1. Introduction
In the present era of rapid methods (including molecular and serological ones), the method involving culture and isolation is still recommended by the current regulations for the microbiological analysis of food, medical products, and so on, in order to assess key characteristics of the micro-organisms, such as their antibiotic microbial sensitivities [
1,
2,
3]. One key element is the sterilization of culture media and related materials to ensure the isolation of micro-organisms. However, autoclaving, as a sterilization method, is associated with certain drawbacks, such as high energy and equipment costs, long cycles, and the occupancy of considerable laboratory space, making it potentially unaffordable for laboratories around the world. Moreover, it is known that common autoclaving conditions can degrade key nutrients (such as peptones, agar, yeast extracts, and so on) and generate toxic compounds to which micro-organisms are sensitive [
4]. The generation of reactive oxygen species (ROS) or products of Maillard reactions (such as melanoidins) has been reported as compromising the culturability of certain sensitive species [
1,
5,
6,
7,
8].
A reduction in the autoclaving conditions, or the separation of agar from the rest of the components during autoclaving [
6], have been demonstrated to increase the productivity of different media, particularly for key sensitive species.
There is also interest in the use of alternatives to the autoclave for sterilizing culture media, which could maintain the properties of the medium and even increase microbial growth and selectivity [
6,
8]. Alternative sterilization methods could include filtration, ionizing radiation, or microwave sterilization. However, filtration is not a viable option for large amounts of media and cannot eliminate viruses, while ionizing radiation (e.g., gamma radiation) is not a feasible and economical option for most laboratories [
9,
10], and is typically reserved for key culture media or ingredients.
Microwave technology has been gaining increasing interest as a potential alternative method for the sterilization of culture media, as it can produce high-quality media, while saving significant time and energy, and is considerably more affordable than autoclaves [
11,
12,
13].
It is known that microwave irradiation can destroy micro-organisms though thermal and non-thermal effects [
14,
15,
16,
17,
18]. However, in the current routine by microbiological laboratories, its use is limited to melting certain solid media or to regenerate media intended for anaerobic growth, such as sulfite polymyxin sulfadiazine (SPS). To date, however, few studies have explored the utility of microwaves to prepare culture media from dehydrated ingredients in comparison with autoclaves. In a previous study, the use of microwaves was demonstrated to be an efficient alternative for the preparation of Sabouraud dextrose agar (SDA) [
7]. However, the additional devices required to ensure safety (for gas elimination) and to maintain the tightness of the media container during the microwave process are issues that have not yet properly been assessed. In normal conditions, common bottles do not resist the generated pressure and, due to the loosening of normal caps, leaks of the medium from the container are common.
In the present study, the use of a microwave was compared to autoclave sterilization for the production of common microbiological culture media, including tryptone soy agar (TSA), tryptone soy broth (TSB), Sabouraud 4% dextrose agar (SDA), Sabouraud dextrose broth, and violet red bile glucose agar (VRBG), through the implementation of an innovative device based on the use of a water bath made out of polycarbonate, and an easily perforated rubber cap, allowing medium depressurization. A commonly used power level (1000 W) and different exposure times (min) were assessed.
2. Materials and Methods
Manufacturing of culture media
All media (TSA, TSB, SDA, Sabouraud dextrose broth, and VRBG) were manufactured at Reactivos para Diagnóstico (RPD), S.L. (Sentmenat, Barcelona, Spain) from dehydrated ingredients, which were mixed in 250 mL of deionized water and resuspended, and sterilized using the following: (1) moist heat (121 °C for 15 min) in an autoclave (AES-150 model, Raypa, Barcelona, Spain) or by boiling (VRBG); (2) a water bath in a microwave, in a glass bottle with a special rubber cap, specially designed for this purpose (
Figure 1). The cap was punctured with a 28 mm sterile needle, and capped with a sticker to allow gas elimination and to prevent contamination. For the microwave method, the bottle with the rehydrated medium was placed in a polycarbonate tube (14 cm high and 8 cm in diameter) filled with water (
Figure 1), covering the whole part of the filled bottle, similar to a water bath. This tube allowed a homogeneous temperature distribution and a safety barrier to be created. The cycles were programmed, with different durations (from 3 to 20 min), at 1000 W, in a conventional microwave oven (Cecotec ProClean 5110 Inox, Valencia, España).
