Extraction of His Bundle Pacing Lead: More Difficult than Coronary Sinus Lead Extraction: An Analysis of 3897 Lead Extraction Procedures Including 27 His and 253 Coronary Sinus Lead Removals

by
Paweł Stefańczyk
1,
Wojciech Jacheć
2,
Andrzej Kutarski
3,
Paweł Dąbrowski
1,
Andrzej Głowniak
3 and
Dorota Nowosielecka
4,*
1
Department of Cardiology, The Pope John Paul II Province Hospital of Zamość, 22-400 Zamość, Poland
2
2nd Department of Cardiology, Faculty of Medical Sciences in Zabrze, Medical University of Silesia, 40-055 Katowice, Poland
3
Department of Cardiology, Medical University of Lublin, 20-093 Lublin, Poland
4
Department of Cardiac Surgery, The Pope John Paul II Province Hospital of Zamość, 22-400 Zamość, Poland
*
Author to whom correspondence should be addressed.
Biomedicines 2024, 12(6), 1154; https://doi.org/10.3390/biomedicines12061154
Submission received: 24 March 2024 / Revised: 8 May 2024 / Accepted: 20 May 2024 / Published: 23 May 2024
(This article belongs to the Section Molecular and Translational Medicine)

Abstract

:
Background: Experience with the transvenous extraction of leads used for His bundle pacing (HBP) is limited. Methods: Analysis of 3897 extractions including 27 HBP and 253 LVP (left ventricular pacing) leads. Results: The main reason for HBP lead extraction was lead failure (59.26%). The age of HBP and LVP leads (54.52 vs. 50.20 months) was comparable, whereas procedure difficulties were related to the LVP lead dwell time. The extraction of HBP leads > 40 months old was longer than the removal of younger leads (8.57 vs. 3.87 min), procedure difficulties occurred in 14.29%, and advanced tools were required in 28.57%. There were no major complications. The extraction time of dysfunctional or infected leads was similar in the HBP and LVP groups (log-rank p = 0.868) but shorter when compared to groups with other leads. Survival after the procedure did not differ between HBP and LVP groups but was shorter than in the remaining patients. Conclusions: 1. HBP is used in CRT-D systems for resynchronisation of the failing heart in 33.33%. 2. Extraction of HBP leads is most frequently performed for non-infectious indications (59.26%) and most often because of lead dysfunction (33.33%). 3. The extraction of “old” (>40 months) HBP leads is longer (8.57 vs. 3.87 min) and more difficult than the removal of “young” leads due to unexpected procedure difficulties (14.29%) and the use of second line/advanced tools (28.57%), but it does not entail the risk of major complications and procedure-related death and is comparable to those encountered in the extraction of LVP leads of a similar age. 4. Survival after lead extraction was comparable between HBP and LVP groups but shorter compared to patients who underwent the removal of other leads.

Graphical Abstract

1. What’s New?

Conduction System Pacing (CSP) delivered by His Bundle Pacing (HBP) or Left Bundle Pacing (LBP) appears to compete with biventricular pacing but may be of limited durability due to more frequent pacing/sensing failures than in other leads. Transvenous lead extraction is of key importance in the proper management of dysfunctional or infected leads. The lumenless Medtronic Select Secure 3830 lead is most popular for CSP, but experience with such lead extraction is very limited, especially if the leads are older than 4 years. This study presents a broad analysis of data to compare the difficulty, complexity, and complications of CSP and LVP lead extraction.

2. Introduction

Conduction system pacing (CSP) delivered by His Bundle Pacing (HBP) or Left Bundle Pacing (LBP) is a relatively new development in the field of more physiological pacing and cardiac resynchronisation therapy (CRT). Although the concept of HBP was created 23 years ago [1], it was widely introduced only a few years ago [2,3,4,5,6,7,8] both as an option for the resynchronisation of a failing heart [2,9] and as a non-desynchronising option for ventricular pacing in patients without indications for CRT [3,4,5,6,7,8]. Between 2018 and 2021, HBP was the most prevalent choice [2,3,4,5,6,7,8,9], and since 2019, it has been gradually replaced by LBP [10,11,12,13,14]. The effectiveness of CPS device implantation is 67–92% [1,3,7], whereas the durability of HBP is shorter than that of other types of cardiac pacing. The rate of dysfunctional HBP leads was 14.3% in the initial period [1], and now, it has ranged from 4.3 to 12.1% over a period of several years [2,3,4,5,6,7], but in the longer term, it increases to 53% [8]. It is expected that with the increasing popularity of this type of system, there will be more problems in the long-term follow-up of patients with CSP [15]. Transvenous lead extraction (TLE) is the first-line option for the management of dysfunctional, infected, and abandoned leads [16]. In most patients referred for CSP, the thin lumenless Medtronic Select Secure 3830 lead has been used, originally intended for children for whom its small diameter had a clear advantage over standard leads [15]. Experience with the extraction of these leads in 33 children after several (2–5) years of operation has been described [17,18,19]. However, our understanding of CSP lead extraction is very limited, as there is only one report presenting 30 extractions of HBP leads [20], one case series (nine cases) [21], and two case reports [22,23]. The lead implant duration in the cited studies was 25 months or significantly less, except one HBP lead with a dwell time of 14.3 years [22]. The knowledge of LBP lead extraction is even more modest, as it is based on six case reports [24,25,26,27,28,29].

3. Goal of the Study

The goals of this study were as follows: 1. to analyse the clinical and technical aspects of HBP lead extraction, including the influence of the lead dwell time on the effectiveness and complications of TLE; 2. to compare the difficulty and complexity of HBP and conventional LVP lead extraction, as well as survival after TLE in these groups.

4. Methods

4.1. Study Population

Data from 3897 transvenous lead extraction (TLE) procedures (27, HBP group; 253, LVP group; 3617, control group; TLE of other than HBP or LVP leads) performed between March 2006 and May 2023 at a single high-volume centre were reviewed.

4.2. Lead Extraction Procedure

4.2.1. Definitions

Indications for lead removal, procedure effectiveness, and complications were defined according to the lead management recommendations (2017 HRS consensus and 2018 EHRA guidelines) [16,30]. The risk of major complications (MCs) related to TLE was assessed using the SAFeTY TLE score, an online tool available at http://alamay2.linuxpl.info/kalkulator/ and http://usuwanieelektrod.pl/kalkulatory (accessed on 28 January 2020) [31].
The EROS score was used for the prediction of significant procedural complications that required emergent surgical intervention [32]. The assessment of procedure complexity was based on the MB—Mazzone-Bontempi (score), showing the need for the use of advanced tools to achieve TLE success [33], LED—Lead Extraction Difficulty index referring to lead extraction difficulty based on fluoroscopy times [34], Advanced LE Techniques (ALET) score to predict the necessity of using advanced extraction techniques [35], and the Complex Indicator of the Difficulty of the TLE (CID-TLE) and LECOM score based on time of lead extraction and use of metal sheath or Evolution/TightRail, lasso-catheters, or basket catheters or the use of an alternative approach [36].
Unexpected difficulties during the extraction procedure, so-called “technical problems”, were defined as the situations that increased the complexity of the procedure but were not complications [37].

