1. Introduction
Multi-organ-on-a-chip is a biomimetic platform that utilizes microfluidic chip and three-dimension (3D) cell culture to model human physiology and pathology in vitro. Since the publication of a lung-on-a-chip in 2010 [
1], this field has seen over a decade of development, resulting in various organ-specific chips, including liver-on-a-chip [
2,
3,
4], intestine-on-a-chip [
2,
5,
6,
7], brain-on-a-chip [
8,
9,
10], heart-on-a-chip [
11,
12], etc. These chips have already shown great potential for drug discovery and precision medicine.
In order to gain a deeper understanding of the interactions between tissues and organs, as well as their coordinated responses to external stimuli, it is essential to develop more sophisticated physiological and pathological models that incorporate interconnected organ communication networks [
13,
14]. One approach entailed the interconnection of multiple single-organ chips via small tubes, with pumps facilitating the flow of culture medium through the tubing to simulate blood flow, which enabled the exchange of substances and interactions among organs and tissues [
15,
16]. However, the introduction of external tubing led to an overall increase in the volume of fluid, which in turn resulted in a reduction in the concentration of organoid and tissue secretions within the microenvironment. This ultimately impeded the study of interactions between tissues and organoids. Another approach was to develop multi-organ chips, in which the channels and culture chambers were integrated into a single chip. Various tissues and organoids were cultured in designated culture chambers, which were interconnected through microfluidic channels [
17].
To further emulate the niches of the interior and exterior of blood vessels, a vascular barrier structure was incorporated between the culture chamber and the perfusion channel. This structure permitted the permeation of compound molecules, ions, cytokines, and exosomes, thereby facilitating communications between chambers and channels [
2,
18,
19,
20]. The construction of vascular barriers frequently employed a “sandwich” structure, wherein the culture chamber was situated above the fluid channel, separated by a porous membrane or scaffold seeded with vascular endothelial cells. A notable limitation of this approach was its restricted capacity for microscopic imaging. During brightfield imaging, the dense pores of the membrane impeded the observation of cells growing on the membrane. Furthermore, the in situ observation of the chip was impeded by the fact that the membrane or scaffold was situated at a greater distance from the objective, which made it challenging to distinguish between the different cells that were growing on either side of the membrane and to conduct high-resolution imaging of the tissues that were located above the membrane or scaffold. Consequently, the real-time observation and detection of processes such as the traversal of nanoparticles, immune cells, or circulating tumor cells across the barrier proved to be an arduous task.
To address these imaging challenges, the culture chambers and biomimetic blood vessels can be arranged in a horizontal pattern, with a hydrogel “wall” functionalized with endothelial cells serving as the vascular barrier. MIMETAS has developed chips featuring several parallel, enclosed channels, one of which is filled with type I rat tail collagen [
21,
22]. This hydrogel-filled channel communicates with adjacent parallel channels but is confined by raised “phaseguide” structures that prevent the hydrogel from overflowing into the side channels. Endothelial cells are then introduced to form an endothelial layer supported by the hydrogel wall. Other researchers have also utilized hydrogel in the construction of vascular network. A fibrin matrix was constructed in a gap in the bottom of the culture chamber and was connected to the culture medium reservoirs [
23]. The open culture chambers could accommodate organoids, while human umbilical vein endothelial cells (HUVECs) and human lung fibroblasts within the fibrin could form capillary networks under the influence of cytokines and shear stress, thereby simulating the capillary microenvironment of the organoids.
The brain and the heart are two of the most vital organs in the human body. The study of communication between brain and heart is significant for the advancement of our understanding of disease. Recently, researchers have identified the persistence of pro-inflammatory changes in monocytes/macrophages in multiple organs, particularly the heart, following a stroke. The changes have been linked to the development of cardiac fibrosis and dysfunction. This phenomenon has been observed in both mice and stroke patients. It has been demonstrated that IL-1β plays a pivotal role in the epigenetic alterations associated with innate immune memory [
24]. Additionally, damage to the endothelium of blood vessels has the potential to impact the brain and heart. Endothelial cells exhibit high secretory and pro-inflammatory properties in response to external injurious stimuli, and the pro-inflammatory cytokines and chemokines they produce transmit injury signals to parenchymal organs [
25].
