Evaluation of the Tribocorrosion Behavior of Ti-6Al-4V Biomedical Alloy in Simulated Oral Environments

by
Sónia I. G. Fangaia
1,2,3,*,
Ana Messias
1,4,
Fernando A. D. R. A. Guerra
1,3,
Ana C. F. Ribeiro
2,
Artur J. M. Valente
2 and
Pedro M. G. Nicolau
1,3
1
Institute of Implantology and Prosthodontics, Faculty of Medicine, University of Coimbra, 3000-075 Coimbra, Portugal
2
Coimbra Chemistry Centre (CQC), Department of Chemistry, University of Coimbra, 3004-535 Coimbra, Portugal
3
Center for Innovation and Research in Oral Sciences (CIROS), Faculty of Medicine, University of Coimbra, 3000-075 Coimbra, Portugal
4
Center of Mechanical Engineering Materials and Processes (CEMMPRE), Departamento de Engenharia Mecânica, University of Coimbra, 3030-788 Coimbra, Portugal
*
Author to whom correspondence should be addressed.
Processes 2024, 12(7), 1283; https://doi.org/10.3390/pr12071283
Submission received: 23 May 2024 / Revised: 14 June 2024 / Accepted: 18 June 2024 / Published: 21 June 2024
(This article belongs to the Special Issue Processing, Manufacturing and Properties of Metal and Alloys)

Abstract

:
The sliding wear of Ti-6Al-4V alloys coexisting with dental amalgam in a simulated temperature-controlled cell was evaluated. Disc-shaped samples of Ti-6Al-4V (n = 30) and spherical silver amalgam (n = 30) were prepared. Discs were subjected to wear while immersed in artificial and fluoridated saliva as follows: Ti-6Al-4V–Ti-6Al-4V (G1); amalgam–amalgam (G2), and Ti-6Al-4V–amalgam (G3). Samples were analyzed for mass variation, volume loss, and surface roughness. Wear tracks were characterized by scanning electron microscopy. Wearing induced significant mass loss for all groups except G3 in fluoridated saliva: Ti-6Al-4V (p = 0.045) and amalgam (p = 0.732). These samples presented an increase in mean surface roughness (p = 0.032 and 0.010, respectively). Overall, Ti-6Al-4V showed 0.07 mm3 (95% CI: [0.06–0.07]) higher wear track volume. Ti-6Al-4V has a higher mass loss when subjected to fluoridated media but no significant roughness variation. Fluor-containing substances should be avoided over Ti-6Al-4V alloys placed in areas of mechanical wear, especially if dental amalgam is also present.

1. Introduction

Metal alloys used in the oral cavity for restorative or prosthetic purposes are subjected to a wet environment, where saliva can act as an electrolytic solution that promotes electrochemical reactions between the different alloys. Therefore, metal restorations permanently held in saliva can undergo progressive degradation due to corrosion [1]. Up to now, however, there is little evidence that multiple metal restorations on the same medium increase wear. Nonetheless, empirical clinical observations frequently detect signs of wear in dental restorations; this phenomenon could be intensified by the coexistence of different metal alloys in the oral cavity, such as amalgam, cobalt–chromium, and titanium alloys.
The expansion of dental implants and implant-supported prostheses increased the development and optimization of metallic alloys [2], namely titanium alloys, such as Ti-6Al-4V. This alloy has higher strength than commercially pure grade IV titanium and can be used to manufacture implant components (namely abutments) and the frameworks of fixed and removable prostheses (tooth and implant-supported).
In the oral cavity, Ti-6Al-4V biomedical alloys are subjected to a process called tribocorrosion, which is the combination of corrosion and wear action, culminating in irreversible material degradation, in which wear probably contributes to the initiating mechanism [3,4,5,6]. Adequate wear resistance is a requirement of a dental alloy used in oral rehabilitation to ensure the durability of the restorations. On the contrary, a high wear rate may be related to the elongation of antagonists, tilting, and movement of teeth and other dysfunctions [7].
Although titanium and its alloys are considered corrosion-resistant, they are susceptible to acidic environments and to the presence of fluoride ions [8,9]. According to the literature, fluoride ions present in toothpaste, mouthwashes, and fluoridated acidic prophylactic gels may be harmful to the corrosion resistance of titanium-based materials for dental use. The fluoride ion can disrupt the protective oxide layer present on the surface of Ti alloys, which has been considered responsible for the corrosion resistance of the mentioned alloys [3,10,11,12]. This is an issue to take into account, because fluoride is a fundamental agent in the prevention of tooth decay [13].
Several studies evaluate the effect of fluoride on the corrosion behavior of titanium isolated or coupled with amalgam but under static conditions [1,14,15,16]. Other authors also evaluated tribocorrosion, particularly stress corrosion [11], fretting corrosion [17,18], fatigue [19,20,21,22], and elastic tensile strain [10] but on dental alloys isolated.
The literature is sparse regarding the evaluation of different alloys in the same medium under dynamic wear testing conditions, with and without fluoride. Thus, the purpose of the present study was to evaluate the sliding wear of two dental alloys coexisting in a simulated temperature-controlled cell containing artificial saliva, in the presence and absence of fluoride ions. The fluoridated saliva with a concentration of 225 ppm of fluoride was used since almost all fluoride mouthwashes on sale have this concentration [13]. We hypothesized that due to the galvanic phenomenon, the simultaneous presence of amalgam and Ti-6Al-4V samples in a wet medium leads to higher wear and surface roughness in amalgam and lower wear and surface roughness in Ti-6Al-4V, compared to the samples isolated. Additionally, we hypothesized that the fluoridation of the wet medium would contribute to an increase in the wear of Ti-6Al-4V samples due to the disruption of the superficial oxide layer, regardless of the presence of the amalgam.