4.2.2. Techniques of HBP Lead Extraction

The Medtronic Select Secure 3830 lumenless (Medtronic, MN, USA) lead was developed for use in the paediatric population due to its small diameter. Later, it proved to be excellent also for CSP. The lead is characterised by an exposed nonretractable helix [15]. These features affect the technical aspects of the TLE procedure. The impossibility of introducing a locking stylet requires the use a semi-rigid lead extender, e.g., Bulldog System (Cook Medical, Bloomington, IN, USA) [15,19,22]. Although paediatric experience has shown that a significant percentage of leads can be removed through manual traction alone [17,18,19], this technique was used only in patients with device infection. In the present study, most of the extraction procedures were performed to replace a dysfunctional lead; therefore, we tried to maintain the original venous access. As a rule, we used polypropylene sheaths of the smallest possible diameter to reduce the risk of dislodgement of the remaining functional leads. For practical reasons, we used a sharply bent standard stylet and a fixation ligature to extend the lead (Figure 1 and Figure 2).
It should be noted that in the case of the short age of the leads (<40 months), the problem was rather too early disruption of the lead from scar tissue, which could result in the accidental removal of the lead and loss of venous access. Therefore, when the tip of the lead was freed after gentle traction, we usually held it with a lasso inserted from the femoral access. This made it possible to obtain a safe “rail” effect for the polypropylene sheath [15,21]. In only a few cases, it was necessary to replace the sheath with a new one with a larger diameter, and Evolution had to be used only when we preferred to extract leads using non-powered mechanical polypropylene sheaths, bearing in mind that the remaining functional leads should be kept untouched if not planned to be removed (Figure 2).

4.3. Dataset and Statistical Methods

4.3.1. Creation of Subgroups for Analysis

Figure 3 illustrates how the study participants were divided into groups. The division of the study cohort into groups is presented in Figure 3. Groups 5, 6, and 7 were identified to assess the difficulty and efficiency of HBP and LVP lead extraction.
Groups 1, 2, 3, and 4 were identified to investigate the effect of the HBP and LVP lead implant duration on the difficulty and efficiency of the extraction procedure. HBP and LVP subgroups were selected based on the median age of the electrodes (HBP, 40 months; LVP, 44 months) (Figure 3).

4.3.2. Statistical Analysis

Continuous variables are presented as the mean ± standard deviation. The categorical variables are presented as counts and percentages. The significance of differences between the groups was determined using the non-parametric Chi2 test with Yates correction or the unpaired Mann–Whitney U test, as appropriate. The Bonferroni correction was applied for a comparison among groups 6, 5, and 7 (considering a p-value < 0.0166 as statistically significant). To determine the effect of HBP and LVP leads on the survival of the CIED system and survival of patients after TLE, Kaplan–Meier curves were plotted, and differences in their course were assessed using the log rank test. A p-value less than 0.05 was considered statistically significant. Statistical analysis was performed with Statistica 13.3 (TIBCO Software Inc., Tulsa, OK, USA).

4.3.3. Approval of the Bioethics Committee

All patients gave their informed written consent to undergo TLE and use anonymous data from their medical records, approved by the Bioethics Committee at the Regional Chamber of Physicians in Lublin no. 288/2018/KB/VII (approval date: 27 November 2018). The study was carried out in accordance with the ethical standards of the 1964 Declaration of Helsinki.

5. Results

In the HBP lead group consisting of 27 patients with a median of age of 70.52 years, there were 29.63% women, and 40.74% of the procedures were performed for infectious reasons. In the LVP lead group consisting of 253 patients with a median of age of 68.87 years, there were 20.95% women, and 62.05% of the procedures were performed for infectious reasons. In the remaining 3.617 patients (extraction of other than HBP and LVP leads) with an average age of 65.84 years, there were 39.37% women, and 29.08% of the procedures were performed for infectious reasons (Table 1).
Generally, the HBP and LVP groups, as well as their subgroups with “old” and “young” leads, did not differ from each other, i.e., they were made up of very similar populations. The HBP patients had a slightly higher LVEF than the LVP patients because HBP was not used to resynchronise the failing heart in some patients. Obviously, the control group had a higher LVEF and a significantly lower rate of a worse NYHA class than the patients with HBP and LVP leads. The complexity of the systems and multiple previous CIED-related procedures increased the risk of systemic infection expressed as the PADIT score, more in patients with LVP than HBP leads compared to the control group. Patients with older LVP leads were more likely to have mechanical lead damage and dysfunction, and this phenomenon was not observed in those with HBP leads (Table 1).
HBP leads were used more often in standard pacing systems (66.67%) than in CRTD systems (33.33%) with the opposite order for left ventricular leads, but in that case, the LVP lead was a component of the CRT-P system. Although the mean age of the oldest lead removed in the “old” HBP and LVP subgroups was similar to the age of the oldest lead removed in the control group (105.9, 90.74 and 103.5 months, respectively), the age of the oldest lead removed per patient in the HBP and LBP groups was lower than that in the control group (74.81, 68.12, and 103.5 months, respectively). The lead (HBP and LVP) implant duration was similar (54.52 and 50.20 months). Patients with systems containing HBP and LVP leads had more leads in the heart (2.52, 3.04 vs. 1.87) and more procedures (1.92, 2.09 vs. 1.84) before TLE than patients in the control group (Table 2).
Scores of risk factors for major complications or procedure complexity indicated greater difficulty and increased risk during the extraction of older leads in the HBP and LVP subgroups (SAFeTY LE, MB, LED, ALET, LECOM), but only the LECOM score indicated an increase in procedure complexity in the case of LVP versus HBP leads (27.09 vs. 19.41%). The LVP group was characterised by a higher rate of CIED-related procedures and a higher rate of removed passive fixation leads (excluding LV leads). This was due to historical reasons (in earlier years, LVP was used significantly more often than HBP) (Table 2).
The extraction duration was longer in the “old lead” subgroups, while the lead extraction time did not differ significantly between the HBP and LVP lead groups and the control group. The extraction of a single HBP/LVP lead was twice as long in the subgroups with “old” leads, while it did not differ in relation to the entire HBP and LVP groups (3.19 vs. 3.05 min). In 85.19% and 70.75% of patients, HBP and LVP leads were removed using mechanical dilation and when interpreting these percentages, it should be borne in mind that, respectively, 59.26% and 37.94% of the procedures were performed for non-infectious indications with the aim to replace the lead for a new one while maintaining the original venous access. Gentle screw-out and gentle traction to remove the HBP and LVP lead were mainly used in patients with device infection (14.81 and 28.97% respectively) (Table 3).
The procedure difficulty was similar in all the study groups and subgroups. The values of the CID-TLE score were higher in the subgroups with old leads but did not differ significantly among the HBP, LVP, and control group. Similarly, there were no differences in the number of more complex procedures (2 and more CID score points) (18.52, 28.06, and 18.30%, respectively).
Rates of major complications (MCs), rescue cardiac surgery, and procedure-related deaths were very low and did not differ between the study groups. Due to a higher rate of partial radiographic success (retained tip or <4 cm lead fragment) in the “old” LVP lead subgroup (more passive fixation leads removed), the rate of procedural success was slightly lower.
The FU period in survivors was shorter for patients with removed HBP leads, as this technique was introduced significantly later than LVP (Table 3).
The presence of an ICD lead (CRTD) before TLE was the only risk factor for death in patients undergoing the removal of His pacing leads (Table 4).
The time from implantation to TLE for non-infectious indications did not differ between the groups (log rank p = 0.868), and the time to TLE for infectious indications was comparable in the HBP and LVP groups and was shorter when compared to the group with other leads. Survival after TLE did not differ between HBP and LVP groups but was lower when compared to the group with other leads (Figure 4).

