Next Article in Journal
Screening for Lactation Risk Factors in Pregnancy
Previous Article in Journal
Low-Voltage Tri-Electrode Electrostatic Actuator Using Solid Gap-Spacing Materials
 
 
Font Type:
Arial Georgia Verdana
Font Size:
Aa Aa Aa
Line Spacing:
Column Width:
Background:
Abstract

MEMS Valves with Molecular Flow Regime Orifices †

by
Alvise Bagolini
1,*,
Raffaele Correale
2,
Antonino Picciotto
3 and
Leandro Lorenzelli
1
1
MicroSystems Technology Group, CMM, Fondazione Bruno Kessler, 38123 Trento, Italy
2
Nanotech Analysis SRL, Corso Umberto 65, 10128 Torino, Italy
3
Micro-Nano Facility Group, CMM, Fondazione Bruno Kessler, 38123 Trento, Italy
*
Author to whom correspondence should be addressed.
Presented at the XXXV EUROSENSORS Conference, Lecce, Italy, 10–13 September 2023.
Proceedings 2024, 97(1), 73; https://doi.org/10.3390/proceedings2024097073
Published: 21 March 2024

Abstract

:
In this work, a novel, silicon-based micro-electromechanical valve that includes a submicrometric orifice and can operate at pressure gradients of 1 bar was used to enhance sampling for gas chromatograph mass spectrometers. The valve is based on a membrane-in-membrane design and operates with thermomechanical actuation. It includes a pin to enable self-cleaning. Prototypes were fabricated and preliminary testing was performed.

1. Introduction

Great improvement in gas chromatograph mass spectrometers (GC-MSs) is possible using nanometer-scale orifices [1] as sampling points and smart gas interfaces towards atmospheric pressure. At high pressure gradients, orifices with diameters of about 100 nm operate under a molecular flow regime (MFR) [2]. A flow through conductance in an MFR does not give origin to gas collective motions, preventing condensation, chemical reactions and clogging [3]. Such inlet diameters enable gas flows in the range from 10−5 to 10−7 mbar L/s that guarantee the same concentrations of the external gases into the ionization chamber [4,5]. Such devices will be exposed to a 1 bar pressure gradient, with the sensing chamber on one side (in vacuum) and the analyte-containing ambient (at atmospheric pressure) on the other. A membrane with submicron orifices capable of consistently withstanding 1 bar pressure gradient was developed by the authors using a novel membrane-in-membrane structure [6].
We hereby report the development of an analogous nanoscale orifice membrane with a thermomechanically actuated valve to enable its operation and self-cleaning.

2. Materials and Methods

The device structure (Figure 1) was modified in order to include the thermomechanical valve component, but the same membrane-in-membrane approach was adopted. An inductively coupled, plasma-enhanced vapor deposition (ICP-CVD) silicon nitride film was developed for this device, demonstrating low residual stress after annealing. Based on the simulation results, a batch of prototype devices was designed and microfabricated on 6-inch SEMI standard silicon wafers. An example of the fabricated devices is reported in Figure 2.

3. Results

The first fabrication batch was successfully completed. The devices were bonded on copper circular substrates having a passing hole. These substrates provide a mounting support in the vacuum system, and are equipped with a flat cable connection to the MEMS device. Thermoelectric valve actuation was demonstrated at ambient pressure, and resistance towards the pressure gradient was demonstrated under a full vacuum by mounting the device in a quadrupole mass filter spectrometer system that operated at a background pressure ranging from 10−9 to 10−10 mbar.

Author Contributions

Conceptualization, A.B. and R.C.; Methodology, A.B., R.C. and L.L.; Material Characterization, A.B. and A.P.; Fabrication and testing, A.B. and R.C. All authors have read and agreed to the published version of the manuscript.

Funding

This research was funded by Nanotech Analysis SRL.

Institutional Review Board Statement

Not applicable.

Informed Consent Statement

Not applicable.

Data Availability Statement

The raw data supporting the conclusions of this article will be made available by the authors on request.

Conflicts of Interest

The affiliated company Nanotech Analysis SRL has no conflict of interest with this paper.

References

  1. Kurth:, M.L.; Gramotnev, D.K. Nanofluidic delivery of molecules: Integrated plasmonic sensing with nanoholes. Microfluid. Nanofluidics 2013, 14, 743–751. [Google Scholar] [CrossRef]
  2. Lafferty, J.M. Foundation of Vacuum Science and Technology; John Wiley & Sons, Inc.: Hoboken, NJ, USA, 1998. [Google Scholar]
  3. Correale, R. A new generation of nanotechnological instruments: Application to Homeland Security domain. In Proceedings of the 8th National Conference on Cyber Warfare (CWC), Information Warfare vs Cyber Warfare: Differences and Applications, Milan, Italy, 24 November 2017. [Google Scholar]
  4. Correale, R. A nanotechnology approach for the realization of a novel generation of miniature and portable analytical instruments. In Proceedings of the Euro Nano Forum 2017 Conference, Valletta, Malta, 21–23 June 2017. [Google Scholar]
  5. Tosi, P.; Correale, R.; Lu, W.; Bassi, D. Production of the molecular Dication ArN2+ ions in the Reaction Ar2+ + N2. Phys. Rev. Lett. 1999, 2, 450. [Google Scholar] [CrossRef]
  6. Bagolini, A.; Correale, R.; Picciotto, A.; Di Lorenzo, M.; Scapinello, M. MEMS Membranes with Nanoscale Holes for Analytical Applications. Membranes 2021, 11, 74. [Google Scholar] [CrossRef] [PubMed]
Figure 1. Membrane and bridge valve section.
Figure 1. Membrane and bridge valve section.
Proceedings 97 00073 g001
Figure 2. SEM image of the fabricated membrane with a bridge valve.
Figure 2. SEM image of the fabricated membrane with a bridge valve.
Proceedings 97 00073 g002
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content.

Share and Cite

MDPI and ACS Style

Bagolini, A.; Correale, R.; Picciotto, A.; Lorenzelli, L. MEMS Valves with Molecular Flow Regime Orifices. Proceedings 2024, 97, 73. https://doi.org/10.3390/proceedings2024097073

AMA Style

Bagolini A, Correale R, Picciotto A, Lorenzelli L. MEMS Valves with Molecular Flow Regime Orifices. Proceedings. 2024; 97(1):73. https://doi.org/10.3390/proceedings2024097073

Chicago/Turabian Style

Bagolini, Alvise, Raffaele Correale, Antonino Picciotto, and Leandro Lorenzelli. 2024. "MEMS Valves with Molecular Flow Regime Orifices" Proceedings 97, no. 1: 73. https://doi.org/10.3390/proceedings2024097073

Article Metrics

Back to TopTop