unit, the ordinary differential equations are listed:
m 1       m 2       m 3 x ¨ 1 x ¨ 2 x ¨ 3 + c 1 + c 2 c 2   c 2 c 2 + 2 c 3 c 4 c 4   c 4 c 4 x ˙ 1 x ˙ 2 x ˙ 3 + k 1 + k 2 k 2   k 2 k 2 + 2 k 3 k 4 k 4   k 4 k 4 x ˙ 1 x ˙ 2 x ˙ 3 = 0 0 F sin ( w t )
where k1 is the transducer stiffness; k2 is the elastic ring stiffness. K3 is the sum of the connecting shaft stiffness and the piezoelectric vibrator stiffness.
In fatigue systems, changing the differential equations of a discrete system can be carried out by changing the intrinsic frequency of the system by changing the mass of the mass block, since the mass block is replaceable. Five values were averaged over the 0.999 kg to 3.999 kg range to perform a comprehensive frequency sweep analysis from 0 to 250 Hz. Using this method, it is possible to check the correlation between the mass change and the corresponding intrinsic frequency. The force load–frequency diagrams for the five mass blocks are shown in Figure 9; the greater the mass of the mass blocks, the lower the intrinsic frequency of the system.
Express the piezoelectric vibrator simple harmonic excitation force F(t) as:
F ( t ) = F e i w t ,
Then the steady-state response of this system can be expressed as:
x 1 ( t ) = X 1 e i w t , x 2 ( t ) = X 2 e i w t , x 3 ( t ) = X 3 e i w t ,
where X1 and X2 are complex quantities, taking values related to the system parameters, piezoelectric vibrator simple harmonic excitation force.
Equations (2) and (3) are brought into the vibrational differential Equation (1) and expressed in matrix form:
Z 11 ( w ) Z 12 ( w ) Z 13 ( w ) Z 21 ( w ) Z 22 ( w ) Z 23 ( w ) Z 31 ( w ) Z 32 ( w ) Z 33 ( w ) X 1 X 2 X 3 = 0 0 F ,
where
Z 11 ( w ) = w 2 m 1 + i w ( c 1 + c s ( t ) ) + k 1 + k s ( t ) Z 12 ( w ) = Z 21 ( w ) = i w ( c s ( t ) ) k s ( t ) Z 13 ( w ) = Z 31 ( w ) = 0 Z 22 ( w ) = w 2 m 2 + i w ( c s ( t ) + 2 c 2 + c 3 ) + k s ( t ) + 2 k 2 + k 3 Z 23 ( w ) = Z 32 ( w ) = i w ( c 3 ) k 3 Z 33 ( w ) = w 2 m 3 + i w ( c 3 ) + k 3 ,
If,
Z ( w ) = Z 11 ( w ) Z 12 ( w ) Z 13 ( w ) Z 21 ( w ) Z 22 ( w ) Z 23 ( w ) Z 31 ( w ) Z 32 ( w ) Z 33 ( w ) ,
then Equation (4) can be expressed as:
Z ( w ) X = F ,
where Z ( w ) is the impedance matrix, X is the displacement column vector, and F is the excitation amplitude column vector.
X = Z ( w ) 1 F ,
where
Z ( w ) 1 = 1 D a 11 a 12 a 13 a 21 a 22 a 23 a 31 a 32 a 33 ,
where
D = Z 11 ( w ) Z 22 ( w ) Z 33 ( w ) Z 11 ( w ) Z 23 2 ( w ) Z 33 ( w ) Z 12 2 ( w ) a 11 = Z 22 ( w ) Z 33 ( w ) Z 23 2 ( w ) a 12 = a 21 = Z 12 ( w ) Z 33 ( w ) a 13 = a 31 = Z 12 ( w ) Z 23 ( w ) a 22 = Z 11 ( w ) Z 33 ( w ) a 23 = a 32 = Z 11 ( w ) Z 23 ( w ) a 33 = Z 11 ( w ) Z 22 ( w ) Z 12 2 ( w ) ,
Thus, the steady–state response amplitude of the system can be obtained as a function of the excitation frequency w:
X 1 ( w ) = Z 12 ( w ) Z 23 ( w ) F Z 11 ( w ) Z 22 ( w ) Z 33 ( w ) Z 11 ( w ) Z 23 2 ( w ) Z 33 ( w ) Z 12 2 ( w ) X 2 ( w ) = Z 11 ( w ) Z 23 ( w ) F Z 11 ( w ) Z 22 ( w ) Z 33 ( w ) Z 11 ( w ) Z 23 2 ( w ) Z 33 ( w ) Z 12 2 ( w ) X 3 ( w ) = ( Z 11 ( w ) Z 22 ( w ) Z 12 2 ( w ) ) F Z 11 ( w ) Z 22 ( w ) Z 33 ( w ) Z 11 ( w ) Z 23 2 ( w ) Z 33 ( w ) Z 12 2 ( w ) ,
When the system damping is neglected, the steady-state response of the system can be found with respect to the excitation frequency w:
x 1 ( t ) = k s ( t ) k 3 sin w t D n x 2 ( t ) = k 3 sin w t D n w 2 m 1 + k 1 + k s ( t ) x 3 ( t ) = sin w t D n [ w 2 m 1 + k 1 + k s ( t ) w 2 m 2 + k s ( t ) + 2 k 2 + k 3 k s 2 ( t ) ] ,
where
D n = w 2 m 1 + k 1 + k s ( t ) w 2 m 2 + k s ( t ) + 2 k 2 + k 3 w 2 m 3 + k 3 w 2 m 1 + k 1 + k s ( t ) k 3 2 w 2 m 3 + k 3 k s 2 ( t ) ,