6. Discussion

The haemodynamic benefits of conduction system pacing (CSP) have long been known [1], but the intervention has become increasingly popular only in recent years [2,3,4,5,6,7,8,9]. The use of this type of physiological ventricular pacing has been extensively described in various situations: in patients with a normal heart to prevent post-pacing cardiomyopathy (bradycardia indications) [1,3,4,5,6,7,8,15], as a method of ventricular resynchronisation in patients with heart failure and wide QRS complexes [2,3,5,7,15], and as a component of resynchronisation systems (CRT-D), in which the electrode stimulating the conduction system (HBP, LBP) is used for resynchronisation and the ICD electrode is meant for high-voltage therapy only [5,9,15], a dysfunctional left ventricular transvenous pacing lead is replaced by a CSP lead [15] or implanted when LVP lead implantation fails [9,15].
In recent years, several limitations of HBP have been recognised, i.e., an unacceptable increase in the pacing threshold or a decrease in the sensed ventricular potential values, which makes a dozen or more percent of the leads cease to fulfil their role within a few years [1,2,3,4,5,6,7,8]. Today, the seemingly less sophisticated LBP lead guarantees better long-term performance, values of the sensed ventricular potential are much higher, the risk of early dislodgement is practically non-existent, and the pacing threshold values increase less frequently (in 8%) [15]. At present, we have an increasing population of patients with HBP leads implanted several years ago, and the demand for LBP lead replacement is likely to grow in the coming years [15]. The difficult technique of implanting leads into the CSP prolongs the implantation procedure, which increases the risk of infectious complications. The basic component of optimal lead management is transvenous lead extraction [16], and the odds are that the interest in CSP lead extraction techniques will increase dramatically.
Thin lumenless Select Secure 3885 leads are commonly used for CSP. Experience with their extraction comes from the paediatric population (34 children, lead extraction a few years after implantation), and three reports showed that 41–100% can be removed through manual traction [17,18,19]. The knowledge and experience of the extraction of CSP leads is very limited as it comes from one report of 30 HBP lead extractions [20], one case series (9 cases) [21], and 2 case reports [22,23] and one review paper [15]. Of the 41 HBP lead removals, mechanical dilatation was necessary in 10 patients (24.4%) [20,21,22,23]. There is no study on LBP lead extraction, apart from 6 case reports. There is no available literature on extraction failure or rupture of the Select Secure 3880 lead during extraction.
This study showed that the EF and NYHA class in the HBP group were better than in the LVP group. Patients with HBP and LVP had more leads in the heart and more CIED-related procedures before TLE than patients in the control group. The age of extracted HBP or LVP leads did not differ significantly (54.52 vs. 50.20 months). The extraction of “old” (>40 months) HBP leads was longer than the removal of “young” (<40 months) HBP leads (8.57 and 3.87 min). Mechanical dilatation was necessary in 69.23% and 100% of HBP leads, but it should be borne in mind that in 35.71% and 46.15%, respectively, the procedures had been performed for non-infectious indications with the intention to replace the lead for a new one while maintaining the original venous access. Unexpected procedure difficulties (“technical problems) appeared in 14.29%, and second line/advanced tools were used in 28.57% during the extraction of “old” HBP leads only. Difficulties in extracting HBP leads were comparable to those encountered in the extraction of LVP leads of similar age. Consistent with the cited reports [20,21,22,23] in this study, there were no major complications, rescue cardiac surgery, or procedure-related death during HBP and LVP lead extraction. The follow-up duration was shorter in patients with removed HBP leads, as this technique was introduced much later than LVP. Kaplan–Meier analysis showed comparable survival of patients with LVP and HIS lead extraction; however, it was shorter than in the control group.

7. Conclusions

  • HBP is used in CRT-D systems for resynchronisation of the failing heart in 33.33%.
  • The extraction of HBP leads is most frequently performed for non-infectious indications (59.26%) and most often because of lead dysfunction (33.33%).
  • The extraction of “old” (>40 months) HBP leads is longer (8.57 vs. 3.87 min) and more difficult than the removal of “young” leads due to unexpected procedure difficulties (14.29%) and the use of second line/advanced tools (28.57%), but it does not entail the risk of major complications and procedure-related death and is comparable to those encountered in the extraction of LVP leads of a similar age.
  • Survival after lead extraction was comparable between HBP and LVP groups but shorter compared to patients who underwent the removal of other leads.

Study Limitations

All procedures were performed using all types of mechanical systems but not laser-powered sheaths. The study aimed to assess the effectiveness and outcomes of HBP and LVP lead extraction. But complexity and major complications apply to the entire procedure and not to the extraction of one type of the lead(s). As removal of the HBP and LVP leads was often accompanied by the extraction of other pacemaker/ICD leads, we could never be sure which of these leads caused the complication. Patients with HBP and LVP leads often had abandoned leads or newer leads implanted during upgrade procedures. Also, procedure difficulties resulted not only from the fact of removing the specific (HBP and LVP) lead but also from the presence of additional leads. And last but not least, this is presentation of a single very experienced centre. For this reason, the outcomes may not represent the overall safety and efficacy of TLE especially in patients with a long implant duration.

Author Contributions

P.S.—preliminary draft preparing, W.J.—methodology, statistical study, investigation, A.K.—final draft editing and supervision, P.D.—data curation, A.G.—data curation, D.N.—data curation, corresponding author. All authors have read and agreed to the published version of the manuscript.

Funding

Association “Heart with Help”, Department of Cardiology, The Pope John Paul II Province Hospital 22-400 Zamość, Poland.

Institutional Review Board Statement

The study was performed according to the principles expressed in the Declaration of Helsinki and approved by the Bioethics Committee at the Regional Chamber of Physicians in Lublin (no. 288/2018/KB/VII). Approval date: 27 November 2018.

Informed Consent Statement

Informed consent was obtained from all subjects involved in the study.

Data Availability Statement

Readers can access the data supporting the conclusions of the study at www.usuwanieelektrod.pl.

Conflicts of Interest

The authors declare no conflicts of interest.

