Next Article in Journal
Infrared Wavefront Sensing for Adaptive Optics Assisted Galactic Center Observations with the VLT Interferometer and GRAVITY: Operation and Results
Previous Article in Journal
Open-Source Digitally Replicable Lab-Grade Scales
Previous Article in Special Issue
A Software Improvement Technique for Platinum Resistance Thermometers
 
 
Font Type:
Arial Georgia Verdana
Font Size:
Aa Aa Aa
Line Spacing:
Column Width:
Background:
Article

CT-Dose Measurement of the Spinal Cord Region Using XR-QA2 Radiochromic Films and TLD 100H Dosimeters

by
Martina Pace
1,2,
Giuseppe Stella
1,2,*,
Letizia Barone Tonghi
2,3,
Stefania Mazzaglia
1,2 and
Anna Maria Gueli
1,2
1
PH3DRA labs, Dipartimento di Fisica e Astronomia “Ettore Majorana”, Umiversità di Catania & INFN-CHNet Sez CT, via Santa Sofia 64, 95123 Catania, Italy
2
Scuola di Specializzazione in Fisica Medica, Dipartimento di Scienze Mediche Chirurgiche e Tecnologie Avanzate ‘‘G. F. Ingrassia’’, Scuola Facoltà di Medicina, Via Santa Sofia 78, 95123 Catania, Italy
3
P.O. Garibaldi Nesima, Via Palermo, 636, 95122 Catania CT, Italy
*
Author to whom correspondence should be addressed.
Instruments 2020, 4(3), 19; https://doi.org/10.3390/instruments4030019
Submission received: 17 December 2019 / Revised: 13 June 2020 / Accepted: 24 June 2020 / Published: 3 July 2020

Abstract

:
The European Directive 2013/59/Euratom focuses on the radiation protection and patient dosimetry. In particular, the dose absorbed by serial organs such as the spinal cord due to diagnostic exams plays a crucial role in the appraisal of medical exposure. In this work, a comparison between dose measurements performed with GafchromicTM XR-QA2 and TLD 100H is presented. The dosimeters—after a calibration procedure through an X-ray tube—were placed within the thorax region of an anthropomorphic phantom corresponding to the spinal cord area exposed to a thoracic CT procedure. The mean-dose value was measured with GafchromicTM first, and it was then compared to the dose value obtained with TLDs, resulting in a good agreement between the two dosimetric methodologies. Additionally, the results showed that—due to the usage of the automatic exposure control (EC) system—the discrepancy between the two methods is proportional to the current output of the CT system.

1. Introduction

The evolution and improvement of imaging techniques has contributed to increasing the number of examinations carried out during the hospital routine: clinical imaging is a fundamental tool employed by specialists in the diagnosis of various diseases. However, wide use of these techniques has led to an increase in the exposure to ionizing radiation, highlighting the dangers and the associated risks. As of today, one of the most-employed clinical imaging techniques is the Computed Tomography (CT) [1].
Over the past years, considerable effort has been put into addressing the assessment the organ doses for many different types of CT scanners (cone- and fan-beam) using different CT scan protocols and dosimetry systems such as, GafchromicTM and TLD crystals [2,3].
Although TLD dosimeters can be used to obtain punctual information of the absorbed dose, they do not allow for accurate measurements of strong localized dose gradients—especially when multi-slice scanners are employed [4]. The TLD 100H (LiF: Mg, Cu, P) dosimeters are the most-used in clinical dosimetry and in applications where high sensitivity at low doses is needed.
The TLD-100H dosimeters have been characterized in terms of glow curve deconvolution, energy and angular dependence [5,6,7]. In fact, it has been found that TLD-100H dosimeters are characterized by a linear response inside the diagnostic energy interval and their sensitivity variation is under 3% at tube potentials ranging from 40 kVp to 125 kVp [5].
Radiochromic films are employed to obtain two-dimensional dose distributions with high-resolution [2,8,9,10]. The GafchromicTM XR-QA2 is the most recent radiochromic film used to carry out measurements of CT dose distribution [3]. In general, the XR-QA model allows high-resolution slice-by-slice dose distribution measurements inside an organ volume, as well as spatial- and punctual-dose measurements, providing a potential tool to perform 2D- and 3D-CT dosimetry [2,11].
Recent studies have shown that the difference in dose response between XR-QA films and TLD arise when the film surface sensitive area is parallel to the central axis of the CT beam [2]. In addition, further studies have shown that—under certain configurations—the angular dependence of the XR-QA films can introduce significant uncertainties [2,12]. Nonetheless, it was established that a complete CT scan introduces a 2% overall uncertainty [2].
The aim of this work is to carry out a comparison between measurements performed with GafchromicTM XR-QA2 and TLD 100H, placed within an inhomogeneous anthropomorphic phantom and exposed to a thoracic CT scan. The radiochromic films and the TLDs were placed within the thorax region corresponding to the spinal cord region.