3.2. Simulation Analysis

In order to further investigate theoretically the influence of the mass M of the counterweight mass block and the stiffness k2 of the elastic ring on the vibration characteristics of the fatigue tester, the relevant parameters were determined, as shown in Table 6.
According to the above Equation (1), the elastic ring stiffness and mass block mass are used as parameters affecting the intrinsic frequency of the fatigue testing machine, by making Dn = 0 in Equation (13), the intrinsic frequency of the system can be solved for different parameters, The relationship is shown in Figure 10.
From Figure 10, it can be seen that for the same counterweight mass block mass, increasing the elastic ring stiffness (corresponding to increasing the thickness of the elastic ring at the flexible place Tb in the simulation analysis) gradually increases the system intrinsic frequency; for the same elastic ring stiffness, increasing the mass of the counterweight mass block (which corresponds to increasing the thickness of counterweight mass block a in the simulation analysis) results in a gradual decrease in the intrinsic frequency of the system, which is in accordance with the facts and with the results of the simulation analysis. The overall analysis of the fatigue testing machine was carried out using the finite element method, and the results are shown in Figure 11. The fatigue tester achieved the optimal solution of the modal vibration pattern at an operating frequency of 187 Hz. At this frequency, the piezoelectric actuator vibrates the fatigue tester vertically up and down with no displacement of the rest of the machine, as shown in Figure 12.
According to the above simulation process, the specimen with the central width w = 0.8 mm, the thickness of the counterweight mass block a = 0 mm:2 mm:30 mm, and the thickness of the elastic ring at the flexible place b = 0.1 mm:0.1 mm:1 mm were chosen, and the three-dimensional relationship between the intrinsic frequency of the fatigue testing machine and the thicknesses of the counterweight mass block a, and the thickness of the elastic ring at the flexible place b was obtained, which is shown in Figure 13.
From Figure 13 it can be seen that for the same counterweight mass block thickness a, increase the thickness of the elastic ring flexible place b, and fatigue testing machine system intrinsic frequency will gradually increase; for the same elastic ring flexible thickness b, increase the counterweight mass block thickness a, and fatigue testing machine system intrinsic frequency will gradually decrease.

4. Trial Tests

4.1. Test System Construction and Test Methods

Using the developed piezoelectric fatigue testing machine system, several sets of specimens of different sizes were processed on the basis of selected materials, the influence laws of different parameters on the resonance frequency of the system, and the stresses acting on the specimens were investigated, and finally the specimens were subjected to tension–tension fatigue tests. Shown in Figure 14 is the fatigue testing machine performance test system, including fatigue testing machine, signal generator (UTI-T, UTG7052B), phase shifter, power amplifier (SA-PA010), computer, impedance analyser (ZX70A), transducer (FCW-2620), amplifier (T098H), acquisition card (USB DAQ-7606I), and acquisition software (V190521). The characteristics and output performance of the piezoelectric fatigue tester system were investigated by selecting specimens with different stiffnesses, counterweights with different masses, and setting different preset loads and driving voltages. The dimensions of the fatigue testing machine are shown in Table 7.
Prior to the start of the test, the specimen is clamped between the upper and lower fixtures, a suitable initial static load is applied to the specimen using the preset load adjustment mechanism, and the static load is maintained using the upper locking bolt. A precision impedance analyser is used to determine the intrinsic frequency of the system at this point, and an alternating voltage is applied to the piezoelectric vibrator at the intrinsic frequency. After continuously increasing the RMS value of the driving voltage, the transducer is utilized to determine the alternating load applied to the specimen at different voltages and recorded in real time using the VI acquisition software; The characteristics and output performance of the fatigue testing machine system are investigated by means of specimens of different stiffness.