The only difference in the resulting culture media was the sterilization method. The reference media were sterilized using common autoclaving conditions, as specified by the manufacturer. The TSA and SDA were autoclaved at 121 °C for 15 min, and the VRBG was boiled for 1 min until completely dissolved. Using the microwave method, all culture media were sterilized in a microwave oven, as mentioned above. Solid culture media were maintained in bottles at room temperature, until being dispensed onto Petri plates (with the bacterial inoculum, using the pour plate method or without it using spiral inoculation).
All culture media were pH adjusted, if necessary (to pH values of 5.6 ± 0.2 for SDA; 7.3 ± 0.2 for TSA, and 7.4 ± 0.2 for VRBG), to achieve the physicochemical properties required for each medium.
Micro-organism strains used
The microbial strains used in the productivity and selectivity assays were obtained from a reference strain collection (American Type Culture Collection, ATCC). The reference strains were sub-cultured once to obtain stock strains, from which the stock and working cultures were prepared. The strains used (
Table 1) were those recommended in ISO 11133:2014/amended 1:2018 for the quality control of TSA, TSB, SDA, Sabouraud dextrose broth, and VRBG [
19].
Inocula preparation
The stock culture strains were maintained in either TSA (bacteria) or SDA (molds and yeasts) slant tubes (RPD S.L.) under refrigeration. The inocula of the studied culture were prepared according to ISO 11133 [
19]. Each strain was verified before use in non-selective media (TSA for bacteria, and SDA for molds and yeasts) and then cultured in appropriately selected and differential media, according to ISO 17025:2017 (ISO 17025) [
20]. Stock dilutions were prepared in maximum recovery diluent (RPD S.L.) using turbidimetry (around 0.5 McFarland units) (DEN-1B densitometer, Grant Instruments, Cambridgeshire, UK), from which serial dilutions were prepared using the same diluent. For the productivity assays, serial dilutions were prepared to inoculate 50–120 CFU per plate. For the selectivity assays (
S. aureus and
E. faecalis species), serial dilutions were prepared to inoculate ≥10
4 CFU per plate, according to ISO 11133:2014/amended 1:2018 [
19].
Productivity assays
Productivity assays were performed using the pour plate method, according to the corresponding ISO (ISO 4833-1:2014 [
21]; ISO 21149:2017 [
22]; ISO 16212:217 [
23]; ISO 21528-2:2004 [
24]) and European Pharmacopoeia (Ph. Eur., 10th ed.) [
3], with the culture media required for each strain (
Table 1). A microbial suspension (1 mL) containing 50–120 CFU (with a precision limit of ±20 CFU) was plated on sterile Petri dishes, and then the culture medium was poured onto the Petri dish and mixed with the inoculum.
In the productivity assays, to ensure the exact CFU were inoculated by the pour plate method in each Petri dish, a microbial suspension containing 50–120 CFU (100 µL) was also inoculated in the TSA or SDA, using the spiral plate method (Eddyjet, IULmicro, Barcelona, Spain), as a control.
Selectivity assays
Selectivity assays for VRBG were performed using the pour plate technique, according to ISO 6579 [
25], ISO 21528-2:2017 [
24], and ISO 11133:2014/amended 1: 2018 [
19], with
S. aureus (ATCC, equivalent to WDCM 00032) and
E. faecalis (ATCC 19433, equivalent to WDCM 00009), inoculated at ≥10
4 CFU (1 mL).
Seeding method
Two different assays, quantitative (solid media) and qualitative (liquid media), were performed to compare the quality of the media sterilized using both methods with different micro-organisms (
Table 1). In the quantitative assays, samples of 1 mL were plated in triplicate onto sterile Petri dishes, using the pour plate method. Then, the medium (20–25 mL), after melting in a boiling bath, was added once it was cooled to 40–45 °C, in both sterilization methods. In the qualitative assays, 1 mL (50–120 CFU) was inoculated in triplicate in broth bottles (TSB and Sabouraud dextrose broth, 250 mL) and then incubated for the required time and temperature.
The TSA plates and TSB bottles were incubated at 32.5 ± 2.5 °C for 24–48 h, the SDA plates and Sabouraud dextrose broth at 22.5 ± 2.5 °C for 2–5 days, and the VRBG plates and broth at 37 ± 2.5 °C for 24–48 h. For each medium and strain, inoculation was repeated in triplicate.
To provide a quality control for the qualitative assays (broth), 50 μL of each strain was inoculated in a sterile TSB bottle, sterilized in an autoclave, and incubated at 30 ± 1 °C for 24–48 h.
