Next Article in Journal
LipiSensors: Exploiting Lipid Nanoemulsions to Fabricate Ionophore-Based Nanosensors
Next Article in Special Issue
Screen-Printed Sensor for Low-Cost Chloride Analysis in Sweat for Rapid Diagnosis and Monitoring of Cystic Fibrosis
Previous Article in Journal / Special Issue
Electrochemiluminescence Biosensors Using Screen-Printed Electrodes
 
 
Font Type:
Arial Georgia Verdana
Font Size:
Aa Aa Aa
Line Spacing:
Column Width:
Background:
Article

Exploiting Laser-Ablation ICP-MS for the Characterization of Salt-Derived Bismuth Films on Screen-Printed Electrodes: A Preliminary Investigation

1
Dipartimento di Scienze Teoriche ed Applicate, Università degli Studi dell’Insubria, Via Dunant, 3, 22100 Varese, Italy
2
Dipartimento di Scienza ed Alta Tecnologia, Università degli Studi dell’Insubria, Via Valleggio, 11, 21100 Como, Italy
*
Author to whom correspondence should be addressed.
Biosensors 2020, 10(9), 119; https://doi.org/10.3390/bios10090119
Submission received: 30 July 2020 / Revised: 31 August 2020 / Accepted: 8 September 2020 / Published: 9 September 2020

Abstract

:
The use of insoluble bismuth salts, typically BiPO4, is known to be a viable alternative to classical Bi3+ ion electrochemical reduction for the preparation of bismuth film electrodes (BiFE) on screen-printed electrodes. The freshly prepared electrodes are indefinitely stable, and the active bismuth film is simply formed by in situ reduction. Two aspects are still to be investigated, namely the bismuth distribution on the working electrode and the possible residual presence of the counteranion, namely phosphate. High-vacuum techniques such as electron microscopy or spectroscopy, which are commonly employed for this purpose, cannot be safely used: the bismuth surface is well-known to reconstruct and recrystallize under the electron beam in vacuum. Here, we demonstrate the suitability and the effectiveness of laser ablation ICP-MS (LA-ICP-MS, a technique that vaporizes and analyzes the surface material under flowing helium at atmospheric pressure) for the characterization of BiFE. Fast and stable measurements of bismuth and phosphorous distribution are achieved with the advantage of a minimum alteration of the sample surface, avoiding possible interferences. This investigation evidenced how, upon reductive activation, the bismuth film is distributed with a radial symmetry and the phosphate counteranion is completely absent on the working electrode surface.

