Next Article in Journal
Public Health Value of a Hypothetical Pneumococcal Conjugate Vaccine (PCV) Introduction: A Case Study
Next Article in Special Issue
Immunogenicity Evaluating of the Multivalent COVID-19 Inactivated Vaccine against the SARS-CoV-2 Variants
Previous Article in Journal
COVID-19 Vaccination Hesitancy among Healthcare Workers—A Review
Previous Article in Special Issue
Major Complex Trait for Early De Novo Programming ‘CoV-MAC-TED’ Detected in Human Nasal Epithelial Cells Infected by Two SARS-CoV-2 Variants Is Promising to Help in Designing Therapeutic Strategies
 
 
Article
Peer-Review Record

Methodology of Purification of Inactivated Cell-Culture-Grown SARS-CoV-2 Using Size-Exclusion Chromatography

Vaccines 2022, 10(6), 949; https://doi.org/10.3390/vaccines10060949
by Anastasia A. Kovpak 1,*, Anastasia N. Piniaeva 1, Oleg A. Gerasimov 1, Irina O. Tcelykh 1, Mayya Y. Ermakova 1, Anna N. Zyrina 1, Dmitry V. Danilov 1, Yury Y. Ivin 1, Liubov I. Kozlovskaya 1,2 and Aydar A. Ishmukhametov 1,2
Reviewer 1: Anonymous
Reviewer 2: Anonymous
Reviewer 3:
Vaccines 2022, 10(6), 949; https://doi.org/10.3390/vaccines10060949
Submission received: 25 April 2022 / Revised: 8 June 2022 / Accepted: 9 June 2022 / Published: 15 June 2022
(This article belongs to the Special Issue Advances in COVID-19 Vaccines Development)

Round 1

Reviewer 1 Report

The presented manuscript perfectly fits into the global trend of research on vaccines against SARS-CoV-2 virus. In the presented work, the authors check the suitability of of four size exclusion chromatography resins for purification of viral antigen for the human COVID-19 vaccine. The presented results and interesting and useful in terms of COVID-19 research. The manuscript can be accepted after minor revision.

  1. "2.1 Cells and Viruses": I think the Authors should resume isolation, passages and description very shortly. Otherwise reader has to check position [13] but in my opinion, it should be done only by someone really interested in copying/verifying the results. Typical read should have a fast view of the procedure in the paper.
  2. Table 1: when describing the Resin, please remove the company and country of origin. These should be placed directly in the text (2.3 Chromatography)
  3. Please improve the quality of the Figure 2. The colors should be uniformed and resolution and general quality improved.
  4. The Authors have to elaborate on the results in the conclusions: two sentences are not enough. Please compare your results with other techniques, what else can be done (any future experiments?)

Author Response

The presented manuscript perfectly fits into the global trend of research on vaccines against SARS-CoV-2 virus. In the presented work, the authors check the suitability of of four size exclusion chromatography resins for purification of viral antigen for the human COVID-19 vaccine. The presented results and interesting and useful in terms of COVID-19 research. The manuscript can be accepted after minor revision.

1. "2.1 Cells and Viruses": I think the Authors should resume isolation, passages and description very shortly. Otherwise reader has to check position [13] but in my opinion, it should be done only by someone really interested in copying/verifying the results. Typical read should have a fast view of the procedure in the paper.

Response: manuscript text was amended with a short description of the strain isolation, passage history and characterization.

2. Table 1: when describing the Resin, please remove the company and country of origin. These should be placed directly in the text (2.3 Chromatography)

Response: Corrected

3. Please improve the quality of the Figure 2. The colors should be uniformed and resolution and general quality improved.

Response: Corrected

4. The Authors have to elaborate on the results in the conclusions: two sentences are not enough. Please compare your results with other techniques, what else can be done (any future experiments?)

Response: Amended

 

Reviewer 2 Report

Please describe in more detail the conditions of size-exclusion chromatography.

In the introduction i suggest the authors to take some informations from this articles 

DOI 10.3390/microorganisms8111704, 

DOI

10.3390/microorganisms9040793, 

DOI

10.2147/RMHP.S284557, 

DOI

10.3390/antiox10060881

Could the authors please mention what are the differences from the other published article  mentioned in the bibliography?

There is no discussion chapter, in the results and discussion chapter the authors presented just their results and not aalso discussions.