2. Materials and Methods

2.1. Dosimeters

The TLD 100H is a thermoluminescent dosimeter suited for medical physics dosimetry applications; it is dose sensitive inside the range 1 pGy—10 Gy and its size is 3.2 × 3.2 × 0.89 mm3, as reported in the Table 1 (www.thermofisher.com).
The emission of luminescence from TLD 100H were measured using a Risø TL/OSL-DA-10 reader. The 9635QA photomultiplier tube combined with Corning 7–59 and Schott BG-12 optical filters were used to detect luminescence emission.
The GafchromicTM XR-QA2 film is designed as a quality-assurance (QA) tool for radiology and it is radiation sensitive within the range 20–200 kVp and dose sensitive within the range 0.1–20 cGy (http://www.ashland.com). It is a reflective-type film consisting of five layers: a 25-µm-thick radiation-sensitive layer is stacked to a 97-µm-thick white and reflective polyester by means of a 3-µm-thick surface layer. A 97-µm-protective yellow polyester is over-coated on the top of the sensitive layer by means of a 15-µm-thick pressure sensitive adhesive.

2.2. Calibration Dosimeter Procedures

Irradiation was delivered with a Samsung GC 80 X-ray tube by employing an experimental setup characterized by the following: ΔV = 140 kVp, HVL = 5.7-mm Al, source-detector distance SDD = 60 cm, a 10 × 10 cm2 open field. The calibration of the X-ray tube was performed in terms of radiation output in air (mGy/mAs) by employing an DCT10-RS/Lemo ionization chamber connected to its electrometer (RTI Electronics, Montville, NJ, USA).
In order to produce the calibration curves for the dosimeters, for each dose point three GafchromicTM samples (15 mm × 15 mm) and three TLDs were irradiated in air with growing dose values.
The radiochromic films image acquisition was performed before the irradiation and 72 h after the irradiation [11,13], using an Epson Expression 10,000 XL flatbed scanner. The scanner that we employed has a negligible sensitivity variation in the central region that is used for the reading. This was set up using the reflective mode because of the film opaque property [2,10,14], an RGB image with a resolution of 72 dpi in 48-bit and no color corrections was applied [11,15]. The image analysis was performed in the red channel and the pixel-by-pixel background fog subtraction method was applied [11,16] using ImageJ software (available at https://imagej.net).
The TLDs signals after irradiation were acquired with a heating rate of 1 K/s from RT to 513.15 K in a nitrogen atmosphere, in order to prevent sensitivity changes due to high temperature. The signal bleaching before irradiation was performed by heating the TLD in the oven at 513.15 K for 10 min and by cooling it down at room temperature [17,18]. During the luminescence acquisitions, the laboratory was illuminated by using Ilford DL10 lamps filtered with Ilford #902 [19,20] in order to avoid any optical bleaching [20].

2.3. CT Measurements

An Alderson RandoTM woman anthropomorphic phantom was used in order to imitate the geometry of an adult woman patient. This phantom consists of 11 numbered transverse sections; each 25 mm thick (Figure 1a). The different holes present in each slab were filled with tissue-equivalent material (Figure 1b).
XR-QA2 films samples were cut in 4 cm × 8 cm pieces and applied onto the surface of the 17th, 18th and 19th phantom slabs, as in Figure 2a–c. A CT scan was then carried out. In particular, a four-slice CT scanner (GE Lightspeed) was used to scan the area between the 17th and 19th phantom slabs along the z-axis. The scan parameters were helical mode, AEC system, isocenter in the 18th phantom slab; length of scan equal to 153.75 mm; 3.75 mm slice-thickness; 140 kVp and pitch equal to 1. The same CT scan was repeated other 5 times, after placing the TLDs on the surface of the 17th, 18th and 19th slabs, in the position identified in Figure 2a–c, by the white points.