1. Introduction

The pioneering works by Wang and Hocevar on bismuth [1,2] had the merit to draw considerable interest in this “green” element to prepare electroactive surface films as a low-toxicity alternative to conventional mercury films. In fact, bismuth shows evident analogies with mercury in forming alloys and adsorptive complexes with many metals [3]. Although many years have passed since the introduction of voltammetry and stripping for the detection of trace heavy metals, research on these electroanalytical techniques is still a hot topic, since many advantages may be seen compared to classical optical techniques, not simply their low cost. Interestingly, increased interest to “green” metal modifiers other than bismuth is also observed [4,5,6,7]. It is finally noteworthy that bismuth, thanks to its very low toxicity, will play a definite role for the electroanalytical studies of organic as well as biological molecules [8,9,10].
The preparation of an active bismuth film on a screen-printed electrode (BiFE-SPE) is generally carried out under the conventional electrochemical reduction of Bi3+ ion in aqueous electrolytes. This process is not a facile one, since it is reported to be sensitive to the nature of the bismuth precursor, the possible occurrence of kinetics limitations, and the sensitivity to the cleanliness of the carbon surface of the working electrode [11,12]. This wide range of experimental conditions leading to the optimal and active bismuth film may be rationalized on the basis of the rich chemistry of bismuth in aqueous solutions [13]. In this context, a decisive advantage was gained applying surface modification with polymers: this treatment permits to mitigate interferences and to improve the mechanical stability of the bismuth film [14,15]. As an additional benefit, some polymeric layers are reported to have a synergistic sensing effect (e.g., using carboxymethyl cellulose which is capable of chelating cadmium ions) [16].
An alternative and innovative approach was proposed by Brainina et al. [17]. This strategy involves the in situ formation of an insoluble bismuth salt (e.g., BiPO4) during the electrode preparation with the polymeric film. The bismuth film is then formed by reductive activation. However, the deposition of the insoluble salt precursor may be critical, and the following activation to the active bismuth surface may be complicated by liquid/solid phase kinetics. These conditions require a precise microscopical characterization of the bismuth film and the electrode surface, both before and after the electrochemical activation to ensure electrode performances. Two aspects are of main interest for their characterization: the bismuth distribution and the possible residual presence of the counteranion of the insoluble salt, namely phosphate.
A number of groups have used scanning and transmission electron microscopy [4,5,6,10,17,18,19] for the investigation of these features on bare and modified metal films on SPE, showing very different shapes and morphology. However, bismuth (or mercury) is a soft metal, and possible changes of the phase state of the thin film have been evidenced, due to the combined effect of high-vacuum and high-energy beam conditions [20]. Therefore, it should be taken into consideration the possibility that the observed shape and morphology of the metal surface may be profoundly dependent upon the experimental conditions of the high-vacuum experiment. Surface spectroscopies, primarily X-ray photoelectron spectroscopy (XPS), are also reported to be used to characterize metal films, as well as nanomodifiers, on screen-printed electrodes [21,22,23]. XPS is, in fact, very effective to probe the chemical composition of the outermost few nanometers of materials and the electrochemical activity strictly depends on the first atomic layers in contact with the electrolytic solution. However, no imaging information is likely to be obtained.
In this paper, the laser ablation ICP-MS technique has been proposed as a potential substitute (or complementary) technique of electron microscopy or surface spectroscopy for the investigation of the bismuth surface. Lasers have been used for the modification and functionalization of electrodes, and laser-induced forward transfer (LIFT) has emerged as a powerful physisorption method for the deposition of various materials on the electrode surface [24].
Here, we focused only on the analytical use of a scanning laser beam that vaporizes the surface material under flowing helium at atmospheric pressure. Such conditions should be fit for purpose without compromising bismuth stability. Although this technique possesses a low resolution, it has been already used for many applications for solid material characterization, giving information on the surface topology and on the qualitative and quantitative elemental composition [25,26]. We demonstrate here that the application of this technique is also suitable in the case of the analysis of bismuth films on screen-printed electrodes for electroanalysis. Laser ablation ICP-MS can permit a fast and stable measurement of the distribution of this metal on the electrode, with the advantage of a reduced alteration of the sample surface, avoiding possible interferences.

2. Materials and Methods

2.1. Apparatus and Reagents

Screen-printed electrodes (SPEs) have been a kind gift by Prof. Giuseppe Palleschi and his group at the University of Roma “Tor Vergata”. The printing procedures of the electrodes on a flexible polyester film have been described in the literature [27,28,29].
An optical microscopy view of the electrode is shown in Figure 1A, where its three-electrode structure is evidenced. The carbon working electrode is in the center, with a diameter of 0.3 cm, along with the silver pseudo-reference electrode on the top right, and the surrounding carbon counter-electrode on the left.
Hydrochloric and nitric acid were TraceSelect from Fluka, and they were purified by sub-boiling distillation in a quartz-Teflon apparatus [30]. Chemical reagents (disodium hydrogen phosphate, potassium nitrate, sodium sulphate) were purchased from Aldrich. Poly (sodium-4-styrene sulfonate) (PSS, Aldrich, St. Louis, MO, USA) was used as such.
Standard stock solutions of Cd and Pb were purchased from Aldrich, and bismuth stock solution was prepared from bismuth nitrate pentahydrate (CarloErba). These 1000-mg/L solutions were diluted as required for the preparation of standards.
For standards and solutions, as well as glassware cleaning, high-purity MilliQ water from MilliPore was used.

2.2. Preparation of Bismuth Films Screen-Printed Electrodes from Insoluble Bismuth Phosphate

Pristine screen-printed electrodes have been used as received after cleaning in HCl solution. Polystyrene sulfonate (PSS)-modified Bi/PO4 electrodes were prepared via deposition of the polymer layer and of the insoluble bismuth phosphate, following the 2-step recipe in Reference [13]:
Step 1: 2 μL of a 25 mM solution of PSS is drop-casted onto the working electrode with a microsyringe and left for drying.
Step 2: equal volumes of Na2HPO4 (4 mM) and Bi(NO3)3 (2 mM) were mixed, and 5 μL of this solution is then drop-casted onto the working electrode and left for drying. These electrodes can be stored in air at room temperature.
Prior of their use for the electrochemical analysis, the active bismuth film is formed by cycling eight times in 0.01 M HCl (pH = 2) under the electroanalytical conditions for the differential pulse stripping analysis described below in Section 2.3. The activated electrodes must be stored in a diluted acid solution without contact to air.
For the nitrate or sulphate interference study, sodium sulphate or potassium nitrate were added equimolar to 0.01 M HCl during repetitive scan experiments.

