A Novel Microfluidic Formaldehyde Microanalyser for Continuous Real-Time Monitoring in Indoor Air: Analytical Development and Validation †

,
and
1
Institute of Chemistry and Processes for Energy, Environment and Health (ICPEES), University of Strasbourg and CNRS (UMR 7515), 25 rue Becquerel, 67087 Strasbourg, France
2
In’Air Solutions, 1 rue Blessig, 67084 Strasbourg, France
*
Author to whom correspondence should be addressed.
Presented at the 5th International Symposium on Sensor Science (I3S 2017), Barcelona, Spain, 27–29 September 2017.
Formaldehyde is a major and harmful pollutant of indoor air due to its multiple sources and its carcinogenic effect. This work reports the development of a novel analytical method based on microfluidic technologies for the detection of low airborne Formaldehyde concentrations, representative of those found in indoor air, i.e., 10–100 µg m−3. The new analytical technique operates as follows: (1) gas sampling; (2) gaseous Formaldehyde uptake into the aqueous solution using an annular gas/liquid flow at room temperature; (3) derivatization reaction with acetylacetone solution at 65 °C producing 3,5-Diacetyl-1,4-dihydrolutidine (DDL) and (4) fluorimetric DDL detection.
Laboratory experiments were performed to determine the experimental conditions permitting to obtain a stable annular flow, i.e., gas to liquid flow rate ratios greater than 1000. From liquid and gas calibrations, an uptake yield of 100% and a detection limit of 1 µg m−3 were determined. Finally, our portable instrument is fully controlled by homemade software and has a response time of 10 min, a temporal resolution of 2 s and an autonomy of 100 h with 100 mL reagent. Finally, this formaldehyde microanalyser was then deployed during several field campaigns and compared with the ISO 16000-3 reference method, i.e., the active sampling on DNPH cartridges.