Ultrasound Cyclo Plasty for Treatment of Surgery-Naïve Open-Angle Glaucoma Patients: A Prospective, Multicenter, 2-Year Follow-Up Trial
Abstract
:1. Introduction
2. Materials and Methods
2.1. UCP Procedure
2.2. Patient Assessments
2.2.1. Efficacy Parameters:
- Mean IOP (mmHg) at each follow-up visit compared to baseline IOP;
- Mean IOP variation compared to baseline (%) at each follow-up visit;
- Number of ocular hypotensive medications at each follow-up visit.
2.2.2. Success Rate:
- Complete success: IOP lowering >20% (<21 mmHg), and without supplemental glaucoma medication compared to baseline;
- Qualified success: 6 < IOP < 21 mmHg (IOP lowering < 20%) and reduction of glaucoma medications compared to baseline.
2.2.3. Safety Parameters:
- Rate of per-operative device and/or procedure-related adverse events;
- Rate of post-operative device and/or procedure-related complications and adverse events at each follow-up visit;
- Best corrected visual acuity (BCVA) scored with reference to Logarithm of the Minimum Angle of Resolution (LogMAR).
2.3. Statistical Analysis
3. Results
4. Discussion
5. Conclusions
Author Contributions
Funding
Institutional Review Board Statement
Informed Consent Statement
Acknowledgments
Conflicts of Interest
Manufacturer Name
References
- De Roetth, A., Jr. Cryosurgery for the treatment of glaucoma. Trans. Am. Ophthalmol. Soc. 1965, 63, 189–204. [Google Scholar] [CrossRef] [Green Version]
- Maus, M.; Katz, L.J. Choroidal detachment, flat anterior chamber, and hypotony as complications of neodymium: YAG laser cyclophotocoagulation. Ophthalmology 1990, 97, 69–72. [Google Scholar] [CrossRef]
- Uram, M. Ophthalmic laser microendoscopeciliary process ablation in the management of neovascular glaucoma. Ophthalmology 1992, 99, 1823–1828. [Google Scholar] [CrossRef]
- Kosoko, O.; Gaasterland, D.E.; Pollack, I.P.; Enger, C.L. Long-term outcome of initial ciliary ablation with contact diode laser transscleral cyclophotocoagulation for severe glaucoma. The Diode Laser Ciliary Ablation Study Group. Ophthalmology 1996, 103, 1294–1302. [Google Scholar] [CrossRef]
- Sabri, K.; Vernon, S.A. Scleral perforation following trans-scleral cyclodiode. Br. J. Ophthalmol. 1999, 83, 502–503. [Google Scholar] [CrossRef] [Green Version]
- Vernon, S.A.; Koppens, J.M.; Menon, G.J.; Negi, A.K. Diode laser cycloablation in adult glaucoma: Long-term results of a standard protocol and review of current literature. Clin. Exp. Ophthalmol. 2006, 34, 411–420. [Google Scholar] [CrossRef]
- Coleman, D.J.; Lizzi, F.L.; Driller, J.; Rosado, A.L.; Chang, S.; Iwamoto, T.; Rosenthal, D. Therapeutic ultrasound in the treatment of glaucoma. I. Experimental model. Ophthalmology 1985, 92, 339–346. [Google Scholar] [CrossRef]