2.3. Measurement Procedures

PalmSens (Palm Instrument BV, Houten, The Netherlands) and Amel4330 (Amel srl, Milano, Italy) instruments were used. All instrument control as well as the full data analysis were handled by the Windows-based VApeak software.
A special cell was designed in collaboration with Amel, in order to optimize mass transport to the electrodes. To avoid the drawbacks of the internal pseudo-reference electrode, an external 3 M KCl Ag/AgCl reference electrode was used in order to reference all potentials.
The analysis of cadmium and lead was performed in 0.01 M HCl, or 0.01 M HNO3, using differential pulse stripping voltammetry (DPV). High-purity nitrogen for degassing and blanketing was used throughout the entire stripping experiment. Both the activation of the electroactive bismuth film and the stripping analysis were done under the sane conditions. Potential was initially set at −1.05 V for 60 s at a stirring rate of 300 rpm; then, stirring was stopped, and the DPV analysis started, scanning the potential from −0.95 to −0.25 V, with a step potential of 5 mV, pulse time of 100 ms, and pulse amplitude of 50 mV.
Peak areas measured as (μA × mV) were used for quantitation. The best-fit linear calibration line was calculated as y = mx + q. Analytical sensitivity was defined as the slope (m) of the best-fit calibration line. Limits of detection (LOD) were calculated as 3σ/m and reported as μg/L. σ was determined from the standard deviation of 10 replicate analyses on a standard Pb and Cd sample with concentrations near the limit of detection.

2.4. LA-ICP-MS Analysis

An inductively coupled plasma-mass spectrometry (ICP-MS), model Thermo Scientific ICAP Q, coupled with a fourth harmonic Nd: YAG 266 nm laser system, model New Wave UP Series 266, was used for the laser ablation-ICP-MS (LA-ICP-MS) measurements. The sample chamber for laser scanning (a viscous film sample chamber was employed in this work, see reference [31] for details concerning the design and analytical performances) was flushed with high-purity helium, that was then interfaced with the ICP torch. A careful setting of the instrumental parameters has been done in order to optimize the mass signal, in terms both of speed and signal-to-noise ratio. The optimized parameters of the laser ablation ICP-MS instrument are reported in Table 1.
Optical microscopy views of the screen-printed electrode prior and after a full laser ablation scan are reported in Figure 1A,B, respectively. The six laser scans are clearly visible as thin dark lines. During the laser scan, the ablated material is transported by the helium flow into the ICP torch, and mass signals at 13C, 31P, and 209Bi are continuously recorded using a dwell time of 40 ms. The 13C signal is used as background to reference phosphorous and bismuth m/z signals in order to improve the signal stability and signal-to-noise ratios.

3. Results

3.1. Electrode Preparation

The concentration of bismuth in the pre-deposited BiPO4 phase as a precursor of the active Bi0 metal phase in screen-printed electrodes is known to have a profound effect on analytical performances. In both References [13,17], the best performances are shown by the lowest Bi concentration, with an ideal value around 1 mM in terms of peak shape and increased linearity at low analyte concentrations. For laser ablation studies, instead, the need to increase bismuth loadings was immediately evident, with respect to the best working low-loading electrode, in order to increase signal-to-noise ratios. Therefore, in this paper, bismuth film electrodes have been prepared and studied using an intermediate concentration of bismuth, namely 2 mM, trying to find a good compromise between analytical performance and S/N ratios in ICP-MS.
These bismuth film electrodes (BiFE) have been tested in the trace analysis of Cd and Pb under differential pulse stripping and the results are reported in Figure 2.
Baselines are very reproducible and free of noise; peaks are found at the expected potential, i.e., around −800 mV for Cd and around −530 mV for Pb. They show only a minor broadening observed as a shoulder, which is slightly more evident for Pb. For this reason, calibration curves are done on the integrated signals.
Calibration curves are very straight, and limits of linearity were not reached at the maximum concentration of 75 μg/L used in the calibration. Low-concentration nonlinearity is negligible and it can be seen by the small deviation of the X-intercept from zero (see inset in Figure 2). The non-linearity of bismuth film electrodes is well-known but does not significantly affect the limits of detection (LODs) for Cd and Pb which are in the μg/L range around 1.2 μg/L for Cd and 0.9 μg/L for Pb. These results compare well with those reported for BiPO4-derived BiFEs.
The stability of the bismuth phase has been checked by running a replicate analytical scan of Pb and Cd at 15 μg/L and checking baselines after each repetition. Baselines are always stable and free of noise, and the integrated analytical signals are not significantly decreased after 20 replicates. This is in full agreement with those previously reported, and, therefore, this is not reported here.
The effect of anions different from chloride in the matrix has been studied, in order to understand whether residual phosphate ions after activation may affect the analytical performance. Adding nitrate ions during a repetitive scan indicated a small shift in the peak maxima of about +20 mV and a small, but significant, loss of signal of about 10% (Figure 3).
This effect is then linked to a decrease in the analytical sensitivity, as evidenced in Figure 4 by comparing the calibration lines for Pb analysis in HCl or HNO3 matrix solutions at pH = 2. A similar effect was found with sulphate ion (data not shown here).
Therefore, a detailed characterization was thought to be interesting, both for investigating the structure of the active bismuth layer as well as for fully understanding the chemistry occurring during the reductive activation of the bismuth salt layer upon activation and electrochemical analysis, to understand whether phosphate ions are fully removed prior the analysis, or are still playing a role in the electrochemical analysis.