4.2. System Intrinsic Frequency Impact Factor Testing

The specimen with the width w = 0.8 mm in the middle is selected, and the impedance characterization is carried out under different combinations of counterweight mass block and elastic ring, respectively, and the measured results are compared with the simulation and theoretical results. The test curve graph is shown in Figure 15.
From Figure 15, it can be clearly seen that increasing the thickness a of the counterweight mass block gradually decreases the intrinsic frequency of the system for the same thickness Tb at the elastic ring flexibility. For the same counterweight mass block thickness a, increasing the thickness Tb at the flexible place of the elastic ring, the intrinsic frequency of the system increases gradually, which is consistent with the facts and with the trend of the simulation and theoretical results. Moreover, the measured value of the system’s intrinsic frequency is closer to the simulated and theoretical values, and the maximum error is kept within 10 Hz, which verifies the correctness of the simulation and theoretical analysis.

4.3. Specimen Force Impact Test

In order to study the relationship between the voltage and the stiffness of the specimen on the stresses acting on the specimen, an elastic ring with a thickness Tb = 0.7 mm at the flexible place was selected, and the preset load F0 = 10 N. The thicknesses a of the counterweight mass block were taken to be 0 mm, 10 mm, 20 mm, and 30 mm, respectively. The relationship of its effect on the stresses acting on the specimen by the fatigue testing machine was tested by selecting different specimens (w = 0.8 mm, 2 mm, and 3 mm) and voltages. The test relationship is shown in Figure 16.
From Figure 16, it can be concluded that the stress amplitude acting on the specimen gradually increases with the increase in driving voltage, At the same counterweight mass block and the same voltage, the amplitude of the stress acting on the specimen gradually increases as the specimen stiffness increases. When the thickness of the specimen w is 3 mm, the thickness of the counterweight mass block a is 30 mm, and the maximum stress acting on the specimen is about 14.8 N at a driving voltage of 60 V peak to peak.

4.4. Tests of the Effect of Preset Loads on Specimen Forces

In order to study the relationship between voltage and preset load on the stress acting on the specimen, based on the elastic ring with thickness Tb = 0.7 mm at the flexible place and the specimen with thickness w = 2 mm in the middle, when the counterweight mass block a was taken as 0 mm, 10 mm, 20 mm, and 30 mm, respectively, the effect of the preset load (F0 = 20 N, 40 N, and 60 N) and the voltage was tested on the stress acting on the specimen in the fatigue tester by setting up different preset loads. The relationship is shown in Figure 17.
As can be seen from Figure 17, the stress amplitude acting on the specimen increases gradually with the increase in driving voltage, and the stress amplitude acting on the specimen increases gradually when the preset load increases at the same counterweight mass block and the same voltage. The stress amplitude acting on the specimen ranges from about 1 N to 32 N within a driving voltage peak-to-peak range of 1 V to 200 V.