Colony count
The colony count was performed either manually or using an automatic colony counter (SphereFlash®, IULmicro, Barcelona, Spain).
Physicochemical properties
To check that each culture medium had the required pH values, the pH was measured with a pH meter (Crison GLP21, Barcelona, Spain), before and after sterilization (before pouring onto a sterile Petri dish). The pH was measured at different times after sterilization in a microwave oven (7 min, 10 min, 13 min, and 15 min) and after autoclaving sterilization (control).
Assessment of temperatures achieved in the microwave during sterilization
The temperature of the glass bottle was measured with an infrared thermometer (Helect, Shenzhen, China) every 10 s during the sterilization cycle, as an indicative measure of the degrees reached inside the microwave oven.
Sterility test
In order to determine the necessary duration at which the culture medium was completely sterile in the microwave oven equipped with a water bath, 250 mL bottles of TSA were produced, and several sterilization durations (3 min, 5 min, and 7 min) were tested at 1000 W, without inoculating the micro-organisms in the medium. Dehydrated culture medium was mixed with 250 mL of deionized water, after which the bottle was closed with the special cap and sticker (
Figure 1) and placed in a water bath in a microwave oven for the mentioned durations. The resulting plates were incubated at 32.5 ± 2.5 °C for 24–48 h.
With the media prepared for the optimal exposure time to obtain sterile plates, a quality control of the sterility test was performed, according to ISO 11333 (ISO 11133:2014/amended 1:2018) [
19], keeping the plates and broth for 7 days, as follows: the TSA plates and TSB bottles were incubated at 32.5 ± 2.5 °C, the SDA plates and Sabouraud dextrose broth at 22.5 ± 2.5 °C, and the VRBG plates at 37 ± 2.5 °C.
Kinetics of microbial destruction in the microwave oven
The bacterial strain Escherichia coli ATCC 8739 was obtained from slant tubes (RPD S.L.). Spores from Bacillus subtilis ATCC 6633 were obtained after a thermal shock at 80 °C for 10 min after 7-day cultures (maintained at 32.5 ± 2.5°C) and Geobacillus stearothermophilus ATCC 7953 spores were taken from ampoules (Sterikon® plus bioindicator, MERCK, Rahway, NJ, USA). Then, 50 μL of suspension containing 106 CFU/mL bacteria or spores was injected into a test bottle (5 × 104/bottle), which was exposed to a microwave at a power of 1000 W for periods of time ranging from 2.5 min to 20 min (depending on the destruction of each strain), with measurements taken every 30 s. The suspension was then cooled and poured into a sterile Petri dish and left to solidify. After incubation at 32.5 ± 2.5 °C for 48 h (E. coli ATCC 8739 and B. subtilis ATCC 6633) and 56 ± 2.5 °C (G. stearothermophilus ATCC 7953), the number of colonies was obtained. These tests were performed in triplicate.
Statistical Analysis
The data were analyzed using R software version 4.3.1. Generalized linear mixed models were used to model the bacterial counts (CFU), in which the linear predictor contains random effects, as well as fixed effects. Colonies (response) were modeled using Poisson’s distribution, and in the case of overdispersion, the negative binomial distribution model was used. The different media were compared pairwise through multiple comparisons, using the Tukey method.
The pH values of the different media, both reference and modified, were compared using the ANOVA test. The different media were compared pairwise, with a post hoc analysis using the Tukey test. A p-value less than 0.05 was considered statistically significant.
4. Discussion
The sterilization of microbiological culture media using a conventional microwave could become a simple and affordable method for most laboratories. However, when establishing a new protocol for the sterilization of culture media, several parameters must be considered, particularly exposure times and safety issues, as microwaves are not specifically designed for this purpose. Most of the studies carried out to study the sterilization capacity of microwave ovens have been performed on food samples, for which the maintenance of sterility is not necessary [
26,
27], or on surgical materials [
28]. Previous studies have described the effects on micro-organisms (either live or in their sporulated form) when they are sterilized in a microwave oven, being thermal and non-thermal [
7,
11,
12]. Only the study performed by Bazana et al. assessed the utility of a microwave for the preparation of SDA, adding a UV treatment during the gelling process [
7].
In our study, for the first time, an innovative and simple system for the preparation of culture media was developed, based on the use of a water bath made from polycarbonate covering the rehydrated media, in order to ensure the homogeneous distribution of the temperature and to act as safety containment to avoid any possible event related to high temperatures and possible container breakage.