Author Response

Please describe in more detail the conditions of size-exclusion chromatography.

Response: The data on SEC conditions is presented in Section 2.3. Text was amended.

 

In the introduction i suggest the authors to take some informations from this articles

DOI 10.3390/microorganisms8111704,

 DOI 10.3390/microorganisms9040793,

 DOI 10.2147/RMHP.S284557,

 DOI 10.3390/antiox10060881

Could the authors please mention what are the differences from the other published article mentioned in the bibliography?

Response: These manuscripts mainly describe the data and experience collected by clinical specialists during the peak of the COVID-19 pandemic. As our manuscript focuses on the technological process of inactivated coronavirus antigen purification, we decided not to overload the paper with clinical data on COVID-19 and included only main facts to put the research into context. The technology presented is not specific for SARS-CoV-2, and can be used for purification of other CoVs’ antigen.

 

There is no discussion chapter, in the results and discussion chapter the authors presented just their results and not aalso discussions.

Response: The Discussion was added and elaborated.

 

Reviewer 3 Report

we read with interest the article by Kovpak et al discussing the methodology of using size exclusion column chromatography for isolating and purifying the viral samples ( SARS-CoV-2). the article is written in a good way however; it has some major deficits that need to be answered.

First the article falls in the methodology category and thus the whole method section should be described in an analytical level in terms of amount of proteins loaded (mg) and describe fraction collection and recovery rate as well as purity to be assessed by Western blotting to show that the separation of size exclusion in the different fractions is achieving the size separation. These comments are related to the methodology section.

 

In terms of concepts, the authors do not describe what are they achieving specifically, they mention viral antigen, are they purifying the S protein and if so what is the molecular weight of the pure protein and what is the activity level of this protein after separation, all these data and experiments should be performed. the way it is described is vague and it is described as viral antigen?

In addition, the article describes a methodology of the size exclusion, the authors should state if there is a gap of knowledge in achieving and performing this method successfully and if there are challenges, if not, this report carries no scientific value.

 

The ELISA is described in a very simplistic method where the amount and time of incubation and antibodies concentrations and sources are not described also the results if these ELISA reading should be presented and the raw values should be presented as supplementary data.

 

The introduction should discuss previous work on protein separation using size exclusion and why the authors are presenting an article about it. 

 

 

 

Author Response

we read with interest the article by Kovpak et al discussing the methodology of using size exclusion column chromatography for isolating and purifying the viral samples ( SARS-CoV-2). the article is written in a good way however; it has some major deficits that need to be answered.

First the article falls in the methodology category and thus the whole method section should be described in an analytical level in terms of amount of proteins loaded (mg) and describe fraction collection and recovery rate as well as purity to be assessed by Western blotting to show that the separation of size exclusion in the different fractions is achieving the size separation. These comments are related to the methodology section.

Response: The present manuscript describes only the process of the viral antigen purification. The characteristics of the purified antigen itself (SDS-PAGE and WB) are presented in [13].

 

In terms of concepts, the authors do not describe what are they achieving specifically, they mention viral antigen, are they purifying the S protein and if so what is the molecular weight of the pure protein and what is the activity level of this protein after separation, all these data and experiments should be performed. the way it is described is vague and it is described as viral antigen?

Response: In this manuscript, we investigated the applicability of gel filtration sorbents for the purification of inactivated whole-virion coronavirus particles obtained via reproduction in Vero cells. Therefore, the antigen is not a single viral protein S, but a whole viral particle capable of inducing an immune response.

Response: Introduction part was amended to put the present work into context.

The ELISA is described in a very simplistic method where the amount and time of incubation and antibodies concentrations and sources are not described also the results if these ELISA reading should be presented and the raw values should be presented as supplementary data.

Response: Materials and Methods section was supplemented with detailed description of the ELISA procedure. The whole procedure is fairly basic.

The introduction should discuss previous work on protein separation using size exclusion and why the authors are presenting an article about it. 

Response: Introduction part was amended to put the present work into context.

Round 2

Reviewer 3 Report

Thank you for the clarification and I think the work would be much stronger if this work focused on the protein ID and the virology aspect of this work. In the present work it is a technical report of a methodology applying size exclusion separation.

I would request that the title be changed into:  Methodology of Purification of Inactivated Cell Culture Grown SARS-CoV-2 by Size-Exclusion Chromatography

Author Response

The title has been changed as recommended

Back to TopTop