4.5. Fatigue Testing Machine Fatigue Testing Performance Verification

Finally, in order to verify the fatigue testing performance of the fatigue testing machine, a specimen with a central width w = 0.8 mm was selected and subjected to tensile–extensional fatigue testing with the following test steps: The load Fb applied to three or more specimens of the same specification at the time of damage is measured in advance and the average value is calculated; A suitable stress ratio R is selected and the maximum, minimum, and average loads applied to the specimen are determined according to the stress levels at each level; Determine the value of the applied preset load based on the average load corresponding to the stress levels at each level; adjust the signal generator to resonate the vibration system; The power supply voltage is adjusted so that the force applied to the specimen meets both the maximum and minimum load requirements; At this time to start timing, to be specimen fatigue rupture time accumulation and driving voltage frequency of the product of the number of cycles of the load. In determining the strength limit Fb of the specimen, limited by the limited test conditions, the specimen was manually tensile-tested using this fatigue testing machine, in which the loading was carried out as slowly and uniformly as possible. The three tensile test curves are shown in Figure 18.
From Figure 18, it can be seen that the strength of the specimen increases steeply as the specimen is subjected to an increase in tensile time, and after the yielding stage, the fracture strength of the specimen returns to 0. From this, it can be determined that the average strength limit of the specimen is F b 120 N .
In accordance with the maximum load Fmax = 0.55, Fb = 66 N load level on the specimen for a set of fatigue tests, and through the acquisition software real-time recording of the dynamic load acting on the specimen, and then after using MATLAB software for the processing of the collected load data, the results of the test results obtained by the curve are shown in Figure 19.
As can be seen from Figure 19, the fatigue test load curve can be divided into three more obvious stages. In the first 500 s range, the load acting on the specimen does not change significantly with time. In the range of 500 s~780 s, the static and dynamic loads on the specimen are attenuated more obviously. After 780 s, the static and dynamic loads acting on the specimen change drastically. These three stages correspond to the generally recognized three stages of fatigue fracture: crack initiation, crack extension, and fatigue fracture. This phenomenon is preliminarily analysed due to the change in stiffness parameters of the specimen during the fatigue test. It can also be seen that the designed fatigue testing machine basically meets the demand for fatigue loading of tiny devices and can realize the function of fatigue testing for them.

5. Conclusions

In response to the need for a piezoelectric fatigue tester capable of generating small alternating loads, a down-drive piezoelectric high-frequency fatigue testing machine prototype was developed with a load range of 0.1 N to 100 N. Firstly, the structural composition and working principle of the device were analysed, and a spring–damping system model was established to analyse the effects of mass block mass and spring stiffness on the intrinsic frequency of the fatigue testing machine. Next, the experimental platform was set up, and tensile fatigue tests were performed on steel specimens in the shape of a dog bone. The test results show that the amplitude of the stress acting on the specimen increases gradually with the increase in the driving voltage, the increase in the specimen stiffness, the increase in the preset load, and the increase in the fast mass of the counterweight mass. Through the load change curves during the fatigue test, it was found that both the static and dynamic loads acting on the specimen decayed after 500 s. The designed piezoelectric fatigue testing machine is experimentally verified to be suitable for fatigue testing with small dynamic loads and high loading accuracy. It provides a technical reference for subsequent scholars to study the piezoelectric fatigue testing machine in the range of less than 100 N, which has great application value in the field of fatigue testing technology for reliability and durability determination of microelectromechanical devices and bionic devices products.

Author Contributions

Methodology, Z.Y.; writing—original draft preparation, S.Z.; formal analysis, J.L. All authors have read and agreed to the published version of the manuscript.

Funding

This research was funded by the Natural Science Foundation of China, grant number 51875234.

Institutional Review Board Statement

Not applicable.

Informed Consent Statement

Not applicable.

Data Availability Statement

The data used to support the findings of this study are available from the corresponding author upon request.

Conflicts of Interest

The authors declare no conflicts of interest.