To evaluate the new sterilization system, we assessed the microbiological properties of media prepared in a microwave (for different exposure times) in comparison with those of autoclaved media.
We started by first establishing the minimum power and time required for the complete sterilization of the culture medium (TSA), while measuring the temperature of the culture medium as it was exposed to the microwave oven.
We established that an exposure duration of 7 min at a power of 1000 W was sufficient to sterilize the culture medium. Based on the improvement observed in the growth of an important number of species in media exposed to a microwave, in comparison to the same autoclaved media, we can suggest that these conditions are softer than the autoclaving conditions and, thus, less degradation of nutrients and less generation of toxic compounds, such as reactive oxygen species (ROS), would occur under this microwave treatment.
Due to contamination risk and the bottle design, temperature measurement was only possible on the glass bottle, being an indirect measure of the temperature reached in the culture medium. Glass was the selected material for the bottle as it maintains the temperature of the culture medium during the sterilization and resists pressure increases.
Regarding the temperature measurement, a sustained exponential increase was observed over time. At the minimum sterilization time (7 min), the temperature of the glass bottle was around 90 °C, thus indicating a higher temperature of the culture media at this time.
When comparing the temperature curves of the autoclave and microwave oven, a higher temperature was reached in both at 15–16 min [
29]. However, the whole sterilization process was drastically reduced in the microwave method as, after 7 min, sterilization was achieved and the process of gas elimination was quicker, whereas the whole process in the autoclave, including the temperature increase, sterilization (commonly 21 min), and pressure decrease, may take up to 1.5 h [
29]. Moreover, in the autoclave, higher pressures are likely to be reached than in the microwave system, due to the larger size of the autoclave and the autoclaving conditions. This is the most notable difference between the two devices, making sterilization in a microwave oven a faster and safer process, without the inconvenience of needing to wait for the pressure to decrease. Further studies are needed to assess the pressure and temperature variations inside the microwave system, including the culture medium, similar to the existing measurements of these parameters in autoclaves [
29,
30,
31].
A comparison of media productivity using both systems was performed with culture media (solids and liquids) widely used in laboratories for the microbiological quality analysis of food, cosmetics, and pharmaceutical products. In addition, the media exposed for the different exposure times in the microwave were assessed, with an overall view of how the sterilization method and sterilization times affected the microbiological properties of the culture media.
The tendency of yeasts in an SDA medium to grow better in a microwave-sterilized medium than in an autoclaved medium could be attributed to the generation of toxic compounds; for example, due to Maillard reactions, the culture media darkens, being produced after the long exposure of the medium to high temperatures [
4,
5,
12], which can inhibit the growth of certain micro-organisms [
7,
32,
33]. Culture media containing simple sugars, such as SDA, are more likely to generate these Maillard reaction compounds than other culture media, and this process could be more pronounced in the autoclave than in the microwave system [
7]. For SDA, a correlation of medium browning with the exposure time in the microwave was observed. Exposures of 7–10 min rendered a clearer medium than the rest of the exposure times and the autoclave method. This could explain the increased growth, in certain cases, in media exposed to the microwave versus the autoclave.
For both yeasts (i.e.,
C. albicans and
S. cerevisiae), increased growth was observed with longer exposure times in the microwave oven, possibly due to their tolerance to pH variations [
34]. On the other hand, the growth of
A. brasiliensis in SDA was very similar in all conditions assayed. Growth in regard to SDA sterilized with this system deserves further research, as increased productivity could speed up detection in microbiological controls, food, and medicinal products that are characterized as requiring prolonged periods of incubation.
Improvements in the growth of some micro-organisms were also observed in the TSA sterilized in the microwave, particularly
S. aureus,
B. subtilis, and
L. monocytogenes, probably due to the lower generation of toxic compounds, such as ROS, the oxidation of certain compounds produced to a greater extent at higher temperatures, and the Maillard reaction between the agar and peptones [
6]. The increased growth of
S. aureus and
L. monocytogenes could be particularly relevant for detection in food and clinical samples, where the usefulness of the microwave system should be tested with respect to the corresponding selective media.
In the cases of P. aeruginosa, C. perfringens, and A. brasiliensis, lower growth was observed in the TSA sterilized in the microwave compared to the autoclaved media. TSA is not the optimum medium for the growth of A. brasiliensis and, as reported above, growth in regard to SDA prepared with the new system yielded similar growth to the autoclaved SDA. In the case of C. perfringens, in routine practice, selective culture media are used, such as SPS or tryptose sulfite cycloserine (TSC), which could also be assessed using this system.