3.2. Laser Ablation-ICP-MS Characterization

The characterization of bismuth film deposited onto the screen-printed working electrode was carried out by scanning the high-power UV laser beam through the X-axis of the dry electrode, while the ablated material is analyzed in real time by ICP-MS.
The analysis is first performed on an untreated screen-printed electrode right after the preparation of the BiPO4 layer via drop-casting deposition onto the carbon working electrode. The first result on a single scan line is reported in Figure 5, where the traces corresponding at the two isotopes of 209Bi and 31P are compared. The signal is, however, a little bit noisy, and a 10-point moving average smoothing is routinely applied throughout the paper.
It is immediately evident that the background signal for 209Bi is very low, since no bismuth is expected to be present on the pristine (untreated) screen-printed electrode. The 209Bi signal is almost perfectly confined at the center of the scan, which corresponds exactly to the borders of the carbon working electrode, with minimal spreading outside (please note in Figure 5 that the 209Bi signal is appreciable for a length of 0.3 cm, i.e., the diameter of the working electrode). This may demonstrate that the careful choice of the drop-casting procedure may lead to the preferential formation of the BiPO4 phase on the carbon surface. Obviously, this may be helped by the hydrophobic character of the plastic surface of the screen-printed electrode (SPE), preventing unwanted spreading during evaporation of the two subsequent layers of PSS and Na2HPO4 + Bi(NO3)3 solution.
Looking in more detail at the concentration profile of 209Bi, the presence of peaks and valleys reveals the inhomogeneous nature of the Bi distribution over the working electrode. Since the intensity signal is essentially correlated with the thickness of the bismuth layer onto the surface, the irregular shape of the bismuth signal may suggest an uneven Bi distribution with the presence of much thicker aggregates giving the most intense signals.
It is interesting to compare the traces of 209Bi and 31P on the same scanning line. The phosphorous signal is much less intense than the bismuth one (please note that different scales were used for 209Bi and 31P signals in Figure 5): this is a result of the much lower sensitivity of m/z = 31 of phosphorous with respect to the m/z = 209 signal of bismuth, which is characterized by a very high sensitivity. Secondly, some phosphorous has to be pristinely present in the polymeric base of the electrode, leading to high background values. Both facts have the consequence that the 31P signal shows low S/N ratios, and its plots are intrinsically less diagnostic than 209Bi. Thirdly, it is noteworthy to evidence that the concentration profile of 31P closely resembles that of 209Bi, although it is less defined as said before. This is a clear indication that BiPO4 is the prevailing chemical form present on the electrode surface, with only a residual presence of Bi-rich or P-rich phases which could be formed upon evaporating the aqueous impregnation solution of Na2HPO4 and Bi(NO3)3.
In Figure 6, the contouring plots of 209Bi (Figure 6A) and 31P (Figure 6B) are shown.
The spatial resolution is, unfortunately, quite poor due to the spot size of the laser beam which is necessary for having a reasonable signal for phosphorous. In any case, it can be immediately evidenced the presence of large aggregates of a microcrystal, where bismuth and phosphorous seem to be simultaneously present in the form of BiPO4 crystals.
The laser ablation ICP-MS results on the used electrodes after electrochemical activation are reported in Figure 7 and show a totally different chemical and structural situation.
In fact, the plot at m/z = 209 (Figure 7A) shows a pretty regular, spherical distribution of bismuth, with the maximum concentration on the center of the graphite surface of the working electrode. Instead, the phosphorous signal at m/z = 31 of Figure 7B is completely lost on the electrode surface, with some residual signals being only left at the borders, where the potential density during the electrochemical measurement is minimal. These results are very important, in that they confirm our previous hypothesis that no phosphorous should be left upon electrochemical activation of the BiPO4 surface. The possible effect of anions different from chloride, that was shown to slightly decrease the analytical sensitivity and peak intensity, is to be independent from the residual presence of phosphate, or phosphorous species, after electrode activation, as it was previously supposed. This is to be related to two possible causes: (i) to the different diffusivity of the anions within the PSS layer, or (ii) to the complexing properties of chloride ions, with respect to nitrate or sulphate. This second hypothesis is thought to be the prevailing factor, as it may be evidenced by the positive potential shift in the peak position of Pb and Cd. In fact, bismuth surface activation is due to a reductive elimination reaction when negative potential is applied, leading to the reduction of Bi(III) to Bi(0) and the parallel elimination of phosphate ions for the redox balance. The experimental conditions that have been chosen then allow the zerovalent bismuth phase to slowly grow, reaching its maximum thickness at the center of the working electrode. In fact, the kinetics of this reductive elimination process is mainly governed by the solid–liquid electrolyte interphase. This regular shape of the bismuth surface may also explain the good reproducibility and stability of the stripping analyses.