References

  1. Wu, B.; Chu, W.; Xia, B.; Zhou, Y. Construction of Spring-Shaped UHMWPE Fiber-Based Soft Actuators with Stable/Fast Actuating Response and Large Actuating Stroke. ACS Appl. Polym. Mater. 2024, 6, 5216–5225. [Google Scholar] [CrossRef]
  2. Tian, X.; Sun, Y.; Li, Z.; Wang, H.; Wang, Z.; Wang, H.; Zhu, J.; Yang, Z. Design and Experiment Using Flexible Bumps for Piezoelectric-Driven Hydraulically Amplified Braille Dot Display. Micromachines 2021, 12, 795. [Google Scholar] [CrossRef] [PubMed]
  3. Tian, X.; Liu, J.; Hou, J.; Gai, H.; Yang, J.; Sun, Z. Study of the Power Generation Performance of Impact Piezoelectric Energy Capture Devices. Micromachines 2023, 14, 1013. [Google Scholar] [CrossRef] [PubMed]
  4. Cha, Y.; Chae, W.; Kim, H.; Walcott, H.; Peterson, S.D.; Porfiri, M. Energy harvesting from a piezoelectric biomimetic fish tail. Renew. Energy 2016, 86, 449–458. [Google Scholar] [CrossRef]
  5. Kattimani, M.A.; Khatib, M.I.; Ghori, M.M.; Sajjad, M.; Jahangir, S.; Baqtaiyan, H.; Sadiq, M. Design and fabrication of fatigue testing machine. Int. J. Sci. Res. Sci. Eng. Technol. 2020, 7, 295–304. [Google Scholar] [CrossRef]
  6. Huňady, R.; Sivák, P.; Delyová, I.; Bocko, J.; Vavro, J.; Hroncová, D. Upgrade of the Universal Testing Machine for the Possibilities of Fatigue Tests in a Limited Mode. Appl. Sci. 2024, 14, 3973. [Google Scholar] [CrossRef]
  7. Zhang, K.; Collette, M. Experimental investigation of structural system capacity with multiple fatigue cracks. Mar. Struct. 2021, 78, 102943. [Google Scholar] [CrossRef]
  8. Khemaleelakul, T.; Kanjantra, P.; Sirimongkolwattana, S.; Thongngarm, W.; Mamanee, T.; Sirikulrat, N. Comparative Three-Point Bending Flexural Test of Dental Ceramics Using Standard and Self-Adjustable Specimen Support Fixtures. Mater. Trans. 2020, 61, 685–690. [Google Scholar] [CrossRef]
  9. Mishra, A.K.; Lambai, A.; Jayaram, V.; Jaya, B.N. The edge-notched clamped beam bend specimen as a fracture toughness test geometry. Theor. Appl. Fract. Mech. 2020, 105, 102409. [Google Scholar] [CrossRef]
  10. Goanță, V.; Morăraș, C.; Blanari, I. Device for testing at combined cyclic fatigue loading by tensile and torsion, attachable to the universal testing machine. Exp. Tech. 2022, 46, 179–186. [Google Scholar] [CrossRef]
  11. Samiezadeh, S.; Schemitsch, E.H.; Zdero, R.; Bougherara, H. Biomechanical response under stress-controlled tension-tension fatigue of a novel carbon fiber/epoxy intramedullary nail for femur fractures. Med. Eng. Phys. 2020, 80, 26–32. [Google Scholar] [CrossRef] [PubMed]
  12. Chen, G.; Wu, H.; Gao, J.; Lin, Q. Development of a high-frequency and large-stroke fatigue testing system for rubber. Rev. Sci. Instrum. 2017, 88, 045113. [Google Scholar] [CrossRef] [PubMed]
  13. Duarte, L.M.T.; Almaraz, G.M.D.; Pacheco, C.J.T.; Tello, I.F.Z. Device to measure fatigue life of membranes under biaxial loading fatigue: Tension and torsion. Measurement 2020, 152, 107382. [Google Scholar] [CrossRef]
  14. Kong, B.S.; Shin, J.H.; Jang, C.; Kim, H.C. Measurement of fracture toughness of pure tungsten using a small-sized compact tension specimen. Materials 2020, 13, 244. [Google Scholar] [CrossRef]
  15. Le, X. A Probabilistic Fatigue Damage Model for Aluminum 6061-T6 10-Gauge Sheet-Type Material. ASCE-ASME J. Risk Uncertain. Eng. Syst. Part B Mech. Eng. 2021, 7, 034501. [Google Scholar] [CrossRef]
  16. Marques, J.M.E.; Benasciutti, D.; Niesłony, A.; Slavič, J. An overview of fatigue testing systems for metals under uniaxial and multiaxial random loadings. Metals 2021, 11, 447. [Google Scholar] [CrossRef]
  17. Zeng, X.; Wu, Y.; Yang, Z.; Shen, Y.; Ma, L. Research on high-frequency small-load fatigue testing device driven by piezoelectric actuator. Rev. Sci. Instrum. 2019, 90, 076102. [Google Scholar] [CrossRef] [PubMed]