In the case of
P. aeruginosa, the lower growth detected in the microwave media could also be an advantage for the manufacturing of certain selective media, in which the inhibition of
Pseudomonas often constitutes a challenge, as it concerns samples used for the analysis of water samples, where
Pseudomonas is often the main interfering micro-organism [
8].
Finally, for the selective media VRBG, which is commonly prepared by boiling, relevant growth increases were observed, particularly for
E. coli ATCC 25922,
S. ser. Typhimurium, and
P. aeruginosa. This could be attributed to the fact that, in the conventional method, it is necessary to wait until all the components are well dissolved in the medium, after which it is boiled. In contrast, in the novel microwave oven sterilization method, the medium is heated only during the microwave oven exposure time. Furthermore, it was observed that the exposure time with the best microbial growth results was 10 min for
S. ser. Typhimurium,
E. coli ATCC 25922, and
P. aeruginosa. In this medium, the darkening of the medium was also correlated with the exposure time in the microwave. In a previous study, the properties of this medium also improved with the separation of the components (agar vs. the rest of ingredients) during the autoclaving process [
6].
Using the microwave method, liquid media (TSB and Sabouraud dextrose broth) were also analyzed, as they are routinely used in microbiological quality laboratories. All media sterilized in the microwave oven performed in the same way as the media sterilized in the autoclave for all micro-organisms tested. As both media contain glucose, in future studies, it is necessary to quantify the microbial growth obtained, as a lower degradation of glucose could have occurred in the media prepared in the microwave, thus favoring microbial growth. Improvement of the enrichment media is relevant, for example, in the analysis of food. In the case of the detection of L. monocytogenes in food, at present, there is interest in finding alternatives to the Fraser medium. Identification of the best method for medium sterilization could also contribute to obtaining higher recovery rates.
Regarding pH values, we reported that SDA particularly increased its acidification after 15 min of exposure, likely due to the thermal degradation of glucose, which is in agreement with other previous studies [
7].
Regarding the study of microbial destruction using the microwave system, live bacteria (
E. coli) and bacterial endospores (from
B. subtilis and
G. stearothermophilus) were used. In the case of the spores of
G. stearothermophilus, added at 5 × 10
4 spores/bottle, complete destruction was not obtained at 7 min. It should be noted that these endospores are highly thermoresistant and used as a biological sterilization indicator [
35] during autoclaving cycles. However, they are not commonly found in dehydrated culture media and germinate at temperatures higher than 45 °C. For these reasons, their presence at a low concentration in dehydration media would not indicate interference during the most common microbiological analysis. However, more studies should be carried out to optimize their complete elimination using this system.
When the medium was contaminated with the endospores of B. subtilis ATCC 6633, the time needed to reach three decimal reductions was approximately 4.5 min, with a count of 0.33 ± 0.58 CFU remaining when the medium was sterilized in the microwave oven for 7 min. Considering that dehydrated media would not contain high concentrations of Bacillus endospores, these results should not represent an issue, although more studies should be performed to optimize this destruction.
Finally, as expected, complete destruction was obtained for
E. coli ATCC 8739 with an exposure time of 4 min, in agreement with previous studies [
17]. It should be noted that, according to our experience,
Pseudomonas could also be found in dehydrated media. Therefore, the inclusion of this bacteria in future microbial destruction studies would be of interest.
An important aspect of the microwave system is safety. The puncture of the rubber cap and the application of the sticker allows for depressurization, prevents medium contamination, and avoids medium leakage. To increase safety, other safety strategies could be added, such as the delayed opening of the door, allowing for depressurization before opening, and a polycarbonate cover for the water bath. Other safety issues to take into account include avoiding the introduction of metallic materials and certain types of plastic.
With the optimized microwave system, validation studies should be performed in the same way as autoclave sterilization validation studies to assess the efficiency of the new method, according to validation guidelines (ISO 11133:2014/amended 1:2018 (ISO 11133)) [
19]. This system also deserves to be included in the applicable regulations.
In summary, in comparison to classical autoclave sterilization, the work performed with a microwave offers different advantages, such as time saving, as well as immediacy, increased sterilization capacity, and better microbiological properties of certain media, likely due to the lower generation of toxic compounds. Other advantages can be foreseen, such as energy and cost savings, that require assessment in future work. This system could become a common sterilization method for most microbiological media, although more studies should be carried out to confirm this.