Article

Mechanical and Morphological Assessment of an Innovative Textile for Patient Positioning Applications: Comparison to Two Standard Bandage Systems

1
Experimental Orthopedics, Department of Orthopedic and Trauma Surgery, Medical University Innsbruck, 6020 Innsbruck, Austria
2
Department of Orthopedics and Trauma Surgery, Medical University of Innsbruck, 6020 Innsbruck, Austria
*
Author to whom correspondence should be addressed.
Materials 2021, 14(6), 1508; https://doi.org/10.3390/ma14061508
Submission received: 27 January 2021 / Revised: 10 March 2021 / Accepted: 17 March 2021 / Published: 19 March 2021

Abstract

In the healthcare environment, bandage systems are versatile medical devices to position and fix patients’ torsos or extremities. In this study, the mechanical and morphological properties of an innovative patient position system, iFix, were assessed and compared to two commercially available bandages. Morphological properties were investigated using a scanning electron microscope (SEM). The iFix bandage showed anisotropic mechanical properties, with a more rigid behavior in the longitudinal direction and a more elastic behavior in the transverse direction. This behavior results from the organization of the fibers visible in the SEM images. All three materials investigated in this study were able to support similar maximum loads. In cases where a rigid fixation of patient limbs or torso is necessary, the authors recommend the usage of iFix. In vivo studies should be carried out to prove safety in a surgical environment before its clinical usage.
Keywords: patient positioning system; test methods for textiles; stiffness of bandages; bandage systems; fixation of patients patient positioning system; test methods for textiles; stiffness of bandages; bandage systems; fixation of patients

Share and Cite

MDPI and ACS Style

Putzer, D.; Dammerer, D.; Coraça-Huber, D.; Pallua, J.; Schmölz, W.; Nogler, M. Mechanical and Morphological Assessment of an Innovative Textile for Patient Positioning Applications: Comparison to Two Standard Bandage Systems. Materials 2021, 14, 1508. https://doi.org/10.3390/ma14061508

AMA Style

Putzer D, Dammerer D, Coraça-Huber D, Pallua J, Schmölz W, Nogler M. Mechanical and Morphological Assessment of an Innovative Textile for Patient Positioning Applications: Comparison to Two Standard Bandage Systems. Materials. 2021; 14(6):1508. https://doi.org/10.3390/ma14061508

Chicago/Turabian Style

Putzer, David, Dietmar Dammerer, Débora Coraça-Huber, Johannes Pallua, Werner Schmölz, and Michael Nogler. 2021. "Mechanical and Morphological Assessment of an Innovative Textile for Patient Positioning Applications: Comparison to Two Standard Bandage Systems" Materials 14, no. 6: 1508. https://doi.org/10.3390/ma14061508

APA Style

Putzer, D., Dammerer, D., Coraça-Huber, D., Pallua, J., Schmölz, W., & Nogler, M. (2021). Mechanical and Morphological Assessment of an Innovative Textile for Patient Positioning Applications: Comparison to Two Standard Bandage Systems. Materials, 14(6), 1508. https://doi.org/10.3390/ma14061508

Note that from the first issue of 2016, this journal uses article numbers instead of page numbers. See further details here.

Article Metrics

Back to TopTop