2. Materials and Methods

2.1. Preparation of Test Materials

To carry out this study, 10 mm × 3 mm disc-shaped samples of two metallic alloys with application in Dentistry were prepared: a titanium alloy (Ti-6Al-4V) and a spherical silver amalgam alloy with the following composition in percent weight (wt.%): Ag—60 wt.%, Sn—28 wt.%, and Cu—12 wt.%.
Ti-6Al-4V samples (n = 30) were obtained by machining metal rods (titanium/aluminum/vanadium Ti-6Al-4V Rod, Lot LS278237SJP, Goodfellow Cambridge Limited®, Huntingdon, UK). Samples were then mechanically polished for 1 min using abrasive sandpaper grit 1000 (Struers Inc., Cleveland, OH, USA) at 150 rpm and 10 N pressure.
Amalgam samples (n = 30) were prepared using self-activating capsules (Tytin, Kerr Manufacturing Co., Romulus, MI, USA; Lot 3-3069), as indicated by the manufacturer. The capsule was placed and secured in the amalgamator and triturated for 8 s. The mix was immediately condensed after trituration according to Specification No. 1 ANSI/ADA into a 10 mm × 3 mm cylindric mold. After setting, the samples were polished for 15 s using abrasive sandpaper grit 2500 (Struers Inc., Cleveland, OH, USA) at 300 rpm and 5 N pressure.
The polished samples were enclosed in an acrylic resin cylinder, subjected to a mass loss study, and then stored in water for 14 days at 37 °C. After storage, samples were cleaned using distilled water in an ultrasonic cleaner and stored for another 7 days at room temperature.

2.2. Initial Characterization of the Samples

Before being subjected to tribological experiments, the samples were analyzed with regard to initial mass, surface roughness, and Vickers hardness.

2.2.1. Initial Mass

All samples were weighed before the wear tests using an analytical electronic scale (Model AA-200Ds, Denver Instrument Co, Denver, CO, USA), with a 0.0001 g sensitivity.

2.2.2. Surface Roughness

The surface roughness of the samples was determined in a profilometry device (Perthometer Mahr, model Perthen S4P, Göttingen, Germany), using a filter LT (0.800*5) 5600 Mn. The amplitude parameters used were arithmetic average height (Ra) and ten-point height (Rz).
Ra is defined as the average absolute deviation of the roughness irregularities from the mean line over one sampling length [23].
Rz is more sensitive to occasional high peaks or deep valleys than Ra. The ISO system defines this parameter as the difference in height between the average of the five highest peaks and the five lowest valleys along the assessment length of the profile [23].

2.2.3. Vickers Hardness Measurement

Vickers hardness test is based on the resistance that the material offers to the penetration of a diamond pyramid with a square base and an angle between faces of 136°, under a given load. The samples were subjected to forces of 1, 2, 5, 10, and 20 N in a micro Vickers hardness tester (model HMV-2, Shimadzu, Japan) that allows the acquisition, recording, and processing of data for the evaluation of hardness. Five repetitions were performed for each force. This test followed ISO 6507 standards [24].