References

  1. Deshmukh, P.; Casavant, D.A.; Romanyshyn, M.; Anderson, K. Permanent, direct His-bundle pacing: A novel approach to cardiac pacing in patients with normal His-Purkinje activation. Circulation 2000, 101, 869–877. [Google Scholar] [CrossRef] [PubMed]
  2. Sharma, P.S.; Dandamudi, G.; Herweg, B.; Wilson, D.; Singh, R.; Naperkowski, A.; Koneru, J.N.; Ellenbogen, K.A.; Vijayaraman, P. Permanent His-bundle pacing as an alternative to biventricular pacing for cardiac resynchronization therapy: A multicenter experience. Heart Rhythm 2018, 15, 413–420. [Google Scholar] [CrossRef] [PubMed]
  3. Abdelrahman, M.; Subzposh, F.A.; Beer, D.; Durr, B.; Naperkowski, A.; Sun, H.; Oren, J.W.; Dandamudi, G.; Vijayaraman, P. Clinical outcomes of his bundle pacing compared to right ventricular pacing. J. Am. Coll. Cardiol. 2018, 71, 2319–2330. [Google Scholar] [CrossRef] [PubMed]
  4. Jastrzębski, M.; Moskal, P.; Bednarek, A.; Kiełbasa, G.; Czarnecka, D. His-bundle pacing as a standard approach in patients with permanent atrial fibrillation and bradycardia. Pacing Clin. Electrophysiol. 2018, 41, 1508–1515. [Google Scholar] [CrossRef]
  5. Zanon, F.; Abdelrahman, M.; Marcantoni, L.; Naperkowski, A.; Subzposh, F.A.; Pastore, G.; Baracca, E.; Boaretto, G.; Raffagnato, P.; Tiribello, A.; et al. Long term performance and safety of His bundle pacing: A multicenter experience. J. Cardiovasc. Electrophysiol. 2019, 30, 1594–1601. [Google Scholar] [CrossRef] [PubMed]
  6. Keene, D.; Arnold, A.D.; Jastrzębski, M.; Burri, H.; Zweibel, S.; Crespo, E.; Chandrasekaran, B.; Bassi, S.; Joghetaei, N.; Swift, M.; et al. His bundle pacing, learning curve, procedure characteristics, safety, and feasibility: Insights from a large international bservational study. J. Cardiovasc. Electrophysiol. 2019, 30, 1984–1993. [Google Scholar] [CrossRef]
  7. De Leon, J.; Seow, S.C.; Boey, E.; Soh, R.; Tan, E.; Gan, H.H.; Lee, J.Y.; Teo, L.J.T.; Yeo, C.; Tan, V.H.; et al. Adopting permanent His bundle pacing: Learning curves and medium-term outcomes. Europace 2022, 24, 606–613. [Google Scholar] [CrossRef] [PubMed]
  8. Frausing, M.H.J.P.; Bæk, A.L.; Kristensen, J.; Gerdes, C.; Nielsen, J.C.; Kronborg, M.B. Long-term follow-up of selective and non-selective His bundle pacing leads in patients with atrioventricular block. J. Interv. Card. Electrophysiol. 2023; online ahead of print. [Google Scholar] [CrossRef]
  9. Jastrzębski, M.; Moskal, P.; Huybrechts, W.; Curila, K.; Sreekumar, P.; Rademakers, L.M.; Ponnusamy, S.S.; Herweg, B.; Sharma, P.S.; Bednarek, A.; et al. Left bundle branch-optimized cardiac resynchronization therapy (LOT-CRT): Results from an international LBBAP collaborative study group. Heart Rhythm 2022, 19, 13–21. [Google Scholar] [CrossRef]
  10. Zhang, W.; Huang, J.; Qi, Y.; Wang, F.; Guo, L.; Shi, X.; Wu, W.; Zhou, X.; Li, R. Cardiac resynchronization therapy by left bundle branch area pacing in patients with heart failure and left bundle branch block. Heart Rhythm 2019, 16, 1783–1790. [Google Scholar] [CrossRef]
  11. Vijayaraman, P.; Ponnusamy, S.; Cano, Ó.; Sharma, P.S.; Naperkowski, A.; Subsposh, F.A.; Moskal, P.; Bednarek, A.; Dal Forno, A.R.; Young, W.; et al. Left Bundle Branch Area Pacing for Cardiac Resynchronization Therapy: Results From the International LBBAP Collaborative Study Group. JACC Clin. Electrophysiol. 2021, 7, 135–147. [Google Scholar] [CrossRef]
  12. Jastrzębski, M.; Kiełbasa, G.; Cano, O.; Curila, K.; Heckman, L.; De Pooter, J.; Chovanec, M.; Rademakers, L.; Huybrechts, W.; Grieco, D.; et al. Left bundle branch area pacing outcomes: The multicentre European MELOS study. Eur. Heart J. 2022, 43, 4161–4173. [Google Scholar] [CrossRef] [PubMed]
  13. Keene, D.; Anselme, F.; Burri, H.; Pérez, Ó.C.; Čurila, K.; Derndorfer, M.; Foley, P.; Gellér, L.; Glikson, M.; Huybrechts, W.; et al. Conduction system pacing, a European survey: Insights from clinical practice. Europace 2023, 25, euad019. [Google Scholar] [CrossRef]
  14. Burri, H.; Jastrzebski, M.; Cano, Ó.; Čurila, K.; de Pooter, J.; Huang, W.; Israel, C.; Joza, J.; Romero, J.; Vernooy, K.; et al. EHRA clinical consensus statement on conduction system pacing implantation: Executive summary. Endorsed by the Asia-Pacific Heart Rhythm Society (APHRS), Canadian Heart Rhythm Society (CHRS) and Latin-American Heart Rhythm Society (LAHRS). Europace 2023, 25, 1537–1548. [Google Scholar] [CrossRef]
  15. Wijesuriya, N.; Elliott, M.K.; Mehta, V.; Behar, J.M.; Niederer, S.; Wilkoff, B.L.; Rinaldi, C.A. Transvenous lead extraction in conduction system pacing. Front. Physiol. 2022, 13, 993604. [Google Scholar] [CrossRef]
  16. Kusumoto, F.M.; Schoenfeld, M.H.; Wilkoff, B.; Berul, C.I.; Birgersdotter-Green, U.M.; Carrillo, R.; Cha, Y.M.; Clancy, J.; Deharo, J.C.; Ellenbogen, K.A.; et al. 2017 HRS expert consensus statement on cardiovascular implantable electronic device lead management and extraction. Heart Rhythm 2017, 14, e503–e551. [Google Scholar] [CrossRef]
  17. Garnreiter, J.; Whitaker, P.; Pilcher, T.; Etheridge, S.; Saarel, E. Lumenless pacing leads: Performance and extraction in pediatrics and congenital heart disease. Pacing Clin. Electrophysiol. 2015, 38, 42–47. [Google Scholar] [CrossRef] [PubMed]
  18. Perez, A.A.; Woo, F.W.; Tsang, D.C.; Carrillo, R.G. Transvenous lead extractions: Current approaches and future trends. Arrhythm. Electrophysiol. Rev. 2018, 7, 210–217. [Google Scholar] [CrossRef] [PubMed]
  19. Krainski, F.; Miller, J.P.; Pretorius, V.; Birgersdotter-Green, U. What goes in may need to come out: Considerations in the extraction of a lumenless, fixed-screw permanent pacemaker lead. Heart Rhythm. O2 2020, 1, 160–163. [Google Scholar] [CrossRef]
  20. Vijayaraman, P.; Subzposh, F.A.; Naperkowski, A. Extraction of the permanent His bundle pacing lead: Safety outcomes and feasibility of reimplantation. Heart Rhythm 2019, 16, 1196–1503. [Google Scholar] [CrossRef]
  21. Boczar, K.; Ząbek, A.; Dębski, M.; Gajek, J.; Lelakowski, J.; Małecka, B. Transvenous extraction of His bundle pacing lead: New challenge in the field of lead extraction. Cardiol. J. 2019, 26, 805. [Google Scholar] [CrossRef]
  22. Migliore, F.; Dall’Aglio, P.; Falzone, P.V.; Bertaglia, E.; Zanon, F. Extraction of a very old His bundle pacing lead: A safe and effective procedure? Pacing Clin. Electrophysiol. 2021, 44, 1464–1465. [Google Scholar] [CrossRef] [PubMed]
  23. Boczar, K.; Ząbek, A.; Golińska-Grzybała, K.; Gajek, J.; Holcman, K.; Kostkiewicz, M.; Lelakowski, J. Initial experience with transvenous lead extraction of His bundle pacing leads. Kardiol. Pol. 2023, 81, 1015–1017. [Google Scholar] [CrossRef] [PubMed]
  24. Vijayaraman, P. Extraction of left bundle branch pacing lead. JACC Clin. Electrophysiol. 2020, 6, 903–904. [Google Scholar] [CrossRef] [PubMed]
  25. Ponnusamy, S.S.; Vijayaraman, P. Late dislodgement of left bundle branch pacing lead and successful extraction. J. Cardiovasc. Electrophysiol. 2021, 32, 2346–2349. [Google Scholar] [CrossRef] [PubMed]
  26. Migliore, F.; Aruta, P.; Cecchetto, A.; Iliceto, S.; Gerosa, G.; Catanzariti, D. Extraction of left bundle branch pacing lead: A safe procedure? Europace 2021, 23, 1921. [Google Scholar] [CrossRef]