  18. Deng, J.; Liu, Y.X.; Zhang, S.J.; Liu, J. Modeling and experiments of a nano-positioning and high frequency scanning piezoelectric platform based on function module actuator. Sci. China Technol. Sci. 2020, 63, 2541–2552. [Google Scholar] [CrossRef]
  19. Costa, P.; Nwawe, R.; Soares, H.; Reis, L.; Freitas, M.; Chen, Y.; Montalvão, D. Review of multiaxial testing for very high cycle fatigue: From ‘conventional’ to ultrasonic machines. Machines 2020, 8, 25. [Google Scholar] [CrossRef]
  20. Saghaeian, F.; Lederer, M.; Hofer, A.; Todt, J.; Keckes, J.; Khatibi, G. Investigation of high cyclic fatigue behaviour of thin copper films using MEMS structure. Int. J. Fatigue 2019, 128, 105179. [Google Scholar] [CrossRef]
  21. Ma, Z.; Zhao, H.; Li, Q.; Wang, K.; Zhou, X.; Hu, X.; Cheng, H.; Lu, S. Novel in situ device for investigating the tensile and fatigue behaviors of bulk materials. Rev. Sci. Instrum. 2013, 84, 045104. [Google Scholar] [CrossRef] [PubMed]
  22. Andreades, C.; Meo, M.; Ciampa, F. Tensile and fatigue testing of impacted smart CFRP composites with embedded PZT transducers for nonlinear ultrasonic monitoring of damage evolution. Smart Mater. Struct. 2020, 29, 055034. [Google Scholar] [CrossRef]
  23. Li, K. Design and Research of High Frequency Horizontal Small Load Fatigue Testing Machine Drived by Piezoelectric Actuator; Jilin University: Changchun, China, 2017. [Google Scholar]
  24. Ma, L. Research on Piezoelectric High-Frequency Micro Load Fatigue Testing Machine; Jilin Uinversity: Changchun, China, 2018. [Google Scholar]
  25. Ogawa, F.; Shimizu, Y.; Bressan, S.; Morishita, T.; Itoh, T. Bending and torsion fatigue-testing machine developed for multiaxial non-proportional loading. Metals 2019, 9, 1115. [Google Scholar] [CrossRef]
  26. Xuan, J.; Wang, S. Development of hydraulically driven fatigue testing machine for insulators. IEEE Access 2017, 6, 980–988. [Google Scholar] [CrossRef]
  27. Xu, P.; Liu, H.Y.; Men, Y.T.; Xu, B.S.; Lu, J.; Zhang, C.Q. A new type fatigue machine design with high frequency large stroke. Appl. Mech. Mater. 2014, 496, 1522–1525. [Google Scholar] [CrossRef]
Figure 1. Structure of fatigue testing machine. (a) Fatigue testing machine system structure diagram (200 mm × 45 mm × 600 mm); (b) Three-dimensional schematic of a piezoelectric vibrator; (c) Three-dimensional structure of dynamic load adjustment mechanism; (d) Three-dimensional structure of piezoelectric actuator.
Figure 1. Structure of fatigue testing machine. (a) Fatigue testing machine system structure diagram (200 mm × 45 mm × 600 mm); (b) Three-dimensional schematic of a piezoelectric vibrator; (c) Three-dimensional structure of dynamic load adjustment mechanism; (d) Three-dimensional structure of piezoelectric actuator.
Applsci 14 06961 g001aApplsci 14 06961 g001b
Figure 2. Fatigue testing machine force detection system.
Figure 2. Fatigue testing machine force detection system.
Applsci 14 06961 g002
Figure 3. Acquisition Software Interface.
Figure 3. Acquisition Software Interface.
Applsci 14 06961 g003
Figure 4. Cutaway view of piezoelectric vibrator.
Figure 4. Cutaway view of piezoelectric vibrator.
Applsci 14 06961 g004
Figure 5. Cutaway view of a mass block.
Figure 5. Cutaway view of a mass block.
Applsci 14 06961 g005
Figure 6. Cutaway view of an elastic ring.
Figure 6. Cutaway view of an elastic ring.
Applsci 14 06961 g006
Figure 7. Specimen size.
Figure 7. Specimen size.
Applsci 14 06961 g007
Figure 8. Model of spring–mass–damping system for fatigue testing machine.
Figure 8. Model of spring–mass–damping system for fatigue testing machine.
Applsci 14 06961 g008