2.3. Tribocorrosion Experiments

The samples were tested in couples: Ti-6Al-4V + Ti-6Al-4V (G1); amalgam + amalgam (G2); and Ti-6Al-4V + amalgam (G3) and subjected to a wear test.
The wear was carried out in alternative sliding equipment with customizable stroke, speed, and load (Figure 1) (Pedro Nunes Institute, Coimbra, Portugal), which simulates the conditions of use usually given to these materials in the oral cavity. The wear equipment operates by applying a vertical load to the specimen using a counter body, sliding horizontally, and then repeating the cycle. The counter-body type was made of a chemically inert ceramic material (95% Alumina spheres (Al2O3) ¼″, Helmut Kreutz, Haiger, Germany) [25], with a hardness similar to dental enamel; the contact is a punctual type using a counter-plane sphere. The parameters for this test were as follows: a load of 15 N and 2400 cycles at 0.7 Hz.
During the tests, the samples were immersed in an electrolyte solution kept in circulation at 310.15 K, simulating the conditions found in the oral cavity. After the tests, the samples were washed with distilled water, ultrasonically cleaned for 5 min, and dried, before final characterization.
Two electrolyte solutions were considered: artificial saliva and artificial fluoridated saliva; the latter was prepared by using sodium fluoride (NaF) at 225 ppm. Artificial saliva was prepared according to the following composition [26,27]: potassium chloride (KCl, 20 mmol/L), sodium hydrogen carbonate (NaHCO3, 17.9 mmol/L), sodium dihydrogen phosphate (NaH2PO4·H2O, 3.6 mmol/L), potassium thiocyanate (KSCN, 5.1 mmol/L), and lactic acid (0.10 mmol/L). All solutions were prepared using Milli-Q ultrapure water of resistivity >18 MΩcm and analytical grade reagents. The laboratory temperature was 25 ± 1 °C in all experiments.

2.4. Final Characterization of the Samples

After the tribocorrosion experiments, all samples were subject to weighing in the same equipment described in Section 2.2.1, and the wear track profile was acquired by Perthometer Perthen S4P, previously described in Section 2.2.2, in order to evaluate the final surface roughness and the volume of the wear tracks.
The wear track morphology was also observed by scanning electron microscopy (SEM JEOL JSM-5310, Tokyo, Japan) with a voltage of 20 kV.

2.5. Statistical Analysis

All statistical analyses were carried out on SPSS Statistics Version 27.0 (IBM, New York, NY, USA). Data on the variation in mass, wear track volume, and surface roughness variation (Ra and Rz) of the samples were analyzed using a three-way factorial design (three-way ANOVA), considering as independent variables, the material, the electrolyte, and the cylinder combination (pairs of cylinders simultaneously tested). Data were checked for normality and homoscedasticity and for three-way and simple two-way interactions.
Simple comparisons (pre–post and between levels of the independent variables—between alloys or between electrolyte solutions) were performed with non-parametric tests (Kruskal–Wallis and Mann–Whitney U for independent groups, Wilcoxon for paired data), followed by post hoc Bonferroni correction where applicable. The significance level was set at 0.05.

3. Results

3.1. Specimen Mass Variation

The mean and median values of the registered initial and final mass of the Ti-6Al-4V and amalgam samples tested in artificial saliva (AS) and fluoridated artificial saliva (FS) are detailed in Table 1.
Wearing induces a statistically significant mass loss for all groups, except for the amalgam samples in G3 (simultaneous testing of Ti-6Al-4V and amalgam specimens) when submerged in fluoridated saliva (p = 0.732).
There were statistically significant differences in mass loss between G1 and G2 regardless of the electrolyte solution (AS or FS, p = 0.004 and p = 0.026, respectively).
However, there were no statistically significant differences between TiAlV samples (G1 and G3, p > 0.05) or amalgam samples (G2 and G3, p = 0.205) after immersion in artificial saliva.
Three-way ANOVA detected a statistically significant three-way interaction between cylinder material, cylinder combination in testing setup (Ti-6Al-4V–Ti-6Al-4V or amalgam–amalgam or Ti-6Al-4V–amalgam), and type of electrolyte solution: F(1, 52) = 5.40, p = 0.024. There was a statistically significant simple two-way interaction between material and type of electrolyte solution for cylinder combination Ti-6Al-4V–amalgam, F(1, 52) = 8.56, p = 0.01, but not for cylinders of the same material, F(1, 52) = 0.012, p > 0.05. Such findings are graphically replicated in Figure 2, in which it is possible to notice that the fluoridation of saliva promotes a similar increase in mass loss for the samples of Ti-6Al-4V and amalgam when tested separately (full thickness lines representing G1 and G2) but not when tested simultaneously (dashed lines representing Ti-6Al-4V and amalgam samples of G3). In this case, saliva fluoridation leads to higher mass loss of the Ti-6Al-4V samples (approximately twice that of G1) and lower mass loss of the amalgam samples (4 to 5 times lower than that for G2).