This paper introduced an arrayed brain–heart chip comprising a biomimetic vascular channel that encircles the open organoid culture chambers. In each unit, a cerebral organoid and a cardiac organoid were, respectively, cultured in two separate open culture chambers. A fibrin matrix wall was positioned around the perimeter of the culture chamber, with HUVECs cultured on the external surface of the wall. The outer channels were designed to emulate the blood vessels, and perfusion flow was maintained through a rocking shaker. The perfusion method has been utilized for a high-throughput blood–brain barrier chip [
26]. The arrayed chip design permitted both automated pipetting and microscopic imaging. The numerical calculations were used to direct experimental optimizations, including surface hydrophilic treatment and the regulation of the loading volume of fibrinogen. Furthermore, they were utilized to guide the structural design of the convergence angle between the biomimetic vascular channel and the hydrogel channel. The morphology and permeability of the endothelial barrier were subsequently investigated. The utilization of this chip facilitated the co-culture of cerebral and cardiac organoids with an endothelial barrier and demonstrated the feasibility of investigating inter-organoid communication.
2. Materials and Methods
2.1. Fabrication and Assembly of the Chip
The polydimethylsiloxane (PDMS) structure was fabricated using soft lithography as described previously [
27], and briefly described as follows. First, SUEX
® Thick Dry Film Sheets (DJ Microlaminates) with thicknesses of 300 μm and 500 μm were sequentially transferred onto a silicon wafer. Photolithography was then performed using a maskless lithography machine (MicroWriter ML3, Durham Magneto Optics, Cambridge, UK) at energy densities of 12,000 J/cm
2 and 20,000 J/cm
2, respectively, to create a two-level master on a silicon wafer. Subsequently, a PDMS mixture (base elastomer: curing agent = 10:1
w/
w, Sylgard 184, Dow Corning, Midland, MI, USA) was poured onto the SU8 mold and cured at 60 °C for 4 h. The culture chambers and vascular channel inlets and outlets were punched out on the cured PDMS. The PDMS was then irreversibly bonded to plasma-treated glass slides and then sterilized at 120 °C using an autoclave. At last, the ultraviolet-sterilized polymethylmethacrylate (PMMA) was securely attached to the bonded chip, aligning the through-holes in the PMMA with the culture chambers and inlets/outlets of the vascular channel, thus increasing the liquid storage capacity.
2.2. Simulation of Fibrinogen Advancing in the Chip
The numerical simulations in this study were conducted using COMSOL Multiphysics 6.2 software. The spontaneous advancement of the aqueous phase within the chip microchannels was modeled as a two-phase flow problem. A 3D model was constructed for this purpose, utilizing the “Laminar Two-Phase Flow, Phase Field Interface” physics in the “Fluid Flow” module. At 0 °C, the viscosity and surface contact angle of the fibrinogen solution with the chip are approximately equal to that of water. Therefore, the aqueous phase in the simulation was defined as water. The Navier–Stokes equations and phase field equations were solved to track the interface between the water and air phases. The computational domain was divided into two regions: a rectangular region filled initially with water, while the rest was filled with air. The boundary conditions for the simulation were set as follows: (i) the initial velocities of both the water and air phases were set to 0 mm/s; (ii) the initial pressure of the water, both at the inlet and outlet, was set to the standard atmosphere; (iii) all chip walls were assigned no-slip boundary conditions and assumed to have identical contact angles and hydrophilicity. The simulations were based on the geometric parameters of the PDMS chip. In the phase field settings, the interface thickness parameter was set to 6.5 × 10⁻5 m, and the mobility tuning parameter was set to 50 m·s/kg. Additionally, the following assumptions were made while constructing the two-phase flow model: (i) The water phase was modeled as a homogeneous, incompressible Newtonian fluid with a density of 998 kg/m3 and a viscosity of 1.0 × 10⁻3 Pa·s. The air phase had a density of 1.2 kg/m3 and a viscosity of 1.8 × 10⁻5 Pa·s. (ii) No energy exchange or chemical reactions occurred between the computational domain and its surroundings. (iii) Gravitational effects were neglected.
2.3. Simulation of Perfusion Flow Field
To evaluate the perfusion within the chip, flow field simulations were also carried out using COMSOL Multiphysics. The flow field modeling was based on the incompressible Navier−Stokes equations and employed the “Laminar Flow” interface. In this simulation, the computational domain was treated as three-dimensional. The boundary conditions were set as follows: a pressure boundary condition was applied at the inlet, the outlet was maintained at zero pressure, and no-slip boundary conditions were applied to all walls.
2.4. Method for Construction of the Fibrin Matrix Barrier
A working solution of fibrinogen (20430ES, Yeasen Biotechnology, Shanghai, China) with a concentration of 10 mg/mL and thrombin (T8021, Solarbio, Beijing, China) with an activity of 20 U/mL was prepared in advance. The assembled chip was subjected to plasma treatment (200 W, 2 min) and then placed on ice for 2 min. Subsequently, thrombin and fibrinogen solutions were mixed at a 1:4 ratio, and 10 μL of the mixed solution was dropped onto each hydrogel inlet. The chip was then placed in a 37 °C incubator for 30 min to allow the hydrogel to fully polymerize. After polymerization, phosphate-buffered saline (PBS, pH7.4, Thermo Fisher Scientific, Waltham, MA, USA) was added to the chip, which was placed in 37 °C incubator for 12 h to allow for complete swelling and then stored in a clean, moist environment until needed.