- Coleman, D.J.; Lizzi, F.L.; Driller, J.; Rosado, A.L.; Burgess, S.E.P.; Torpey, J.H.; Smith, M.E.; Silverman, R.H.; Yablonski, M.E.; Chang, S.; et al. Therapeutic ultrasound in the treatment of glaucoma. II. Clinical applications. Ophthalmology 1985, 92, 347–353. [Google Scholar] [CrossRef]
- Coleman, D.J.; Lizzi, F.L.; Silverman, R.H.; Dennis, P.H., Jr.; Driller, J.; Rosado, A.; Iwamoto, T. Therapeutic ultrasound. Ultrasound Med. Biol. 1986, 12, 633–638. [Google Scholar] [CrossRef]
- Burgess, S.P.; Silverman, R.H.; Coleman, D.J.; Yablonski, M.E.; Lizzi, F.L.; Driller, J.; Rosado, A.; Dennis, P.H. Treatment of glaucoma with high-intensity focused ultrasound. Ophthalmology 1986, 93, 831–838. [Google Scholar] [CrossRef]
- Valtot, F.; Kopel, J.; Haut, J. Treatment of glaucoma with high intensity focused ultrasound. Int. Ophthalmol. 1989, 13, 167–170. [Google Scholar] [CrossRef]
- Maskin, S.L.; Mandell, A.I.; Smith, J.A.; Wood, R.C.; Terry, S.A. Therapeutic ultrasound for refractory glaucoma: A three-center study. Ophthalmic. Surg. 1989, 20, 186–192. [Google Scholar]
- Sterk, C.C.; vd Valk, P.H.; van Hees, C.L.; van Delft, J.L.; van Best, J.A.; Oosterhuis, J.A. The effect of therapeutic ultrasound on the average of multiple intraocular pressures throughout the day in therapy-resistant glaucoma. Graefes Arch. Clin. Exp. Ophthalmol. 1989, 227, 36–38. [Google Scholar] [CrossRef]
- Aptel, F.; Charrel, T.; Palazzi, X.; Chapelon, J.Y.; Denis, P.; Lafon, C. Histologic effects of a new device for high-intensity focused ultrasound cyclocoagulation. Investig. Ophthalmol. Vis. Sci. 2010, 51, 5092–5098. [Google Scholar] [CrossRef] [PubMed] [Green Version]
- Aptel, F.; Charrel, T.; Lafon, C.; Romano, F.; Chapelon, J.Y.; Nordmann, J.P.; Denis, P. Miniaturized high-intensity focused ultrasound device in patients with Glaucoma clinical pilot study. Investig. Ophthalmol. Vis. Sci. 2011, 52, 8747–8753. [Google Scholar] [CrossRef] [PubMed]
- Denis, P.; Aptel, F.; Rouland, J.F.; Nordmann, J.P.; Lachkar, Y.; Renard, J.P.; Sellem, E.; Baudouin, C.; Bron, A. Cyclocoagulation of the ciliary bodies by high intensity focused ultrasound: Result of a 12-month muticentric study in refractory glaucoma population. Investig. Ophthalmol. Vis. Sci. 2015, 56, 1089–1096. [Google Scholar] [CrossRef] [PubMed] [Green Version]
- Melamed, S.; Goldenfeld, M.; Cotlear, D.; Skaat, A.; Moroz, I. High-Intensity focused ultrasound treatment in refractory glaucoma patients. Results at 1 year of a prospective study. Eur. J. Ophthalmol. 2015, 25, 483–489. [Google Scholar] [CrossRef]
- Aptel, F.; Dupuy, C.; Rouland, J.F. Treatment of refractory Open-Angle Glaucoma using Ultrasonic Circular Cyclocoagulation: A prospective Case series. Curr. Med. Res. Opin. 2014, 38, 1599–1605. [Google Scholar] [CrossRef]
- Aptel, F.; Denis, P.; Rouland, J.F.; Renard, J.P.; Bron, A. Multicenter clinical trial of high-intensity focused ultrasound treatment in glaucoma patients without previous filtering surgery. Acta Ophthalmol. 2016, 94, 268–277. [Google Scholar] [CrossRef] [Green Version]