3.2. Vickers Hardness

The Vickers test revealed that Ti-6Al-4V samples present hardness superior to 3 GPa regardless of the applied load (up to 20 N). Under the same conditions, amalgam samples show lower test values, i.e., approximately 2 GPa.

3.3. Surface Roughness

Prior to the wear test (T0), all groups are characterized by similar arithmetic average surface roughness (Ra). The exceptions are Ti-6Al-4V samples of G3, tested in fluoridated saliva, for which lower Ra values were obtained, as depicted in Table 2.
The wearing test induced an increase in Ra values for all amalgam samples, statistically significant for G2 in saliva (p = 0.001) and G3 in FS (0.010). After the wear test (T1), the Ra values of titanium samples remained relatively similar to those of the pre-test (T0) despite the slight increase in Ti-6Al-4V samples of G3 in FS.
At T0, there were no statistically significant differences in the average maximum height of the profile (Rz) (p = 0.426). After the wear test (T1), there was an increase in Rz of all amalgam samples, as detailed in Table 2 and graphically represented in Figure 3. The analysis of data shows that G3 amalgam samples, after immersion in FS, lead to statistically higher Rz than their counterpart Ti-6Al-4V samples (p = 0.036).
Ti6-Al-4V samples of G1 presented almost null Rz variation regardless of the electrolyte solution (Figure 3). However, Ti-6Al-4V samples of G3 presented a slight increase in Rz in artificial saliva, whereas there was a reduction in the maximum height of the profile for Ti-6Al-4V samples of G3 tested in fluoridated saliva.

3.4. Volume of the Wear Tracks

The mean volumes of the wear tracks of each group are shown in Table 3.
Three-way ANOVA was unable to detect a three-way interaction between material, type of electrolyte solution, and cylinder combination in the volume of the wear track: F(1, 52) = 1.05, p = 0.311. Also, no two-way interactions were identified between any pair of variables (p > 0.05 for all interactions). Therefore only the variable “material” shows a statistically significant main effect (F(1, 52) = 310.68, p < 0.001), with higher wear track volumes equal to 0.07 mm3 (95% CI: [0.06–0.07]).
The absence of simple main effects for electrolyte solution and any type of interactions indicated that there were no statistically significant differences between samples of the same material in different electrolyte solutions, i.e., Ti-6Al-4V in artificial saliva presented comparable volume loss to Ti-6Al-4V in fluoridated artificial saliva (p = 0.315) and a similar situation occurred with amalgam in artificial and fluoridated artificial saliva (p = 0.190). The simultaneous presence of two different materials had no impact on the volume of the wear track for both Ti-6Al-4V and amalgam (p > 0.05 and p = 0.151, respectively).

3.5. SEM Observation of the Wear Tracks

Representative images of the scanning electron microscopy observation of different samples after the wearing tests are shown in Figure 4, Figure 5, Figure 6, Figure 7, Figure 8, Figure 9 and Figure 10.
In Figure 4, amalgam samples tested simultaneously with Ti-6Al-4V (G3) in AS show uniform wear tracks, without significant material destruction. However, some pores are visible, as detailed in Figure 4b. When these two materials are tested simultaneously in FS, it is possible to observe cylindrical particles deposited on the surface of the amalgam samples (Figure 5b), even though no significant destruction can be reported for the wear tracks. From the analysis of micrographs of Ti-6Al-4V (Figure 6), we can conclude that wear causes significant surface destruction, occasionally with micro-cracks in the bottom of the track perpendicular to the sliding direction (Figure 7). It is also possible to observe the accumulation of debris particles in front of the counter body (Figure 6a) as well as the formation of a lamellae structure on the periphery of the wear tracks (Figure 6b). The crumbling material that drags in front of the counter body works as a third body and sometimes facilitates wear as a lubricant.
When these two materials are simultaneously tested in fluoridated saliva, the Ti-6Al-4V samples (Figure 8) show visible tips, material deposit on the bottom of the wear track, and extensive ploughing.
The wear tracks of amalgam samples (G2) tested in artificial saliva and fluoridated saliva are presented in Figure 9a,b, respectively. The wear tracks of these samples show neither morphological differences nor material pullout, although it is possible to see on the surface the same cylindrical particles that are observed in Figure 5b.
Ti-6Al-4V samples of G1 tested in artificial saliva shown in Figure 10a present material deformation aligned in lamellae, small microfractures in a repetitive process, extensive ploughing, large deformation, and very strong abrasion, but no great crumbling. The dragged material shapes a track and not the counter body.
Micrographs of Ti-6Al-4V samples of G2 tested in fluoridated saliva (Figure 10b) reveal a spallation and extensive ploughing on the wear track, demonstrating a strong abrasive wear action of the alumina ball.