4. Conclusions

Our preliminary investigation has shown how laser ablation ICP-MS may play an important role in the characterization of electrochemical surfaces. In fact, the electrode surface is simply dried in a nitrogen atmosphere before placing it under the laser compartment under helium at atmospheric pressure. In this way, no contamination from air is possible, and no restructuring of the mobile bismuth (or mercury) surface can occur, as would be taking place for conventional techniques, such as scanning electron microscopy, using high-energy electron or photon beams under high-vacuum conditions.
This investigation clearly evidenced how the phosphate counteranion is completely lost upon reductive activation in the electrochemical cell, leading to a very clean, “ball-type” bismuth active surface. The formation of such a bismuth surface may account for any anion-type interference, as well as of the non-linearities at low analyte concentrations with increasing bismuth loadings.
Finally, these screen-printed electrodes, based on Brainina’s original ideas, are demonstrated to be viable alternatives to the in situ-prepared bismuth film electrodes. They are indefinitely stable in air, easily activated under the reductive conditions of the stripping electroanalytical experiment, and their final properties may be tuned by varying the bismuth concentration and deposition conditions.

Author Contributions

Conceptualization, C.D., D.M., A.P. and S.R.; methodology, C.D., D.M., A.P. and S.R.; formal analysis, C.D.; data curation, C.D., G.B. and D.S; writing—original draft preparation, C.D., G.B. and D.S.; All authors have read and agreed to the published version of the manuscript.

Funding

Financial support from Fondo di Ateneo per la Ricerca (FAR)—Università degli Studi dell’Insubria is greatly acknowledged.

Acknowledgments

The authors wish to thank Giuseppe Palleschi and his group at the University of Rome “Tor Vergata” for kindly providing the screen-printed electrodes and for scientific discussions. Research collaboration with Amel srl is greatly acknowledged.

Conflicts of Interest

The authors declare no conflict of interest.