3. Results and Discussion

3.1. Dosimeter Dose Calibrations

The estimation of the X-ray tube output factor and the delivered-dose values for GafchromicTM and TLDs calibration procedures are reported in Table 2.
Figure 3 shows the PVisub values as a function of the delivered-dose values shown in Table 2. The PVisub values (red box) represent the measured pixel values subtracted from the background and obtained from a squared ROI (side of 15 pixels).
The data were fitted (blue curve) by the equation:
y = d + a d 1 + ( x c ) b
obtaining a R 2 correlation factor equal to 0.9999.
Figure 4 shows the trend of the TL intensity versus the dose values. The mean TL intensity, including standard deviation, is represented with black squares while for the linear fit a red line is used. The data were fitted by the equation y = A x where A is equal to 12,124 ± 97. The R2 correlation factor is 0.99961. The luminescence emission signals are normalized considering the sensitivity of each single dosimeter. The error propagation theory was applied to uncertainties evaluation starting from statistical counting of luminescence signals.

3.2. Dose Measurements

Figure 5a–c shows the 2D-dose distributions (on the phantom axial plane) obtained by XR-QA2 films. The lowest doses are recorded precisely at the spinal bone that surrounds the spinal cord with his bony parts, while an increasing dose is recorded in the area of the spinal cord.
Table 3 shows the maximum, minimum and average dose values obtained for each slab with XR-QA2 films.
Table 4 shows the maximum, minimum and average dose values obtained for each slab with TLD.
Figure 6a–c show the average dose values for each TLD measurements as a function of the position and the corresponding average dose value recorded by XR-QA2 films. The XR-QA2 film dose values were extracted using a 15 × 15 pixels ROI, centered on the corresponding position of the TLDs (see Figure 2a–c).
The mean-dose values obtained with the two dosimetric methodologies showed a homothetic behavior, due to the same scan parameters and scattering geometries, with TLD dose values being systematically higher than the GafchromicTM.
The results are in good agreement with the recent publications by Rampado et al. [9] and Brady et al. [2]. Figure 7a–c shows the slab-by-slab difference in percentage between the two dosimetric methodologies for each numbered TLD position.
Table 5 shows, for each slab, the maximum, minimum and the mean discrepancy in percentage between TLD and GafchromicTM dose responses.
The results obtained show higher average percentage discrepancy for the 19th slab compared to the 18th slab and, equally, for the 18th slab compared to the 17th slab.
The CT scan used was equipped with the automatic exposure control (EC) system that reduces the radiation dose to the patient based on their physical dimension and tissue absorption properties [15]. Figure 8a shows the behavior of the current output during the scan length. It emerged that larger current output values correspond to a larger discrepancy between dose values of TLDs and GafchromicTM (Figure 8b).

4. Conclusions

In order to consider high dose gradients, e.g., those generated by the bony parts surrounding some organ regions such as the spinal cord, it is helpful to produce high-resolution 2D dosimetric distributions. However, the contributions to the generation of dose distributions are complex and varied, e.g., the angular dependence, the scattering due to the different anatomic structures and the EC system that generates variation in the output current during a CT scan. Despite the complexity of these mechanisms, the XR-QA2 provides a good tool to perform patient dosimetry in CT using anthropomorphic phantoms. The dose values obtained using radiochromic films are systematically lower when compared to the dose values obtained with TLDs which are considered among the most accurate dosimeters for punctual dose estimation. The measurements carried out with TLDs show on average a 4–6% discrepancy with respect to the measurement carried out using GafchromicTM. It is possible to claim that these values of discrepancy are acceptable since their order of magnitude is comparable to the measured errors. Nevertheless, Figure 8a,b shows that the discrepancy between the two methodologies increases for low current output values. The observation of this trend could potentially allow to obtain a slice-by-slice correction factor which depends on the variation of the current output due to the use of the EC system.

Author Contributions

M.P.: conceptualization, investigation, methodology, software, writing original draft; G.S.: conceptualization, data curation, formal analysis, investigation, methodology, writing—review and editing; L.B.T.: project administration, validation; Stefania Mazzaglia: investigation; A.M.G.: conceptualization, validation, supervision, writing—review and editing. All authors have read and agreed to the published version of the manuscript.