4. Discussion

Dental alloy corrosion measured by electrochemical methods has been extensively tested in recent years; however, in our literature review, we found that there is a gap in tribological studies for TI-6Al-4V. Having in mind that there are many patients who have different metal alloys in the oral cavity and corrosion rate, in general, strongly affected by mechanical processes such as wear, a sliding device has been developed that makes it possible to simulate the coexistence of two alloys but also subject them to the action of forces when immersed in an electrolyte solution. In previous studies, it has been reported that galvanic corrosion occurs in amalgam alloys in contact with Ti implants [28]. Bumgardner and Johansson [29] have also found that coupling those two dental alloys may undergo the release of saliva of potentially toxic corrosion products. Although the use of amalgam has fallen into disuse, we still find many patients with teeth restored with this material. So, in the present study, the wearing of two different metallic alloys, amalgam and Ti-6Al-4V, submerged in a wet medium, with and without fluoride, was evaluated.
The main results obtained in this study indicate that, in artificial saliva, there is no significant difference in the average mass loss of the Ti-6Al-4V samples, either when tested alone or together with amalgam. However, fluoridation of the moist environment leads to an increase in mass loss in all the Ti-6Al-4V samples. This can be explained by the delamination of the oxide film on the surface of the Ti-6Al-4V samples due to the presence of fluoride ions [18]. It is important to note that when Ti-6Al-4V is tested simultaneously with amalgam samples, mass loss is twice that of Ti-6Al-4V tested alone.
Even though in this study mass loss was considered a surrogate variable for the wear assessment, other variables like surface roughness and volume of the wear track were assessed. According to DeLong, the volume is the preferred parameter for measuring the wear effect [30].
We obtained the wear volume by measuring the wear crater by profilometry, taking a measurement that crosses the wear track, and multiplying this value by the length of the wear track. Here we were unable to find differences between the wear crater of Ti-6Al-4V samples regardless of the presence of the amalgam or the type of electrolyte solution. Nevertheless, Ti-6Al-4V has, on average, 0.07 mm3 higher volume loss than amalgam. This result is in agreement with those reported by Atar [31], who studied the sliding wear performance of Ti-6Al-4V, 316 stainless steel, and CoCr isolated in dry conditions. They demonstrated that the Ti6-Al-4V has poor wear resistance, as justified by the low shear strength of titanium and its high tendency to scuff off the HCC structure, contrariwise dental amalgam is known for its high resistance to occlusal wear. We have found that no correlation between volume wear mass losses were found. The reason for this may be the high heterogeneity of the craters, as can be seen from SEM images, which leads to different volumes of wear paths.
With regard to the SEM images, in Ti-6Al-4V, extensive shear deformation and signals of abrasive wear were observed. We have also found some areas of particle detachments and delaminated wear compatible with adhesive wear. Our results are in line with the studies published by Atar, Fu et al., and Meddah et al. [31,32,33], even though the conditions of these studies are different. In the Ti-6Al-4V samples, the wear mechanism is formed with the worn material accumulating in the areas surrounding the wear track and, in particular, in front of the counter body, acting as a third body that increases surface degradation.
From the results obtained, we verified that there was extensive carving in the wear tracks as well as a delamination mechanism in the Ti-6Al-4V samples, which was corroborated by some other studies [33,34] that report that despite the high resistance of titanium to corrosion, it is little wear resistant. Among the damage mechanisms seen in Ti-6Al-4V, micro-cracks observed in the wear track act as crack initiation sites, which may coalesce and lead to premature fatigue.
In the amalgam samples, we observed grooves parallel to the sliding direction, compatible with the abrasive wear mechanism. It is also possible to observe cylindrical particles that we believe correspond to oxides; according to Landolt et al. [35], during the sliding contact, metallic wear particles are torn from the first body, and these particles create a metallic third body. Some may undergo electrochemical oxidation, leading to the formation of solid oxides, so it would be of interest in future investigations to analyze the surface chemical structure of these particles. Additionally, as previously described, when we tested the two alloys simultaneously, in a fluoridated medium, the mass of amalgam samples increased, which leads us to suppose that some debris and oxides released from the Ti-6Al-4V are deposited in amalgam samples.
While for all the amalgam samples, the roughness parameters analyzed (Ra and Rz) increased in both media; in the Ti-6Al-4V alloy, when tested simultaneously with amalgam, the Rz value in artificial saliva and the Ra value in fluoridated saliva slightly increased, with a contrasting decrease in the value of Rz. These results may be explained through the development of galvanic currents and by the release and deposition of oxides resulting from this phenomenon, as well as by the wear to which they were subjected.
Regarding Vickers hardness, we consider that it is stable with the loads used, since it does not suffer large variations, for both alloys.
The hardness of the metal alloy must be sufficient to resist occlusal forces without wearing down the opposing teeth. According to El-Bediwi et al. [36], alloys with a Vickers hardness <125 kg/mm2 are susceptible to wear, while with a hardness greater than 340 kg/mm2, there is a risk of wearing down the opposing teeth. Based on the results obtained, both Ti-6Al-4V and amalgam have surface hardness within this range of values.
According to Archard’s Law, the volume of loss material should be proportional to load and sliding distance and inversely proportional to the hardness value of the material [37]. Martin et al. [38] investigated the tribocorrosion behavior of Ti-6Al-4V and established a connection between a lower hardness and high mechanical wear.
In the present study, it was not possible to correlate the hardness of Ti-6Al-4V and wear, which is consistent with the findings of Chan et al. [39]; although the surface hardness of Ti-6Al-4V was higher than that of amalgam, mechanical wear was more pronounced.
It should be noted that in this tribological study, due to equipment limitations, the applied load was lower than that recommended by Heintze for this type of simulation. The author states that the applied load should be between 20 and 120 N, which is similar to that of the biting force in molars [7]. Nevertheless, 15 N was the load chosen by some other authors [40,41,42], and under 15 N load, severe material damage of the Ti-6Al-4V samples was observed, being higher than that of the amalgam samples.
Due to the large difference in galvanic potentials between amalgam and Ti-6Al-4V, it is reasonable that the amalgam suffers greater galvanic corrosion than the more noble alloy, Ti-6Al-4V. As would be expected, in artificial saliva, the Ti-6Al-4V sample does not suffer alterations when tested simultaneously with amalgam. However, this premise does not happen in the presence of fluoride; in this medium, Ti-6Al-4V suffers a greater loss of mass and a decrease in Rz, which makes us suppose that there is a synergistic effect between fluoride and amalgam.