- Posarelli, C.; Covello, G.; Bendinelli, A.; Fogagnolo, P.; Nardi, M.; Figus, M. High-intensity focused ultrasound procedure: The rise of a new noninvasive glaucoma procedure and it’s possible future applications. Surv. Ophthalmol. 2019, 64, 826–834. [Google Scholar] [CrossRef] [PubMed]
- Morais Sarmento, T.; Figueiredo, R.; Garrido, J.; Passos, I.; Rebelo, A.L.; Candeias, A. Ultrasonic circular cyclocoagulation prospective safety and effectiveness study. Int. Ophthalmol. 2021, 41, 3047–3055. [Google Scholar] [CrossRef] [PubMed]
- Shaarawy, T.M.; Sherwood, M.B.; Grehn, F. (Eds.) Guidelines on Design and Reporting of Glaucoma Surgical Trials. In World Glaucoma Associations; Kugler Publications: Amsterdam, The Netherlands, 2009; pp. 15–24. [Google Scholar]
- Giannaccare, G.; Pellegrini, M.; Bernabei, F.; Urbini, L.; Bergamini, F.; Ferro Desideri, L.; Bagnis, A.; Biagini, F.; Cassottana, P.; Del Noce, C.; et al. A 2-year prospective multicenter study of ultrasound cyclo plasty for glaucoma. Sci. Rep. 2021, 11, 12647. [Google Scholar] [CrossRef] [PubMed]
- Torky, M.A.; Al Zafiri, Y.A.; Hagras, S.M.; Khattab, A.M.; Bassiouny, R.M.; Mokbel, T.H. Safety and efficacy of ultrasound ciliaryplasty as a primary intervention in glaucoma patients. Int. J. Ophthalmol. 2019, 12, 597–602. [Google Scholar]
- Leshno, A.; Rubinstein, Y.; Singer, R.; Sher, I.; Rotenstreich, Y.; Melamed, S.; Skaat, A.J. High-intensity Focused Ultrasound Treatment in Moderate Glaucoma Patients: Results of a 2-Year Prospective Clinical Trial. J. Glaucoma 2020, 29, 556–560. [Google Scholar] [CrossRef] [PubMed]
- Rouland, J.F.; Aptel, F. Efficacy and Safety of Ultrasound Cycloplasty for Refractory Glaucoma: A 3-Year Study. J. Glaucoma 2021, 30, 428–435. [Google Scholar] [CrossRef]
- De Gregorio, A.; Pedrotti, E.; Stevan, G.; Morselli, S. Safety and efficacy of multiple cyclocoagulation of ciliary bodies by high intensity focused ultrasound in patients with glaucoma. Graefes Arch. Clin. Exp. Ophthalmol. 2017, 255, 2429e35. [Google Scholar] [CrossRef] [PubMed]
- Aptel, F.; Tadjine, M.; Rouland, J.F. Efficacy and safety of repeated Ultrasound Cyclo Plasty procedures in patients with early or delayed failure after first procedure. J. Glaucoma 2020, 29, 24–30. [Google Scholar] [CrossRef] [PubMed]
- Bolek, B.; Wylęgala, A.; Wylęgala, E. Assessment of Scleral and Conjunctival Thickness of the Eye after Ultrasound Ciliary Plasty. J. Ophthalmol. 2020, 24, 9659014. [Google Scholar]
- Sousa, D.C.; Ferreira, N.P.; Marques-Neves, C.; Somer, A.; Vandewalle, E.; Stalmans, I.; Pinto, L.A. High-intensity focused ultrasound cycloplasty: Analysis of pupil dynamics. J. Curr. Glaucoma Pract. 2018, 12, 102–106. [Google Scholar] [CrossRef] [PubMed]