  27. le Polain de Waroux, J.B.; Wielandts, J.Y.; Gillis, K.; Hilfiker, G.; Sorgente, A.; Capulzini, L.; Geerts, B.; Knecht, S.; Duytschaever, M.; Tavernier, R. Repositioning and extraction of stylet-driven pacing leads with extendable helix used for left bundle branch area pacing. J. Cardiovasc. Electrophysiol. 2021, 32, 1464–1466. [Google Scholar] [CrossRef] [PubMed]
  28. Ayala Valani, L.; Dion, D.; Daneault, B.; Ayala-Paredes, F. Deep septal perforation during left bundle area pacing resulting in left ventricular capture and delayed systemic embolism. Can. J. Cardiol. 2023, 39, 1432–1435. [Google Scholar] [CrossRef] [PubMed]
  29. Agudo, C.A.; Jaén, E.G.; Sánchez, D.J.; Urda, V.C.; Ramos, J.T.; Lozano, I.F. Extraction of a fractured pacemaker lead in the left bundle branch area using a snare via a femoral approach. J. Interv. Card. Electrophysiol. 2023, 66, 239–240. [Google Scholar] [CrossRef] [PubMed]
  30. Bongiorni, M.G.; Burri, H.; Deharo, J.C.; Starck, C.; Kennergren, C.; Saghy, L.; Rao, A.; Tascini, C.; Lever, N.; Kutarski, A.; et al. 2018 EHRA expert consensus statement on lead extraction: Recommendations on definitions, endpoints, research trial design, and data collection requirements for clinical scientific studies and registries: Endorsed by APHRS/HRS/LAHRS. Europace 2018, 20, 1217. [Google Scholar] [CrossRef]
  31. Jacheć, W.; Polewczyk, A.; Polewczyk, M.; Tomasik, A.; Kutarski, A. Transvenous Lead Extraction SAFeTY Score for Risk Stratification and Proper Patient Selection for Removal Procedures Using Mechanical Tools. J. Clin. Med. 2020, 9, 361. [Google Scholar] [CrossRef] [PubMed]
  32. Sidhu, B.S.; Ayis, S.; Gould, J.; Elliott, M.K.; Mehta, V.; Kennergren, C.; Butter, C.; Deharo, J.-C.; Kutarski, A.; Maggioni, A.P.; et al. Risk stratification of patients undergoing transvenous lead extraction with the ELECTRa Registry Outcome Score (EROS): An ESC EHRA EORP European lead extraction ConTRolled ELECTRa registry analysis. Europace 2021, 23, 1462–1471. [Google Scholar] [CrossRef]
  33. Bontempi, L.; Curnis, A.; Della Bella, P.; Cerini, M.; Radinovic, A.; Inama, L.; Melillo, F.; Salghetti, F.; Marzi, A.; Gargaro, A.; et al. The MB score: A new risk stratification index to predict the need for advanced tools in lead extraction procedures. Europace 2020, 22, 613–621. [Google Scholar] [CrossRef]
  34. Bontempi, L.; Vassanelli, F.; Cerini, M.; Inama, L.; Salghetti, F.; Giacopelli, D.; Gargaro, A.; Raweh, A.; Curnis, A. Predicting the difficulty of a transvenous lead extraction procedure: Validation of the LED index. J. Cardiovasc. Electrophysiol. 2017, 28, 811–818. [Google Scholar] [CrossRef]
  35. Mazzone, P.; Tsiachris, D.; Marzi, A.; Ciconte, G.; Paglino, G.; Sora, N.; Sala, S.; Vergara, P.; Gulletta, S.; Della Bella, P. Predictors of advanced lead extraction based on a systematic stepwise approach: Results from a high-volume center. Pacing Clin. Electrophysiol. 2013, 36, 837–844. [Google Scholar] [CrossRef] [PubMed]
  36. Jacheć, W.; Nowosielecka, D.; Ziaja, B.; Polewczyk, A.; Kutarski, A. LECOM (Lead Extraction COMplexity): A New Scoring System for Predicting a Difficult Procedure. J. Clin. Med. 2023, 12, 7568. [Google Scholar] [CrossRef]
  37. Kutarski, A.; Jacheć, W.; Nowosielecka, D.; Polewczyk, A. Unexpected Procedure Difficulties Increasing the Complexity of Transvenous Lead Extraction: The Single Centre Experience with 3721 Procedures. J. Clin. Med. 2023, 15, 2811. [Google Scholar] [CrossRef]
Figure 1. Our “home-made” semi-rigid lead extender using a standard stylet and ligature. Example of the removed HBP leads. General view (A). Extension of lumenless HBP lead with a sharp-angled standard stylet and garter. The system enables easy slipping of a polypropylene sheath onto the lead (B). Scar tissue on ended removed HBP lead (C).
Figure 1. Our “home-made” semi-rigid lead extender using a standard stylet and ligature. Example of the removed HBP leads. General view (A). Extension of lumenless HBP lead with a sharp-angled standard stylet and garter. The system enables easy slipping of a polypropylene sheath onto the lead (B). Scar tissue on ended removed HBP lead (C).
Biomedicines 12 01154 g001
Figure 2. Examples of removed HBP leads with implant durations > 4 years. Scar tissue (AD) is seen on all removed HBP leads. The scarring sometimes stays in place (A,B,D) and sometimes slides off towards the tip of the lead (C).
Figure 2. Examples of removed HBP leads with implant durations > 4 years. Scar tissue (AD) is seen on all removed HBP leads. The scarring sometimes stays in place (A,B,D) and sometimes slides off towards the tip of the lead (C).
Biomedicines 12 01154 g002
Figure 3. Creation of the study groups. TLE—transvenous lead extraction, LV (pacing) lead—a lead designed for pacing of the left ventricle.
Figure 3. Creation of the study groups. TLE—transvenous lead extraction, LV (pacing) lead—a lead designed for pacing of the left ventricle.
Biomedicines 12 01154 g003
Figure 4. Lead survival and patient survival. Lead survival—time since lead implantation to extraction due to infection (A), lead failure (B), and patient survival after CIED extraction (C).
Figure 4. Lead survival and patient survival. Lead survival—time since lead implantation to extraction due to infection (A), lead failure (B), and patient survival after CIED extraction (C).
Biomedicines 12 01154 g004
Table 1. Characteristics of the study groups and main indications for lead extraction.
Table 1. Characteristics of the study groups and main indications for lead extraction.
Patient-Related Risk Factors and Coexisting Indications for TLE (in 3897 Patients)“Young”
(≤40 mths) His Lead
Extraction
“Old”
(>40 mths) His Lead
Extraction
“Young”
(≤44 mths) LV Lead
Extraction
“Old”
(>44 mths) LV Lead
Extraction
All TLEs with LV Lead
Extraction
All TLEs with His Lead
Extraction
TLE
of Other than His and LV Leads
Group number/number of patientsGroup 1
n = 13
mean ± SD
n (%)
Group 2
n = 14
mean ± SD
n (%)
P: 1 vs. 2
Group 3
n = 127
mean ± SD
n (%)
Group 4
n = 126
mean ± SD
n (%)
P: 3 vs. 4
Group 5
n = 253
mean ± SD
n (%)
Group 6
n = 27
mean ± SD
n (%)
P: 5 vs. 6
Group 7
n = 3617
mean ± SD
n (%)
5 vs. 7
6 vs. 7
Patient age during TLE [years]68.38
± 9.24
72.50
± 7.43
p = 0.131
68.43
± 10.58
69.30
± 9.34
p = 0.560
68.87
± 9.97
70.52
± 8.45
p = 0.603
65.84
± 16.05
p = 0.370
p = 0.167
Patient age at first system implantation [years]64.92
± 9.72
63.64
± 8.03
p = 0.961
64.57
± 11.49
61.60
± 9.58
p = 0.012
63.09
± 10.66
64.26
± 8.73
p = 0.971
57.13
± 17.51
p = 0.092
p = 0.001
Female3
(23.08)
5
(35.71)
p = 0.767
27
(21.26)
26
(20.63)
p = 0.974
53
(20.95)
8
(29.63)
p = 0.428
1424
(39.37)
p = 0.001
p = 0.404
Coronary heart disease10
(76.93)
6
(42.86)
p = 0.159
63
(49.67)
55
(57.47)
p = 0.410
118
(46.64)
15
(55.56)
p = 0.498
2041
(56.43)
p = 0.003
p = 0.917
NYHA FC III or IV 4
(30.77)
8
(57.14)
p = 0.505
63
(49.61)
49
(38.89)
p = 0.112
112
(44.27)
12
(44.44)
p = 1.000
491
(13.58)
p = 0.001
p = 0.001
LVEF [%]43.31
± 15.90
44.14
± 11.03
p = 0.699
33.80
± 13.95
33.05
± 12.55
p = 0.929
33.43
± 13.25
43.74
± 13.33
p = 0.001
50.65
± 14.82
p = 0.009
p = 0.001
Charlson co-morbidity
index [points]
5.77
± 3.42