References

  1. Wang, J.; Lu, J.; Hocevar, S.B.; Farias, P.A.; Ogorevc, B. Bismuth-Coated Carbon Electrodes for Anodic Stripping Voltammetry. Anal. Chem. 2000, 72, 3218–3222. [Google Scholar] [CrossRef] [PubMed]
  2. Hoĉevar, S.B.; Ogorevc, B.; Wang, J.; Pihlar, B. A Study on Operational Parameters for Advanced Use of Bismuth Film Electrode in Anodic Stripping Voltammetry. Electroanalysis 2002, 14, 1707–1712. [Google Scholar] [CrossRef]
  3. Grabarczyk, M.; Adamczyk, M. Bismuth film electrode and chloranilic acid as a new alternative for simple, fast and sensitive Ge(IV) quantification by adsorptive stripping voltammetry. RSC Adv. 2018, 8, 15215. [Google Scholar] [CrossRef] [Green Version]
  4. Lu, Y.; Liang, X.; Niyungeko, C.; Zhou, J.; Xu, J.; Tian, G. A review of the identification and detection of heavy metal ions in the environment by voltammetry. Talanta 2018, 178, 324–338. [Google Scholar] [CrossRef]
  5. Yao, Y.; Wu, H.; Ping, J. Simultaneous determination of Cd(II) and Pb(II) ions in honey and milk samples using a single-wallet carbon nanohorns modified screen-printed electrochemical sensor. Food Chem. 2019, 274, 8–15. [Google Scholar] [CrossRef]
  6. Liu, X.; Yao, Y.; Ying, Y.; Ping, J. Recent advances in nanomaterial- enabled screen-printed electrochemical sensors for heavy metal detection. TrAC 2019, 115, 187–202. [Google Scholar] [CrossRef]
  7. Economou, A. Screen-printed electrodes modified with “Green” metals for electrochemical stripping analysis of Toxic Elements. Sensors 2018, 18, 1032. [Google Scholar] [CrossRef] [Green Version]
  8. Lezi, N.; Vyskočil, V.; Economou, A.; Barek, J. Electroanalysis of Organic Compounds at Bismuth Electrodes: A Short Review. In Sensing in Electroanalysis; Kalcher, K., Metelka, R., Svancara, I., Vytras, K., Eds.; University Press Centre: Pardubice, Czech Republic, 2012; Volume 7, pp. 71–78. [Google Scholar]
  9. Tyszczuk-Rotko, K.; Surowiec, K.; Szwagierek, A. Application of Eco-friendly Bismuth Film Electrode for the sensitive determination of Rutin. Curr. Pharm. Anal. 2018, 14, 571–577. [Google Scholar] [CrossRef]
  10. Vladislavic, N.; Buzuk, M.; Brinic, S.; Byljac, M.; Bralic, M. Morphological characterization of ex-situ prepared bismuth film electrodes and their application in the electroanalytical determination of the biomolecules. J. Sol. State Electr. 2016, 20, 2241–2250. [Google Scholar] [CrossRef]
  11. Mandil, A.; Amine, A. Screen-Printed Electrodes Modified by Bismuth Film for the Determination of Released Lead in Moroccan Ceramics. Anal. Lett. 2009, 42, 1245–1257. [Google Scholar] [CrossRef]
  12. Calvo Quintana, J.; Arduini, F.; Amine, A.; Van Velzen, K.; Palleschi, G.; Moscone, D. Part two: Analytical optimisation of a procedure for lead detection in milk by means of bismuth-modified screen-printed electrodes. Anal. Chim. Acta 2012, 736, 92–99. [Google Scholar] [CrossRef] [PubMed]
  13. Dossi, C.; Monticelli, D.; Pozzi, A.; Recchia, S. Exploiting chemistry to improve performance of screen-printed, bismuth film electrodes (SP-BiFE). Biosensors 2016, 6, 38. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  14. Bedin, K.C.; Mitsuyasu, E.Y.; Ronix, A.; Cazetta, A.L.; Pezoti, O.; Almeida, V.C. Inexpensive bismuth-film electrode supported on pencil-lead graphite for determination of Pb(II) and Cd(II) ions by anodic stripping voltammetry. Int. J. Anal. Chem. 2018, 1473706. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  15. Kefala, A.; Economou, A.; Voulgaropoulos, A. A study of Nafion-coated bismuth-film electrodes for the determination of trace metals by anodic stripping voltammetry. Analyst 2004, 129, 1082–1090. [Google Scholar] [CrossRef] [PubMed]
  16. Ning, J.; Luo, X.; Wang, F.; Huang, S.; Wang, J.; Liu, D.; Liu, D.; Chen, D.; Wei, J.; Liu, Y. Synergetic sensing effect of sodium carboxymethyl cellulose and bismuth on cadmium detection by differential pulse anodic stripping voltammetry. Sensors 2019, 19, 5482. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  17. Malakhova, N.A.; Stojko, N.Y.; Brainina, K.Z. Novel approach to bismuth modifying procedure for voltammetric thick film carbon containing electrodes. Electrochem. Comm. 2007, 9, 221–227. [Google Scholar] [CrossRef]
  18. Vladislavic, N.; Buzuk, M.; Buljac, M.; Kozuk, S.; Bralic, M.; Brinic, S. Sensitive electrochemical determination of folic acid using ex-situ prepared bismuth film electrodes. Croat. Chem. Acta 2017, 90, 231–239. [Google Scholar] [CrossRef]
  19. Bernardelly, J.K.B.; Lapolli, F.R.; Gomes da Silva Cruz, C.M.; Floriano, J.B. Determination of zinc and cadmium with characterized electrodes of carbon and polyurethane modified by a bismuth film. Mat. Res. 2011, 14, 366–371. [Google Scholar] [CrossRef]
  20. Yu Kolosov, V.; Yushkov, A.A.; Veretennikov, L.M. Recrystallization and investigation of bismuth thin film IOP by means of electron beams in transmission electron microscopy. J. Phys. Conf. Ser. 2018, 1115, 032087. [Google Scholar] [CrossRef]
  21. Tyszczuk-Rotko, K.; Metelka, R.; Vytřas, K.; Barczak, M.; Sadok, I.; Mirosław, B. A simple and easy way to enhance sensitivity of Sn(IV) on bismuth film electrodes with the use of a mediator. Mon. Chem. Chem. Mon. 2016, 147, 61–68. [Google Scholar] [CrossRef]