Figure 9. Intrinsic frequency of the system for mass.
Figure 9. Intrinsic frequency of the system for mass.
Applsci 14 06961 g009
Figure 10. Plot of mass block and elastic ring stiffness versus intrinsic frequency of fatigue tester.
Figure 10. Plot of mass block and elastic ring stiffness versus intrinsic frequency of fatigue tester.
Applsci 14 06961 g010
Figure 11. Cloud diagram of the first twenty orders of modal analysis of the fatigue testing machine.
Figure 11. Cloud diagram of the first twenty orders of modal analysis of the fatigue testing machine.
Applsci 14 06961 g011aApplsci 14 06961 g011b
Figure 12. Finite element modal analysis of fatigue testing machine.
Figure 12. Finite element modal analysis of fatigue testing machine.
Applsci 14 06961 g012
Figure 13. System intrinsic frequency versus thickness of counterweight mass block a, thickness of elastic ring at flexibility b.
Figure 13. System intrinsic frequency versus thickness of counterweight mass block a, thickness of elastic ring at flexibility b.
Applsci 14 06961 g013
Figure 14. Fatigue testing machine test system.
Figure 14. Fatigue testing machine test system.
Applsci 14 06961 g014
Figure 15. Comparison of simulated, theoretical, and experimental effects of structural parameter on the inherent frequency of the system. (a) Tb = 0.4 mm. (b) Tb = 0.5 mm. (c) Tb = 0.6 mm. (d) Tb = 0.7 mm.
Figure 15. Comparison of simulated, theoretical, and experimental effects of structural parameter on the inherent frequency of the system. (a) Tb = 0.4 mm. (b) Tb = 0.5 mm. (c) Tb = 0.6 mm. (d) Tb = 0.7 mm.
Applsci 14 06961 g015
Figure 16. Stress amplitudes acting on the specimen with different counterweight mass blocks as a function of voltage and specimen stiffness. (a) a = 0 mm. (b) a = 10 mm. (c) a = 20 mm. (d) a = 30 mm.
Figure 16. Stress amplitudes acting on the specimen with different counterweight mass blocks as a function of voltage and specimen stiffness. (a) a = 0 mm. (b) a = 10 mm. (c) a = 20 mm. (d) a = 30 mm.
Applsci 14 06961 g016
Figure 17. Stress amplitudes acting on the specimen as a function of voltage and preset load for different counterweight mass blocks. (a) a = 0 mm. (b) a = 10 mm. (c) a = 20 mm. (d) a = 30 mm.
Figure 17. Stress amplitudes acting on the specimen as a function of voltage and preset load for different counterweight mass blocks. (a) a = 0 mm. (b) a = 10 mm. (c) a = 20 mm. (d) a = 30 mm.
Applsci 14 06961 g017
Figure 18. Tensile test diagram. (a) Tensile test I. (b) Tensile test II. (c) Tensile test III.
Figure 18. Tensile test diagram. (a) Tensile test I. (b) Tensile test II. (c) Tensile test III.
Applsci 14 06961 g018aApplsci 14 06961 g018b
Figure 19. Complete fatigue test load graphs.
Figure 19. Complete fatigue test load graphs.
Applsci 14 06961 g019
Table 1. Material parameters of piezoelectric vibrator.
Table 1. Material parameters of piezoelectric vibrator.
MaterialsDensities (kg·m)Modulus of
Elasticity
(Pa)
Poisson’s RatioThicknesses
(mm)
PZT-87600Tc
Spring steel78102.1 × 10110.288Tm
Table 2. The three main matrices of the PZT-8.
Table 2. The three main matrices of the PZT-8.
Dielectric   Coefficient   Matrix   [ ε s ] Piezoelectric   Matrix   [ e ] / C · m 2 Matrix   of   Elasticity   Coefficients   [ c ] / 10 10 N · m 2
x             y             z     x y z 900 0 0 0 900 0 0 0 600 x             y               z     x y z x y y z x z 0 0 4.1 0 0 4.1 0 0 14.0 0 0 0 0 10.3 0 10.3 0 0       x           y           z           x y           y z               x z     x y z x y y z x z 14.9 8.11 8.11 0 0 0 8.11 14.9 8.11 0 0 0 8.11 8.11 13.2 0 0 0 0 0 0 3.4 0 0 0 0 0 0 3.13 0 0 0 0 0 0 3.13
Table 3. Chemical composition of Q235b.
Table 3. Chemical composition of Q235b.
Chemical Composition of Dog-Bone Specimen (Q235b) (wt%)