- Rivero Santana, A.; Perez-Silguero, D.; Perez Silguero, M.A.; Encinas-Pisa, P. Pupil ovalization and accommodation loss after High Intensity Focused Ultrasound treatment for glaucoma—A case report. J. Curr. Glaucoma Pract. 2019, 13, 77–78. [Google Scholar] [CrossRef] [PubMed]
- Morais Sarmento, T.; Figueiredo, R.; Garrido, J.; Rebelo, A.L. Transient choroidal detachment after ultrasonic circular cyclocoagulation. BMJ Case Rep. 2019, 12, 24–30. [Google Scholar] [CrossRef] [PubMed]
- Luo, Q.; Xue, W.; Wang, Y.; Chen, B.; Wang, S.; Dong, Y.; Ru, Y.; Ge, L. Ultrasound cyclo-plasty in Chinese glaucoma patients: Results of a 6-month prospective clinical study. Ophthalmic Res. 2021, 4. [Google Scholar] [CrossRef]
- Liu, H.T.; Zhang, Q.; Jiang, Z.X.; Xu, Y.X.; Wan, Q.Q.; Tao, L.M. Efficacy and safety of high-dose ultrasound cyclo-plasty procedure in refractory glaucoma. Int. J. Ophthalmol. 2020, 13, 1391–1396. [Google Scholar] [CrossRef] [PubMed]
- Torky, M.A.; Alzafiri, Y.A.; Abdelhameed, A.G.; Awad, E.A. Phaco-UCP; combined phacoemulsification and ultrasound ciliary plasty versus phacoemulsification alone for management of coexisting cataract and open angle glaucoma: A randomized clinical trial. BMC Ophthalmol. 2021, 21, 21–53. [Google Scholar] [CrossRef] [PubMed]
Eyes (Right/Left) | 66 (37/29) |
Age (years), Mean (SD) (range) | 70.4 ± 11.4 (42–90) |
Gender (male/female) | 32/34 (48%/52%) |
Lens status | |
Phakic | 37 (56%) |
Pseudophakic | 29 (44%) |
Type of glaucoma | |
Primary open-angle glaucoma | 54 (82%) |
Exfoliative glaucoma | 11 (16%) |
Pigmentary glaucoma | 1 (2%) |
Previous ocular treatments (total number of procedures) | |
Incisional surgery (trabeculectomy) | 0 |
Cyclodestructive procedure | 0 |
Laser trabeculoplasty (SLT/ALT) | 10 (15%) |
Mean preoperative IOP | 24.3 ± 2.9 [21,22,23,24,25,26,27,28,29,30] |
Pre-operative glaucoma medications | |
Eyedrops | 2.3 ± 1.1 |
Tablets (Acetazolamide) | 8/66 |
Overall Population (A) | Success at Month-24 (B) | ||||
---|---|---|---|---|---|
Mean IOP (No. Patients) (Eyedrops */Tablets **) | Relative IOP Reduction (%) | Success Rate (%) | Mean IOP (No. Patients) (Eyedrops */Tablets **) | Relative IOP Reduction (%) | |
Baseline | 24.3 ± 2.9 (66) (2.3 ± 1.1/8) | NA | NA | 24.0 ± 2.9 (42) (2.4 ± 1.3/4) | NA |
Day 1 | 13.6 ± 4,9 (65) (2.3 ± 1.1/9) | 44% | 94% | 12.9 ± 3,9 (41) (2.4 ± 1.3/4) | 46% |
Month 1 | 16.2 ± 5.1 (65) (2.2 ± 1.1/6) | 33% | 83% | 14.8 ± 4.3 (41) (2.3 ± 1.3/3) | 38% |
Month 3 | 16.0 ± 4.8 (66) (2.3 ± 1.1/5) | 34% | 83% | 14.5 ± 3.7 (42) (2.3 ± 1.4/3) | 40% |
Month 6 | 16.8 ± 5.2 (64) (2.2 ± 1.0/6) | 31% | 85% | 16.3 ± 4.8 (42) (2.1 ± 1.3/3) | 33% |
Month 12 | 16.5 ± 4.3 (60) (2.2 ± 1.0/3) | 32% | 80% | 16.0 ± 4.0 (42) (2.2 ± 1.2/2) | 34% |
Month 18 | 16.3 ± 4.4 (54) (2.1 ± 1.0/2) | 32% | 77% | 15.8 ± 4.4 (40) (2.1 ± 1.2/2) | 34% |
Month 24 | 15.9 ± 3.6 (50) (2.2 ± 1.0/3) | 33% | 77% | 15.4 ± 3.6 (42) (2.2 ± 1.2/2) | 36% |
Postoperative Complications | n (%) |
---|---|