  22. Xie, R.; Zhou, L.; Lan, C.; Fan, F.; Xie, R.; Tan, H.; Xie, T.; Zhao, L. Nanostructured carbon black for simultaneous electrochemical determination of trace lead and cadmium by differential pulse stripping voltammetry. R. Soc. Open Sci. 2018, 5, 180282. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  23. Mazzaracchio, V.; Tomei, M.R.; Cacciotti, I.; Chiodoni, A.; Novara, C.; Castellino, M.; Scordo, G.; Amine, A.; Moscone, D.; Arduini, F. Inside the different types of carbon black as nanomodifiers for screen-printed electrodes. Electrochim. Acta 2019, 317, 673–683. [Google Scholar] [CrossRef]
  24. Milano, F.; Giotta, L.; Chirizzi, D.; Papazoglou, S.; Kryou, C.; De Bartolomeo, A.; De Leo, V.; Guascito, M.R.; Zergioti, I. Phosphate Modified Screen Printed Electrodes by LIFT Treatment for Glucose Detection. Biosensors 2018, 8, 91. [Google Scholar] [CrossRef] [Green Version]
  25. Monticelli, D.; Di Iorio, A.; Ciceri, E.; Castelletti, A.; Dossi, C. Tree ring microanalysis by LA–ICP–MS for environmental monitoring: Validation or refutation? Two case histories. Microchim. Acta 2009, 164, 139–148. [Google Scholar] [CrossRef]
  26. Amman, A.A. Inductively coupled plasma mass spectrometry (ICP MS): A versatile tool. J. Mass Spectrom. 2007, 42, 419–427. [Google Scholar] [CrossRef] [PubMed]
  27. Arduini, F.; Amine, A.; Moscone, D.; Ricci, F.; Palleschi, G. Fast, sensitive and cost-effective detection of nerve agents in the gas phase using a portable instrument and an electrochemical biosensor. Anal. Bioanal. Chem. 2007, 388, 1049–1057. [Google Scholar] [CrossRef]
  28. Arduini, F.; Cassisi, A.; Amine, A.; Ricci, F.; Moscone, D.; Palleschi, G. Electrocatalytic oxidation of thiocholine at chemically modified cobalt hexacyanoferrate screen-printed electrodes. J. Electroanal. Chem. 2009, 626, 66–74. [Google Scholar] [CrossRef]
  29. Palchetti, I.; Laschi, S.; Mascini, M. Miniaturised stripping-based carbon modified sensor for in field analysis of heavy metals. Anal. Chim. Acta 2005, 530, 61–67. [Google Scholar] [CrossRef]
  30. Monticelli, D.; Castelletti, A.; Civati, D.; Recchia, S.; Dossi, C. How to Efficiently Produce Ultrapure Acids. Int. J. Anal. Chem. 2019, 2019, 1–5. [Google Scholar] [CrossRef]
  31. Monticelli, D.; Civati, D.; Giussani, B.; Dossi, C.; Spanu, D.; Recchia, S. A viscous film sample chamber for Laser Ablation Inductively Coupled Plasma—Mass Spectrometry. Talanta 2018, 179, 100–106. [Google Scholar] [CrossRef]
Figure 1. Optical micrographs of a freshly prepared, untreated Bi/PO4/PSS electrode before (A) and after (B) the laser ablation ICP-MS experiment.
Figure 1. Optical micrographs of a freshly prepared, untreated Bi/PO4/PSS electrode before (A) and after (B) the laser ablation ICP-MS experiment.
Biosensors 10 00119 g001
Figure 2. Differential pulse, anodic stripping voltammetry analyses of Cd and Pb solutions, and corresponding calibration curves (inset) at 0, 25, 50 and 75 μg/L on a working Bi/PO4/PSS electrode.
Figure 2. Differential pulse, anodic stripping voltammetry analyses of Cd and Pb solutions, and corresponding calibration curves (inset) at 0, 25, 50 and 75 μg/L on a working Bi/PO4/PSS electrode.
Biosensors 10 00119 g002
Figure 3. Effect of nitrate addition (blue trace) on Pb and Cd analysis in HCl at pH = 2 (red trace) on a working Bi/PO4/PSS electrode.
Figure 3. Effect of nitrate addition (blue trace) on Pb and Cd analysis in HCl at pH = 2 (red trace) on a working Bi/PO4/PSS electrode.
Biosensors 10 00119 g003
Figure 4. Comparison between calibration lines for Pb analysis in 0.01 M HCl (red line) and 0.01 M HNO3 (blue line) on a working Bi/PO4/PSS electrode.
Figure 4. Comparison between calibration lines for Pb analysis in 0.01 M HCl (red line) and 0.01 M HNO3 (blue line) on a working Bi/PO4/PSS electrode.
Biosensors 10 00119 g004
Figure 5. Comparison of the 209 Bi (red line) and 31P (blue line) laser ablation profiles located at 260 µm on a freshly prepared, untreated Bi/PO4/PSS electrode.
Figure 5. Comparison of the 209 Bi (red line) and 31P (blue line) laser ablation profiles located at 260 µm on a freshly prepared, untreated Bi/PO4/PSS electrode.
Biosensors 10 00119 g005
Figure 6. LA-ICP-MS contouring plots for 209Bi ((A), top) and 31P ((B), bottom) on a freshly prepared, untreated Bi/PO4/PSS electrode.
Figure 6. LA-ICP-MS contouring plots for 209Bi ((A), top) and 31P ((B), bottom) on a freshly prepared, untreated Bi/PO4/PSS electrode.
Biosensors 10 00119 g006
Figure 7. LA-ICP-MS contouring plots for 209Bi ((A), top) and 31P ((B), bottom) on a working Bi/PO4/PSS electrode after electrochemical activation and real analysis.
Figure 7. LA-ICP-MS contouring plots for 209Bi ((A), top) and 31P ((B), bottom) on a working Bi/PO4/PSS electrode after electrochemical activation and real analysis.
Biosensors 10 00119 g007
Table 1. List of instrumental parameters for the laser ablation (LA)-ICP-MS experiments.
Table 1. List of instrumental parameters for the laser ablation (LA)-ICP-MS experiments.
Laser Ablation Parameters
Scan Speed20 μm/s
% output20%
Step rate20 Hz
Laser spot size80 μm
Distance between each laser scan 260 μm
Mass Spectrometer Parameters
Dwell Time40 ms
PC Detector3300