2.5. Perfusion Culture of HUVECs and HUVEC-RFPs
HUVECs and HUVECs transfected with red fluorescent protein (HUVEC-RFPs) were obtained from Wuxi Puhe Biomedical Technology Co., Ltd. (Wuxi, China) and were used between passages 4 and 8. All cell types were cultured in EGM-2 (C3162, LONZA, Basel, Switzerland) in an incubator at 37 °C with 5% CO2. Cells were trypsinized, centrifuged, and resuspended to a concentration of 1 × 107 cells/mL.
The chip, with the fibrin barrier already constructed, was prepared by removing the PBS and injecting 50 μL of the cell suspension into the channels of the chip. The chip was then placed in a 37 °C incubator with 5% CO2 for culture. After 12 h, 200 μL EGM-2 was added in the inlet. The chip was then placed on a rocking shaker (IBAC ROCKER, Daxiang Bio, Beijing, China) for perfusion culture. The rocker was set with the following parameters: tilt angle of 10°, swing speed of 20 cycles/min, and a delay of 2 s between swings. Media was replaced daily, and cell growth and behavior were observed and recorded.
2.6. Measurement of Barrier Permeability
To measure the permeability of the barrier, a syringe pump was used to perfuse FITC -dextran solutions with concentrations of 1 mg/mL and molecular weights of 4 kDa and 70 kDa (46944, 46945, Merck, Darmstadt, Germany), respectively, into the chip’s perfusion vascular channel at a flow rate of 20 μL/min. To ensure that the barrier was saturated with the fluorescent solution, no solution was added to the culture chambers initially. After 15 min of perfusion, one of the culture chambers of the brain–heart chip was washed 1–2 times with PBS. To facilitate the collection of fluorescent molecules permeating the barrier, 50 μL of PBS was added to the culture chamber. And then the perfusion of the fluorescent solution was resumed. One minute later, 50 μL of the solution from the culture chamber was transferred to a black 96-well plate. An equal volume of PBS was added back to the culture chamber for the next sampling. Fluorescence intensity was measured using a microplate reader (Spark, Tecan, Mannedorf, Switzerland) and compared with a standard curve. Barrier permeability (
P) was calculated using the equation:
where
C1 and
C0 is the concentration of the sample and the initial solution, respectively, in the culture chamber,
C is the concentration of perfusion solution,
V is the volume of solution in the culture chamber,
A is the area of the barrier, and
t is the perfusion time.
2.7. Organoid Induction and Characterization
Human-induced pluripotent stem cells (iPSCs) (DYR0100, National Collection of Authenticated Cell Cultures, Shanghai, China) were used to induce cerebral and cardiac differentiation.
The induction of cerebral organoid was performed using the STEMdiff™ Cerebral Organoid Kit (08570, 08571, STEMCELL, Vancouver, BC, Canada) according to the provided protocol. Briefly, iPSCs, at a density of 9000 cells per well, were first cultured to form embryoid bodies (EBs) in a 96-well low-attachment U-bottom plate, starting on Day 0 (D0). On Days 2 and 4, 100 μL of EB formation medium was added to each well. Induction began on Day 5 (D5). When dense centers and radial translucent bands were observed in the EBs on Day 7 (D7), the EBs were embedded in Matrigel (354277, Corning, NY, USA) and cultured on a shaker with expansion medium. On Day 10 (D10), the expansion effect was assessed, and if extensive neuroepithelium was observed as evidenced by budding of the organoid surface, the medium was replaced with maturation medium. Media were exchanged every two days, and the culture was continued on a shaker for at least 10 additional days.
Cardiac organoids were derived using the Human iPSC-Derived Cardiac Organoid Differentiation Kit (RIPO-HWM002K, ACRObiosystems, Newark, DE, USA) following the provided protocol. Briefly, iPSCs, at a density of 7500 cells per well, were cultured to form EBs in a 96-well low-attachment U-bottom plate. The formation of EBs was assessed the following day, and EBs with clear, smooth edges and sizes between 300 and 500 μm were deemed suitable for further differentiation. Differentiation started on Day 0 (D0). After removing all media, 200 μL of medium A was added to each well and the culture was incubated for 48 h. Following this, the medium was replaced with medium B, which was changed every 24 h for a total of 4 changes. Subsequently, medium C was introduced, with changes every 24 h for a total of 5 changes. After this period, all cardiac organoids were transferred to a low-attachment 6-well plate and cultured with medium M-M on a shaker. On Day 10 (D10), pulsating cardiac organoids, indicating maturity, were observed.