CSiMnSP
0.120.030.45≤0.0045≤0.0045
Table 4. Mechanical properties of Q235b.
Table 4. Mechanical properties of Q235b.
Yield Strength (MPa)Tensile Strength (MPa)Elongation min (%)Density (Kg/m3)
235375~460267850
Table 5. List of discrete chaos models.
Table 5. List of discrete chaos models.
Mass UnitMass Unit SubunitStructural Schematic
M1M1.1 Force sensorApplsci 14 06961 i001
M1.2 Force sensor substrates
M1.3 Transfer plate
M1.4 Upper fixture
M2M2.1 SpecimensApplsci 14 06961 i002
M2.2 Bottom fixture
M2.3 Lower fixed support plate
M2.4 Vibration plate
M2.5 Elastic ring
M3M3.1 Connection shaftApplsci 14 06961 i003
M3.2 Mass block
M3.3 Circling overhead
M3.4 Piezoelectric vibrator
M3.5 Sump
Table 6. Detailed list of relevant parameters of the fatigue testing machine.
Table 6. Detailed list of relevant parameters of the fatigue testing machine.
Mass UnitParameter Value/kgRigidity Parameter Parameter   Value / ( × 10 6 ) ( N / m ) Damping ParametersParameter Value
M 1 0.2954 k 1 3.33 c 1 16.8568
k s ( t = 0 ) 5.43 c s ( t ) 15.0000
M 2 0.3159 k 2 0.1: 0.1: 1 c 2 17.2519
M 3 0.9990 + M ( 0 : 0.2 : 3 ) k 3 6.21 c 3 20.8610
Table 7. Main structural dimensions of each part of the fatigue testing machine parts.
Table 7. Main structural dimensions of each part of the fatigue testing machine parts.
Part NumberPart NameDimensional ParametersParameter Value (mm)
1FootingCalibre ϕ 30
High degree30
2FoundationLength250
Width200
Thickness45
Centre distance between two holes in the middle130
3Guide barCentre Diameter ϕ 20
Diameter on both sides ϕ 16
High degree488
4Lower support plateLength200
Width35
Thickness25
5Load boardLength200
Width35
Thickness25
6BackboneLength200
Width35
Thickness20
7ScrewLarge diameterM10
8Load handleOuter diameter ϕ 8
9Sensor fixing plateLength60
Width30
Thickness15
10TransducersMiddle diameter ϕ 20
Diameter on both sides ϕ 16
High degree22
11Transducer lower connection plateLength35
Width30
Thickness7
12Fixture fixing plateLength35
Width30
Thickness15
13FixtureWidth15
Specimen clamping range≤4
14SpecimenLength55
Width on both sides10
Middle width0.8
Thickness0.5
15Vibration plateLength53
Width30
Thickness on both sides7
Thickness in the middle13
16Elastic ringLength76
Width10
Thickness15
Flex thickness0.4/0.5/0.6/0.7
17Telegraph rodLength94
Calibre8
Head diameter5
18Mass of counterweightsOuter diameter ϕ 90
Inside diameter ϕ 10
Thickness10/20/30
19CounterweightOuter diameter ϕ 90
Inside diameter ϕ 70
Thickness on all sides13
Thickness in the middle8
20Piezoelectric oscillatorOuter diameter of substrate ϕ 90
Substrate inner diameter ϕ 6
Substrate thickness2
Outer diameter of ceramic disc ϕ 60
Ceramic disc inner diameter ϕ 10
Thickness of ceramic disc1
21Circling in midstreamOuter diameter ϕ 90
Inside diameter ϕ 70
Thickness on all sides10
Thickness at centre5
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content.

Share and Cite

MDPI and ACS Style

Zhang, S.; Yang, Z.; Liu, J. Design and Experimental Study of a Down-Drive Piezoelectric High-Frequency Fatigue Testing Machine. Appl. Sci. 2024, 14, 6961. https://doi.org/10.3390/app14166961

AMA Style

Zhang S, Yang Z, Liu J. Design and Experimental Study of a Down-Drive Piezoelectric High-Frequency Fatigue Testing Machine. Applied Sciences. 2024; 14(16):6961. https://doi.org/10.3390/app14166961

Chicago/Turabian Style

Zhang, Sida, Zhigang Yang, and Jiangwei Liu. 2024. "Design and Experimental Study of a Down-Drive Piezoelectric High-Frequency Fatigue Testing Machine" Applied Sciences 14, no. 16: 6961. https://doi.org/10.3390/app14166961

Note that from the first issue of 2016, this journal uses article numbers instead of page numbers. See further details here.

Article Metrics

Back to TopTop