Share and Cite

MDPI and ACS Style

Dossi, C.; Binda, G.; Monticelli, D.; Pozzi, A.; Recchia, S.; Spanu, D. Exploiting Laser-Ablation ICP-MS for the Characterization of Salt-Derived Bismuth Films on Screen-Printed Electrodes: A Preliminary Investigation. Biosensors 2020, 10, 119. https://doi.org/10.3390/bios10090119

AMA Style

Dossi C, Binda G, Monticelli D, Pozzi A, Recchia S, Spanu D. Exploiting Laser-Ablation ICP-MS for the Characterization of Salt-Derived Bismuth Films on Screen-Printed Electrodes: A Preliminary Investigation. Biosensors. 2020; 10(9):119. https://doi.org/10.3390/bios10090119

Chicago/Turabian Style

Dossi, Carlo, Gilberto Binda, Damiano Monticelli, Andrea Pozzi, Sandro Recchia, and Davide Spanu. 2020. "Exploiting Laser-Ablation ICP-MS for the Characterization of Salt-Derived Bismuth Films on Screen-Printed Electrodes: A Preliminary Investigation" Biosensors 10, no. 9: 119. https://doi.org/10.3390/bios10090119

APA Style

Dossi, C., Binda, G., Monticelli, D., Pozzi, A., Recchia, S., & Spanu, D. (2020). Exploiting Laser-Ablation ICP-MS for the Characterization of Salt-Derived Bismuth Films on Screen-Printed Electrodes: A Preliminary Investigation. Biosensors, 10(9), 119. https://doi.org/10.3390/bios10090119

Note that from the first issue of 2016, this journal uses article numbers instead of page numbers. See further details here.

Article Metrics

Back to TopTop