Funding

This research received no external funding.

Conflicts of Interest

The authors declare no conflicts of interest.

References

  1. Report CEP08007. The imPACT Group 2009 Buyer’s Guided: Multi-Slice CT Scanners. St. George’s Healthcare Trust. March 2009. Available online: https://www.scribd.com/document/224289000/Cep-08007 (accessed on 3 July 2020).
  2. Brady, S.; Yoshizumi, T.; Toncheva, G.; Frush, D. Implementation of radiochromic film dosimetry protocol for volumetric dose assessments to various organs during diagnostic CT procedures. Med. Phys. 2010, 17, 4782–4792. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  3. Purwaningsih, S.; Lubis, L.E.; Pawiro, S.A.; Soejoko, D.S. Measurement of computed tomography dose profile with pitch variation using Gafchromic XR-QA2 and thermoluminescence dosimeter (TLD). J. Phys. 2016, 694, 012046. [Google Scholar] [CrossRef] [Green Version]
  4. Tomic, N.; Quintero, C.; Whiting, B.; Aldelaijan, S.; Bekerat, H.; Liang, L.; Devic, S. Characterization of calibration curves and energy dependence GafchromicTM XR-QA2 model based radiochromic film dosimetry system. Med. Phys. 2014, 6, 062105. [Google Scholar] [CrossRef] [PubMed]
  5. Dong, S.L.; Chu, T.C.; Lan, G.Y.; Wu, T.H.; Lin, Y.C.; Lee, J.S. Characterization of high-sensitivity metal oxide semiconductor field effect transistor dosimeters system and LiF: Mg, Cu, P thermoluminescence dosimeters for use in diagnostic radiology. Appl. Radiat. Isot. 2002, 57, 883–891. [Google Scholar] [CrossRef]
  6. Triolo, A.; Brai, M.; Bartolotta, A.; Marrale, M. Glow curve analysis of TLD-100H irradiated with radiation of different LET: Comparison between two theoretical method. Nucl. Instrum. Method. Phys. Res. Sect. A Accel. Spectrom. Detect. Assoc. Equip. 2006, 560, 413–417. [Google Scholar] [CrossRef]
  7. Carinou, E.; Boziari, A.; Askounis, P.; Mikulis, A.; Kamenopoulou, V. Energy dependence of TLD 100 and MCP-N detectors. Radiat. Meas. 2008, 43, 599–602. [Google Scholar] [CrossRef]
  8. Tomic, N.; Devic, S.; DeBlois, F.; Seuntjens, J. Reference radiochromic film dosimetry in kilovoltage photon beams during CBCT image acquisition. Med. Phys. 2020, 37, 1083–1092. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  9. Rampado, O.; Garelli, E.; Ropolo, R. Computed tomography dose measurements with radiochromic films and a flatbed scanner. Med. Phys. 2010, 37, 189–196. [Google Scholar] [CrossRef] [PubMed]
  10. Asero, G.; Greco, C.; Gueli, A.M.; Raffaele, L.; Spampinato, S. Evaluation of spatial resolution in image acquisition by optical flatbed scanners for radiochromic film dosimetry. J. Instrum. 2016, 11, P03024. [Google Scholar] [CrossRef]
  11. Pace, M.; Tonghi, L.B.; Mazzaglia, S.; Stella, G.; Tuvè, C.; Gueli, A.M. 3-D dose distribution for organ dose measurement in CT thoracic exams using Gafchromic XR-QA2 films. J. Instrum. 2019, 14, P09010. [Google Scholar] [CrossRef]
  12. Giaddui, T.; Cui, Y.; Galvin, J.; Chen, W.; Yu, Y.; Xiao, Y. Characteristics of Gafchromic XRQA2 films for kV image dose measurement. Med. Phys. 2012, 39, 842–850. [Google Scholar] [CrossRef] [PubMed]
  13. Butson, M.J.; Peter, K.N.; Cheung, T.; Metcalfe, P. Radiochromic film for medical radiation dosimetry. Mater. Sci. Eng. R Rep. 2003, 41, 61–120. [Google Scholar] [CrossRef]