Organoids were characterized using immunofluorescence to label specific proteins to confirm organoid identity (
Figure S1). To examine the transport of cardiac troponin I between two organoid culture environments, experiments were conducted using cardiac organoids with impaired cardiomyocytes, which exhibited reduced or ceased pulsation.
2.8. Organoids and Endothelial Cells Co-Culture in the Chip
A suspension of endothelial cells at a concentration of 1 × 107 cells/mL (50 μL) was introduced into the brain–heart chip. After three days of perfusion culture according to 2.5, the endothelial barrier was fully formed. Then, cerebral and cardiac organoids were placed in the two culture chambers, respectively, for further co-culture.
The cardiac medium M-M, cerebral maturation medium, and EGM-2 medium were introduced to the cardiac chamber, cerebral chamber, and perfusion channel inlet, with a volume of 200 μL each. The chip was then placed on a rocking shaker, which was set with the following parameters: tilt angle of 10°, swing speed of 20 cycles/min, and a delay of 2 s between swings. The culture media in each compartment were completely replaced every 24 h.
2.9. Cardiac Troponin I (cTnI) Measurement
After 3 days of co-culture, the media were replaced, and the culture was continued for an additional 12 h. Subsequently, 100 μL of solution was collected from the cardiac and cerebral chambers, respectively. cTnI levels were measured using a fluorescent immunoassay analyzer (Getein1100, Getein Biotech, Nanjing, China). The fresh complete medium corresponding to the organoid was used as a blank control. The final value was obtained by subtracting the blank control value from the measured values.
2.10. Characterization of Fibrin Barrier and Endothelial Barrier
Laser confocal microscopy 3D imaging was used to observe fibrin barriers constructed in chips. First, 50 μL of FITC (46950, Merck) fluorescent solution, at concentration of 100 μg/mL, was added to both the perfusion channel and the culture chamber and incubated for 10 min to facilitate the dyeing of the fibrin matrix. The fluorescent solution was then removed, and the chip was washed twice with an equal volume of PBS. The chip was then observed under a confocal microscope (TCS SP8, Leica, Wetzlar, Germany) with z-stack scanning performed to reconstruct a 3D image.
The morphology of the endothelial barrier on the surface of the fibrin matrix was characterized by cytoskeleton staining. Endothelial cells were fixed with 4% paraformaldehyde solution for 30 min, followed by two washes with PBS. TRITC-Phalloidin (EFL-FA-001, EFL-tech, Suzhou, China) diluted 1:1000, was added to the chip for 30 min to stain the F-actin. The chip was then washed twice with PBS. The stained chip was similarly imaged using a confocal microscope, with z-stack scanning and 3D reconstruction for further analysis.
4. Conclusions
The arrayed brain–heart chip developed in this study, with its biomimetic vasculature surrounding the organoid culture chambers, partially recapitulated the in vivo interconnected multi-organ environment. The structure is horizontally distributed, facilitating microscopic observation. The endothelialized hydrogel barrier not only maintained the specific culture media required by different organoids but also facilitated communication between the organoids. The cerebral and cardiac organoids could be stably co-cultured in the chip for at least one week.
The chip design was simple, and the construction of the hydrogel barrier was straightforward. Once the hydrogel was added to the sample inlet, capillary action completed the construction of the barrier with a success rate approaching 100%, making it highly scalable. The chip array was compatible with the standard 96-well plate layout, facilitating integration with automated liquid handling systems. In addition, the rocking shaker perfusion method allowed for high-throughput processing, making the chip promising for high-throughput drug screening.
During the experiment, it was observed that after one week of co-culture of organoids and endothelial cells, both endothelial cells and cerebral organoids began to penetrate the fibrin. Therefore, if a longer-term in vitro brain–heart connection model is required, alternative hydrogel materials may need to be explored. Furthermore, previous studies have demonstrated angiogenesis within the fibrin hydrogel [
21], suggesting that future studies could incorporate angiogenesis within the current hydrogel barrier to construct bioinspired vascular channels connecting the culture chambers and perfusion channels.
This chip had a modular design that could be easily expanded to create more interconnected organ chips as needed, providing versatile in vitro models for physiological and pathological studies. In the future, the chip could also integrate microelectrode arrays to monitor the electrical activity of electroactive organoids, further expanding its application potential.