6.57
± 3.99
p = 0.980
6.54
± 3.91
5.99
± 3.53
p = 0.403
6.26
± 3.73
6.19
± 3.68
p = 0.972
4.81
± 3.68
p = 0.030
p = 0.001
PADIT score [points]4.69
± 3.38
3.07
± 2.84
p = 0.230
5.52
± 2.81
5.98
± 3.04
p = 0.154
5.75
± 2.93
3.85
± 3.16
p = 0.003
3.54
± 2.82
p = 0.648
p = 0.001
Main indications for TLE—(primary/predominant)
Infective endocarditis with or without pocket infection4
(30.77)
5
(35.71)
p = 0.892
60
(47.24)
63
(50.00)
p = 0.755
123
(48.62)
9
(33.33)
p = 0.190
722
(19.96)
p = 0.001
p = 0.309
Local (isolated) pocket infection2
(15.38)
0
(0.00)
p = 0.430
20
(15.75)
14
(11.11)
p = 0.370
34
(13.44)
2
(7.41)
p = 0.130
330
(9.12)
p = 0.001
p = 0.835
Mechanical lead damage (electrical failure)2
(15.38)
2
(14.29)
p = 0.644
9
(7.09)
21
(16.67)
p = 0.031
30
(11.86)
4
(14.81)
p = 0.891
1026
(28.37)
p = 0.001
p = 0.179
Lead dysfunction (exit/
entry block, dislodgement, perforation, extracardiac pacing)
5
(38.46)
4
(28.57)
p = 0.892
35
(27.56)
14
(11.11)
p = 0.002
49
(19.37)
9
(33.33)
p = 0.164
809
(22.37)
p = 0.025
p = 0.516
Change of pacing mode/upgrading, downgrading and other non-
infectious indications *
0
(0.00)
2
(14.29)
p = 0.496
3
(2.36)
14
(11.11)
p = 0.012
17
(6.72)
3
(11.11)
p = 0.636
730
(20.18)
p = 0.001
p = 0.201
TLE—transvenous lead extraction, LV (pacing) lead—a lead designed for pacing of the left ventricle, mths—months, n—number of patients/procedures, SD—standard deviation, NYHA FC—New York Heart Association (functional class), LVEF—left ventricular ejection fraction, * abandoned lead/prevention of abandonment (atrial fibrillation, redundant leads), threatening/potentially threatening lead (loops, free ends, left heart, lead-derived tricuspid valve defect), other (magnetic resonance imaging, cancer, painful pocket, loss of indications for pacing/implantable cardioverter defibrillator), and re-established venous access (symptomatic occlusion, superior vena cava syndrome, lead replacement/upgrading).
Table 2. System-related, history of pacing-related, and procedure-related risk factors for major complications and procedure complexity.
Table 2. System-related, history of pacing-related, and procedure-related risk factors for major complications and procedure complexity.
System-Related Risk
Factors for TLE
Complexity and Major Complications of TLE
“Young”
(≤40 mths) His Lead
Extraction
“Old”
(>40 mths) His Lead
Extraction
“Young”
(≤44 mths) LV Lead
Extraction
“Old”
(>44 mths) LV Lead
Extraction
All TLEs with LV Lead
Extraction
All TLEs with His Lead
Extraction
TLE
of Other than His and LV Leads
Number of
patients/group number
Group 1
n = 13
mean ± SD
n (%)
Group 2
n = 14
mean ± SD
n (%)
P: 1 vs. 2
Group 3
n = 127
mean ± SD
n (%)
Group 4
n = 126
mean ± SD
n (%)
P: 3 vs. 4
Group 5
n = 253
mean ± SD
n (%)
Group 6
n = 27
mean ± SD
n (%)
P: 5 vs. 6
Group 7
n = 3617
mean ± SD
n (%)
5 vs. 7
6 vs. 7
System and history of pacing
Type of CIED
Pacemaker 7
(53.85)
11
(78.57)
p = 0.341
35
(27.56)
38
(30.16)
p = 0.571
73
(28.85)
18
(66.67)
p < 0.001
2678
(74.04)
p = 0.001
p = 0.516
Implantable cardioverter defibrillator0
(0.00)
0
(0.00)
NA
0
(0.00)
3
(2.38)
p = 0.243
3
(1.19)
0
(0.00)
p = 0.679
845
(23.36)
p = 0.001
p = 0.008
Cardiac
resynchronisation
cardioverter defibrillator
6
(46.15)
3
(21.43)
p = 0.341
92
(72.44)
85
(67.46)
p = 0.467
177
(69.96)
9
(33.33)
p = 0.001
94
(2.60)
p < 0.001
p < 0.001
Dwell time of oldest
extracted lead [months]
41.31
± 35.00
105.93
± 42.40
<0.001
47.87
± 45.20
93.24
± 44.32
p < 0.001
70.46
± 50.13
74.81
± 50.47
p = 0.576
105.19
± 77.55
p = 0.059
p = 0.001
Global lead dwell time before TLE [years]6.24
± 5.33
22.01
± 9.33
<0.001
8.79
± 7.78
21.68
± 9.90
p < 0.001
15.21
± 10.98
14.41
± 11.00
p = 0.680
15.51
± 13.15
p = 0.833
p = 0.369
HBP and LVP lead
implant duration [months]
13.15
± 8.38
92.93
± 38.22
<0.001
21.94
± 16.09
78.45
± 26.98
p < 0.001
50.20
± 35.96
54.52
± 49.12
p = 0.568
NA
NA
NA
Abandoned lead before TLE1
(7.69)
1
(7.14)
p = 0.496
14
(11.02)
20
(15.87)
p = 0.344
34
(13.44)
2
(7.41)
p = 0.557
382
(10.56)
p = 0.118
p = 0.828
Number of leads in the heart before TLE before TLE2.38
± 0.87
2.64
± 0.50
p = 0.590
2.97
± 0.62
3.10
± 0.58
p = 0.175
3.04
± 0.60
2.52
± 0.70
p = 0.003
1.87
± 0.68
p = 0.001
p = 0.001
Number of procedures before lead extraction1.77
± 0.60
2.08
± 0.76
p = 0.294
1.91
± 1.33
2.27
± 0.98
p < 0.001
2.09
± 1.18
1.92
± 0.69
p = 0.830
1.84
± 1.07
p = 0.146
p = 0.001
Scores for evaluation of risk factors for major complications or procedure complexity
Safety score estimated risk of MC (risk of MC) [%]0.60
± 0.50
1.94
± 1.96
p = 0.019
0.98
± 2.72
2.08
± 2.22
p < 0.001
1.54
± 2.53
1.33
± 1.61
p = 0.837
1.73
± 2.98
p = 0.682
p = 0.697
2 and 3 EROS score
[risk of MC]
4
(36.36)
4
(50.00)
p = 0.767
65
(51.18)
68
(53.97)
p = 0.853
133
(52.57)
16
(59.26)
p = 0.646
1399
(38.70)
p = 0.001
p = 0.047
MB score (need for
advanced tools) [points]
1.85
± 1.46
3.21
± 0.58
p = 0.024
2.16
± 1.39
3.45
± 0.93
p < 0.001
2.80
± 1.35
2.56
± 1.28
p = 0.260
2.58
± 1.24
p = 0.782
p = 0.005
LED index (predicted fluoroscopy time) [points]5.54
± 3.10
10.79
± 3.36
p = 0.001
6.28
± 4.16
10.32
± 3.68
p < 0.001
8.29
± 4.42
8.26
± 4.15
p = 0.961
10.23
± 6.53
p = 0.193
p = 0.001
Advanced LE (ALET—need for advanced TLE techniques) scale
(3 and 4 points)
6
(46.15)
6
(42.86)
p = 0.830
68
(53.64)
96
(76.19)
p < 0.001
164
(64.82)
12
(44.44)
p = 0.061
1217
(33.65)
p = 0.001
p = 0.328
LECOM score (expected procedure complexity) [%]16.19
± 10.30
22.39
± 15.70
p = 0.843
23.33
± 18.49
30.88
± 19.11
p < 0.001
27.09
± 19.14
19.41
± 13.50
p = 0.026
19.81
± 18.29
p = 0.391
p = 0.001
TLE-related potential risk factors for major complications and procedure complexity
Number of extracted leads per patient2.23
± 0.83
2.14
± 0.86
p = 0.393
2.46
± 1.01
2.75
± 0.88
p = 0.044
2.60
± 0.96
2.19
± 0.83
p = 0.012
1.59
± 0.66
p = 0.004
p = 0.001
Extraction of abandoned lead(s) (any)1
(7.69)
1
(7.14)
p = 0.496
13
(10.24)
18
(14.29)
p = 0.453
31
(12.25)
2
(7.41)
p = 0.668
351
(9.70)
p = 0.228
p = 0.940
Extraction of passive-
fixation lead
(excluding LV lead)
7
(53.85)
10
(71.43)
p = 0.585
125
(98.43)
124
(98.41)
p = 0.620
249
(98.42)
17
(62.96)
p < 0.001
1985
(54.88)
p = 0.001
p = 0.518
Age of extracted HBP/LVP leads [months]13.15
± 8.38
92.93
± 38.22
p = 0.000
21.94
± 16.09
78.45
± 26.98
p < 0.001
50.20
± 35.96
54.52
± 49.12
p = 0.568
NA
NA
NA
Oldest extracted lead per patient [months]41.31
± 35.00
105.93
± 42.40
<0.001
45.68
± 44.85
90.74
± 41.08
p < 0.001
68.12
± 48.50
74.81
± 50.47
p = 0.458
103.49
± 76.54
p = 0.076
p = 0.001
TLE—transvenous lead extraction, LV (pacing) lead—a lead designed for pacing of the left ventricle, mths—months, n—number of patients/procedures, SD—standard deviation HBP –His Bundle Pacing (lead), LVP—Left Ventricular Pacing (lead), CIED—cardiac implantable electronic devices, MC—major (TLE) complications, EROS—ELECTRa Registry Outcome Score, MB (score)—Mazzone-Bontempi (score), LED (index)—Lead Extraction Difficulty (index), ALET (score)—Advanced Lead Extraction Techniques (score), LECOM score—Lead Extraction COMplexity scoring system, NA—non-applicable.
Table 3. Extraction procedure complexity, major complications, and long-term outcomes.