  14. Alnawaf, H.; Butson, M.J.; Cheung, T.; Peter, K.N. Scanning orientation and polarization effects for XRQA radiochromic film. Phys. Med. 2010, 26, 216–219. [Google Scholar] [CrossRef] [PubMed]
  15. Spampinato, S.; Gueli, A.M.; Milone, P.; Raffaele, L.A. Dosimetric changes with computed tomography automatic tube-current modulation techniques. Radiol. Phys. Tech. 2018, 11, 184–191. [Google Scholar] [CrossRef] [PubMed]
  16. Gueli, A.M.; Cavalli, N.; De Vincolis, R.; Raffaele, L.; Troja, S.O. Background fog subtraction methods in Gafchromic® dosimetry. Radiat. Meas. 2015, 72, 44–52. [Google Scholar] [CrossRef]
  17. Mahajna, S.; Yossian, D.; Horowitz, Y.S. The effect of thermal annealing on defect structure and thermoluminescence in LiF: Mg, Cu, P. Radiat. Meas. 1995, 24, 395–400. [Google Scholar] [CrossRef]
  18. Stella, G.; Mazzaglia, S.; Pace, M.; Tonghi, L.B.; Tuvè, C.; Gueli, A.M. QA for calibration procedures of TLDs 100H in low doses range. J. Instrum. 2019, 14, P06023. [Google Scholar] [CrossRef]
  19. Stella, G.; Fontana, D.; Gueli, A.; Troja, S. Historical mortars dating from OSL signals of fine grain fraction enriched in quartz. Geochronometria 2013, 40, 153–164. [Google Scholar] [CrossRef] [Green Version]
  20. Reshes, G.; Druzhyna, S.; Biderman, S.; Eliyahu, I.; Oster, L.; Horowitz, Y.S. Study of the effect of optical bleaching at selected photon energies on the optical absorption and thermoluminescence of LiF: Mg, Ti (TLD-100). Radiat. Meas. 2017, 106, 26–29. [Google Scholar] [CrossRef]
Figure 1. Alderson RandoTM woman anthropomorphic phantom: (a) front view and measurement area; (b) top view of a slab. The scan axis is along the z direction while x–y is the axial plane.
Figure 1. Alderson RandoTM woman anthropomorphic phantom: (a) front view and measurement area; (b) top view of a slab. The scan axis is along the z direction while x–y is the axial plane.
Instruments 04 00019 g001
Figure 2. Position of radiochromic films (orange squares) and TLDs (white points) on the axial plane of the phantom slabs. (a) 17th slab: numbered positions from 27 to 36; (b) 18th slab: numbered positions from 14 to 26; (c) 19th slab: numbered positions from 1 to 13.
Figure 2. Position of radiochromic films (orange squares) and TLDs (white points) on the axial plane of the phantom slabs. (a) 17th slab: numbered positions from 27 to 36; (b) 18th slab: numbered positions from 14 to 26; (c) 19th slab: numbered positions from 1 to 13.
Instruments 04 00019 g002
Figure 3. Calibration curve for XR-QA2 GafchromicTM.
Figure 3. Calibration curve for XR-QA2 GafchromicTM.
Instruments 04 00019 g003
Figure 4. Calibration curve for TLD 100H.
Figure 4. Calibration curve for TLD 100H.
Instruments 04 00019 g004
Figure 5. Film dose responses (mGy) in color wash: (a) 2D-dose distribution in the spinal cord region for the 17th slab; (b) 2D-dose distribution in the spinal cord region for the 18th slab; (c) 2D-dose distribution in the spinal cord region for the 19th slab.
Figure 5. Film dose responses (mGy) in color wash: (a) 2D-dose distribution in the spinal cord region for the 17th slab; (b) 2D-dose distribution in the spinal cord region for the 18th slab; (c) 2D-dose distribution in the spinal cord region for the 19th slab.
Instruments 04 00019 g005
Figure 6. Average dose values for GafchromicTM XR-QA2 (blue dot) and TLD 100H (red dot): (a) dose values trend for the 17th slab; (b) dose values trend for the 18th slab; (c) dose values trend for the 19th slab.