Anterior chamber inflammation (<1 month) | 41 (62%) |
Conjunctival hyperemia (<1 month) | 25 (38%) |
Transient mild mydriasis * | 13 (19%) |
Superficial punctate keratitis | 10 (15%) |
Pupil peak (pupil deformation) | 5 (7%) |
Transient hypotony (IOP < 6 mmHg) | 3 (5%) |
Ocular pain (<24 h) | 2 (3%) |
Macular edema | 2 (3%) |
Uveitis | 1 (2%) |
Subconjunctival hemorrhage | 1 (2%) |
Corneal edema | 1 (2%) |
Monocular double vision | 1 (2%) |
LogMar Visual Acuity | Baseline | Month 1 | Month 3 | Month 6 | Month 12 | Month 18 | Month 24 | |
---|---|---|---|---|---|---|---|---|
Overall population | n | 66 | 65 | 66 | 63 | 58 | 48 | 43 * |
Mean ± SD | 0.43 ± 0.81 | 0.50 ± 0.80 | 0.46 ± 0.81 | 0.48 ± 0.88 | 0.54 ± 0.93 | 0.57 ±1.00 | 0.40 ± 0.82 | |
Vision Group (A) ** | n | 60 | 59 | 60 | 57 | 52 | 43 | 40 |
Mean ± SD | 0.19 ± 0.24 | 0.27 ± 0.28 | 0.23 ± 0.28 | 0.21 ± 0.31 | 0.28 ± 0.51 | 0.30 ±0.63 | 0.23 ± 0.46 | |
Unchanged | - | 26 (44%) | 40 (67%) | 39 (69%) | 35 (67%) | 27 (63%) | 31 (78%) | |
Loss 1 line | - | 21 (36%) | 12 (20%) | 13 (23%) | 10 (19%) | 11 (26%) | 4 (10%) | |
Loss ≥ 2 lines | - | 12 (20%) | 8 (13%) | 5 (9%) | 7 (14%) | 5 (12%) | 5 (12%) | |
Low vision Group (B) *** | n | 6 | 6 | 6 | 6 | 6 | 5 | 3 |
Mean ± SD | 2.80 ± 0.59 | 2.80 ± 0.59 | 2.80 ± 0.59 | 3.02 ± 0.24 | 2.85 ±0.48 | 2.84 ±0.54 | 2.80 ± 0.75 | |
Unchanged | - | 6 (100%) | 6 (100%) | 4 (67%) | 5 (83%) | 4 (80%) | 2 (67%) | |
Loss 1 line | - | - | - | 1 (16.5%) | - | - | - | |
Loss ≥ 2 lines | - | - | - | 1 (16.5%) | 1 (17%) | 1 (20%) | 1 (33%) |
Publisher’s Note: MDPI stays neutral with regard to jurisdictional claims in published maps and institutional affiliations. |
© 2021 by the authors. Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (https://creativecommons.org/licenses/by/4.0/).
Share and Cite
Figus, M.; Posarelli, C.; Nardi, M.; Stalmans, I.; Vandewalle, E.; Melamed, S.; Skaat, A.; Leshno, A.; Sousa, D.C.; Pinto, L.A. Ultrasound Cyclo Plasty for Treatment of Surgery-Naïve Open-Angle Glaucoma Patients: A Prospective, Multicenter, 2-Year Follow-Up Trial. J. Clin. Med. 2021, 10, 4982. https://doi.org/10.3390/jcm10214982
Figus M, Posarelli C, Nardi M, Stalmans I, Vandewalle E, Melamed S, Skaat A, Leshno A, Sousa DC, Pinto LA. Ultrasound Cyclo Plasty for Treatment of Surgery-Naïve Open-Angle Glaucoma Patients: A Prospective, Multicenter, 2-Year Follow-Up Trial. Journal of Clinical Medicine. 2021; 10(21):4982. https://doi.org/10.3390/jcm10214982
Chicago/Turabian StyleFigus, Michele, Chiara Posarelli, Marco Nardi, Ingeborg Stalmans, Evelien Vandewalle, Shlomo Melamed, Alon Skaat, Ari Leshno, David Cordeiro Sousa, and Luis Abegão Pinto. 2021. "Ultrasound Cyclo Plasty for Treatment of Surgery-Naïve Open-Angle Glaucoma Patients: A Prospective, Multicenter, 2-Year Follow-Up Trial" Journal of Clinical Medicine 10, no. 21: 4982. https://doi.org/10.3390/jcm10214982