Table 3. Extraction procedure complexity, major complications, and long-term outcomes.
Patient-Related Risk
Factors and Coexisting
Indications for TLE
(in 3897 Patients)
“Young”
(≤40 mths) His Lead
Extraction
“Old”
(>40 mths) His Lead
Extraction
“Young”
(≤44 mths) LV Lead
Extraction
“Old”
(>44 mths) LV Lead
Extraction
All TLEs with LV Lead
Extraction
All TLEs with His Lead
Extraction
TLE
of Other than His and LV Leads
Group number/number of patientsGroup 1
n = 13
mean ± SD
n (%)
Group 2
n = 14
mean ± SD
n (%)
P: 1 vs. 2
Group 3
n = 127
mean ± SD
n (%)
Group 4
n = 126
mean ± SD
n (%)
P: 3 vs. 4
Group 5
n = 253
mean ± SD
n (%)
Group 6
n = 27
mean ± SD
n (%)
P: 5 vs. 6
Group 7
n = 3617
mean ± SD
n (%)
5 vs. 7
6 vs. 7
Procedure duration (sheath-to-sheath) [minutes]7.85
± 3.16
19.43
± 17.82
p = 0.057
14.35
± 17.97
25.38
± 27.56
p < 0.001
19.85
± 23.85
13.85
± 14.08
p = 0.052
14.71
± 23.25
p = 0.354
p = 0.001
Mean time of single lead extraction (sheath-to-sheath/number of
extracted leads) [minutes]
3.87
± 1.90
8.57
± 6.97
p = 0.015
5.85
± 5.94
9.31
± 9.58
p < 0.001
7.57
± 8.14
6.31
± 5.63
p = 0.452
9.02
± 13.00
p = 0.161
p = 0.043
Time of single HBP/LVP lead extraction [minutes]2.15
± 2.23
4.14
± 1.96
p = 0.021
2.07
± 2.24
4.05
± 5.19
p < 0.001
3.05
± 4.10
3.19
± 2.29
p = 0.868
NA
NA
NA
Removal of HBP/LVP lead with simple traction 4
(30.77)
0
(0.00)
p = 0.025
63
(49.61)
10
(7.94)
p < 0.001
73
(28.97)
4
(14.81)
p = 0.185
NA
NA
NA
Extraction of HBP/LVP lead using mechanical
dilation
9
(69.23)
14
(100.0)
p = 0.088
64
(50.39)
116
(91.27)
p < 0.001
179
(70.75)
23
(85.19)
p = 0.127
NA
NA
NA
Number of unexpected procedure difficulties (“technical problems”)
per patient
0.00
± 0.00
1.75
± 0.96
p = 0.045
1.57
± 0.85
1.50
± 0.67
p = 0.049
1.52
± 0.72
1.75
± 0.96
p = 0.800
1.45
0.77
p = 0.753
p = 0.908
CID-TLE score [points]0.000.93
± 1.33
p = 0.004
0.43
± 0.94
0.98
± 1.23
p < 0.001
0.70
± 1.12
0.48
± 1.05
p = 0.620
0.54
± 1.12
p = 0.773
p = 0.012
CID-TLE score ≥ 2 points 0
(0.00)
5
(35.71)
p = 0.059
22
(17.32)
49
(38.89)
p < 0.001
71
(28.06)
5
(18.52)
p = 0.405
662
(18.30)
p = 0.001
p = 0.825
Use of additional tools
Evolution (old and new) or TightRail0
(0.00)
1
(7.14)
p = 0.970
1
(0.79)
2
(1.59)
p = 1.000
3
(1.19)
1
(3.70)
p = 0.845
61
(1.69)
p = 0.001
p = 0.952
Metal sheath, lasso-basket catheter0
(0.00)
3
(21.43)
p = 0.247
8
(6.30)
22
(17.46)
p = 0.011
30
(11.86)
3
(11.11)
p = 0.842
499
(13.80)
p = 0.440
p = 0.902
TLE efficacy and complications
Major complications (any)0
(0.00)
0
(0.00)
NA
1
(0.79)
1
(0.79)
p = 0.481
2
(0.79)
0
(0.00)
p = 0.460
78
(2.16)
p = 0.212
p = 0.917
Haemopericardium0
(0.00)
0
(0.00)
NA
1
(0.79)
0
(0.00)
p = 0.997
1
(0.40)
0
(0.00)
p = 0.171
45
(1.24)
p = 0.366
p = 0.771
Rescue cardiac surgery0
(0.00)
0
(0.00)
NA
1
(0.79)
1
(0.79)
p = 1.000
2
(0.79)
0
(0.00)
p = 0.460
38
(1.05)
p = 0.941
p = 0.678
Tricuspid valve damage during TLE (severe)0
(0.00)
0
(0.00)
NA
0
(0.00)
0
(0.00)
NA
0
(0.00)
0
(0.00)
NA
25
(0.69)
p = 0.357
p = 0.461
Death, procedure-related (intra-, post-procedural)0
(0.00)
0
(0.00)
NA
1
(0.79)
0
(0.00)
p = 0.997
1
(0.40)
0
(0.00)
p = 0.171
5
(0.14)
p = 0.859
p = 0.847
Partial radiographic
success (retained tip
or <4 cm lead fragment)
0
(0.00)
0
(0.00)
NA
0
(0.00)
8
(6.35)
p = 0.012
8
(3.16)
0
(0.00)
p = 0.742
145
(4.01)
p = 0.616
p = 0.573
Procedural success13
(100.0)
14
(100.0)
NA
125
(98.43)
118
(93.65)
p = 0.104
243
(96.05)
27
(100.0)
p = 0.613
3444
(95.22)
p = 0.654
p = 0.978
Mortality after TLE (death for each time interval)
All model log rank p < 0.001
Survivors during
follow-up
7
(53.85)
8
(57.14)
p = 0.830
47
(37.01)
59
(46.83)
p = 0.146
106
(41.90)
15
(55.56)
p = 0.247
2264
(62.59)
p = 0.001
p = 0.580
Deaths during follow-up6
(45.15)
6
(42.86)
p = 0.830
80
(62.99)
67
(53.18)
p = 0.146
147
(58.10)
12
(44.44)
p = 0.247
1353
(37.41)
p = 0.001
p = 0.580
Perioperative death
(first two days)
0
(0.00)
0
(0.00)
NA
1
(0.79)
0
(0.00)
p = 0.997
1
(0.40)
0
(0.00)
p = 0.171
13
(0.36)
p = 0.653
p = 0.191
Intra-hospital death
(3rd–30th day)
2
(15.38)
0
(0.00)
p = 0.430
5
(3.94)
4
(3.17)
p = 0.990
9
(3.56)
2
(7.41)
p = 0.647
45
(1.24)
p = 0.006
p = 0.049
Log rank:
p = 0.132
Log rank:
p = 0.522
Log rank:
p = 0.325
Log rank:
p = 0.005
p = 0.010
Death within the first year of follow-up (31–365 days)3
(23.08)
1
(7.14)
p = 0.534
23
(18.11)
19
(15.08)
p = 0.632
42
(16.60)
4
(14.81)
p = 0.972
217
(6.00)
p = 0.001
p = 0.132
One-year follow-up Log rank:
p = 0.022
Log rank:
p = 0.349
Log rank:
p = 0.695
Log rank:
p < 0.001
p = 0.002
Death at 1 to 3 years of follow-up (366–1095 days)1
(7.69)
3
(21.43)
p = 0.644
22
(17.32)
21
(16.67)
p = 0.977
43
(17.00)
4
(14.81)
p = 0.986
306
(8.46)
p = 0.000
p = 0.405
Three-year follow-up Log rank: p = 0.062 Log rank: p = 0.374 Log rank: p = 0.654Log rank:
p < 0.001
p < 0.001
Late death, at >3 years (>1095 days) of follow-up0
(0.00)
2
(14.29)
p = 0.496
29
(22.83)
23
(18.25)
p = 0.456
52
(20.55)
2
(7.41)
p = 0.165
772
(21.34)
p = 0.828
p = 0.135
Death during follow-up Log rank:
p = 0.099
Log rank:
p = 0.354
Log rank:
p = 0.780
Log rank:
p < 0.001
p < 0.001
Follow-up [days]940.8
± 1204
1334
± 1088
p = 0.148
1458
± 1376
1407
± 1239
p = 0.882
1153
± 1135
1432
± 1307
p = 0.337
2122
± 1469
p < 0.001
p < 0.001
The length of the FU
period for survivors [days]
1303.86
± 1451.79
1559
± 1346
p = 1.000
2095.11
± 1548.34
1973
± 1361
p = 0.629
2027
± 1441
1440
± 1351
p = 0.462
2481
± 1516.5
p < 0.001
p < 0.001
The length of the FU
period for non-survivors [days]
284.3
± 314.7
906.0
± 646.0
p = 0.020
1072
± 1137
929.0
± 919.8
p = 0.832
1007
± 1043
595.2
± 583.2
p = 0.243
1578
± 1251.8
p < 0.001
p = 0.003
TLE—transvenous lead extraction, LV (pacing) lead—a lead designed for pacing of the left ventricle, mths—months, n—number of patients/procedures, SD—standard deviation, HBP—His Bundle Pacing (lead), LVP—Left Ventricular Pacing (lead), CID-TLE—Complex Indicator of the Difficulty of the TLE, NA—non-applicable.
Table 4. Differences between survivors and non-survivors of His lead extraction.
Table 4. Differences between survivors and non-survivors of His lead extraction.
His Pacing Lead Extraction Group
Potential Risk Factors
for Shorter Survival
Survivors
n = 15
Non-
Survivors
n = 12
p
Mann–Whitney U Test
Chi2
Female7 (46.67)2 (16.67)0.218
Patient age [years]71.25 ± 7.8269.83 ± 10.140.707
Left ventricular ejection fraction [%]50.18 ± 12.5540.33 ± 13.710.117
Congestive heart failure5 (33.33)7 (58.33)0.553
CHA2DS2–VASc score [points]3.50 ± 1.513.25 ± 1.290.729
Charlson co-morbidity index [points]6.33 ± 4.035.25 ± 3.310.817
PADIT [points]3.33 ± 3.034.33 ± 3.550.525
Oldest extracted lead per patient [years]7.65 ± 4.264.40 ± 3.180.057
Infectious indications for TLE5 (3.33)6 (50.00)0.630
ICD lead presence before TLE1 (6.67)6 (60.00)0.035
CHA2DS2–VASc—(Congestive heart failure, Hypertension, Age 75 years and older, Diabetes, Stroke, Vascular disease, Age 65 to 74 years, Sex Category (female))—ischemic risk stroke score in patients with atrial fibrillation, PADIT—(Prevention of Arrhythmia Device Infection Trial) risk score of device infection, TLE—transvenous lead extraction, ICD—implantable cardioverter defibrillator.
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content.