Figure 6. Average dose values for GafchromicTM XR-QA2 (blue dot) and TLD 100H (red dot): (a) dose values trend for the 17th slab; (b) dose values trend for the 18th slab; (c) dose values trend for the 19th slab.
Instruments 04 00019 g006
Figure 7. The difference in percentage for GafchromicTM and TLD responses: (a) 17th slab; (b) 18th slab; (c) 19th slab.
Figure 7. The difference in percentage for GafchromicTM and TLD responses: (a) 17th slab; (b) 18th slab; (c) 19th slab.
Instruments 04 00019 g007
Figure 8. (a) Current output vs. scan length; the colored squares correspond to the film position between the phantom slabs; (b) mean slice-by-slice %Δ(TLD/GAF) correction factors vs. CT current output.
Figure 8. (a) Current output vs. scan length; the colored squares correspond to the film position between the phantom slabs; (b) mean slice-by-slice %Δ(TLD/GAF) correction factors vs. CT current output.
Instruments 04 00019 g008
Table 1. TLD 100H specifications.
Table 1. TLD 100H specifications.
Emission Wavelength
(nm)
Relative Energy ResponseSize (mm3)Sensitivity
4000.983.2 × 3.2 × 0.891 pGy–10 Gy
Table 2. Calibration step: X-ray-tube output and dose values.
Table 2. Calibration step: X-ray-tube output and dose values.
Tube Output (mGy/mAs)Dose (mGy)
0.45 ± 0.019.11 ± 0.19
18.23 ± 0.38
36.46 ± 0.77
57.41 ± 1.03
72.89 ± 1.37
91.12 ± 1.80
Table 3. Maximum, minimum and average dose value measured with XR-QA2 films for the 17th, 18th and 19th slab.
Table 3. Maximum, minimum and average dose value measured with XR-QA2 films for the 17th, 18th and 19th slab.
XR-QA2
Max Dose
(mGy)
Min Dose
(mGy)
Mean Dose
(mGy)
17th slab25.82 ± 1.3516.79 ± 0.8821.52 ± 2.93
18th slab26.02 ± 1.3617.41 ± 0.9120.95 ± 2.77
19th slab24.05 ± 1.2617.45 ± 0.9120.55 ± 2.11
Table 4. Maximum, minimum and average dose value measured with TLD 100H for the 17th, 18th and 19th slab.
Table 4. Maximum, minimum and average dose value measured with TLD 100H for the 17th, 18th and 19th slab.
TLD 100H
Max Dose
(mGy)
Min Dose
(mGy)
Mean Dose
(mGy)
17th slab27.01 ± 0.6517.01 ± 0.6222.50 ± 3.11
18th slab26.21 ± 0.6818.81 ± 0.6121.90 ± 2.65
19th slab25.35 ± 0.5918.87 ± 0.5821.86 ± 2.17
Table 5. Maximum, minimum and average percentage discrepancy for the 17th, 18th and 19th slab.
Table 5. Maximum, minimum and average percentage discrepancy for the 17th, 18th and 19th slab.
Max %Δ(TLD/GAF)Min %Δ(TLD/GAF)Mean %Δ(TLD/GAF)
17th slab7.91.43.9
18th slab7.70.24.4
19th slab8.245.8

Share and Cite

MDPI and ACS Style

Pace, M.; Stella, G.; Tonghi, L.B.; Mazzaglia, S.; Gueli, A.M. CT-Dose Measurement of the Spinal Cord Region Using XR-QA2 Radiochromic Films and TLD 100H Dosimeters. Instruments 2020, 4, 19. https://doi.org/10.3390/instruments4030019

AMA Style

Pace M, Stella G, Tonghi LB, Mazzaglia S, Gueli AM. CT-Dose Measurement of the Spinal Cord Region Using XR-QA2 Radiochromic Films and TLD 100H Dosimeters. Instruments. 2020; 4(3):19. https://doi.org/10.3390/instruments4030019

Chicago/Turabian Style

Pace, Martina, Giuseppe Stella, Letizia Barone Tonghi, Stefania Mazzaglia, and Anna Maria Gueli. 2020. "CT-Dose Measurement of the Spinal Cord Region Using XR-QA2 Radiochromic Films and TLD 100H Dosimeters" Instruments 4, no. 3: 19. https://doi.org/10.3390/instruments4030019

Article Metrics

Back to TopTop