Share and Cite

MDPI and ACS Style

Stefańczyk, P.; Jacheć, W.; Kutarski, A.; Dąbrowski, P.; Głowniak, A.; Nowosielecka, D. Extraction of His Bundle Pacing Lead: More Difficult than Coronary Sinus Lead Extraction: An Analysis of 3897 Lead Extraction Procedures Including 27 His and 253 Coronary Sinus Lead Removals. Biomedicines 2024, 12, 1154. https://doi.org/10.3390/biomedicines12061154

AMA Style

Stefańczyk P, Jacheć W, Kutarski A, Dąbrowski P, Głowniak A, Nowosielecka D. Extraction of His Bundle Pacing Lead: More Difficult than Coronary Sinus Lead Extraction: An Analysis of 3897 Lead Extraction Procedures Including 27 His and 253 Coronary Sinus Lead Removals. Biomedicines. 2024; 12(6):1154. https://doi.org/10.3390/biomedicines12061154

Chicago/Turabian Style

Stefańczyk, Paweł, Wojciech Jacheć, Andrzej Kutarski, Paweł Dąbrowski, Andrzej Głowniak, and Dorota Nowosielecka. 2024. "Extraction of His Bundle Pacing Lead: More Difficult than Coronary Sinus Lead Extraction: An Analysis of 3897 Lead Extraction Procedures Including 27 His and 253 Coronary Sinus Lead Removals" Biomedicines 12, no. 6: 1154. https://doi.org/10.3390/biomedicines12061154

Note that from the first issue of 2016, this journal uses article numbers instead of page numbers. See further details here.

Article Metrics

Back to TopTop