5. Conclusions

Based on the data presented and discussed, we can conclude that the surface roughness and mass losses obtained for both alloys are in close agreement with the SEM analysis. It can also be concluded that, in artificial saliva when Ti-6Al-4V and amalgam are simultaneously present, no significant changes in the physical parameters of the former are found. However, in a fluoridated medium, when Ti-6Al-4V and amalgam are tested simultaneously, there is a higher mass loss in Ti-6Al-4V, and on the contrary, amalgam samples do not suffer any mass variation. In these conditions, the surface roughness (Rz) of Ti-6Al-4V reduces whereas its counterpart amalgam increases significantly. In fact, apparently, Ti-6Al-4V alloy seems to suffer the synergistic action of fluoride and amalgam. Fluor-containing substances should be avoided over Ti-6Al-4V alloys placed in areas of mechanical wear, especially if dental amalgam is also present.

Author Contributions

Conceptualization, P.M.G.N. and S.I.G.F.; methodology, F.A.D.R.A.G., P.M.G.N. and S.I.G.F.; software, S.I.G.F., A.M. and P.M.G.N.; validation, F.A.D.R.A.G., P.M.G.N. and S.I.G.F.; formal analysis, A.M. and S.I.G.F.; investigation, P.M.G.N. and S.I.G.F.; resources, S.I.G.F., A.M. and A.C.F.R.; data curation, A.M. and S.I.G.F.; writing—original draft preparation, S.I.G.F., A.M. and A.C.F.R.; writing—review and editing, S.I.G.F., P.M.G.N., A.J.M.V. and A.C.F.R.; visualization, S.I.G.F., P.M.G.N., A.J.M.V. and F.A.D.R.A.G.; supervision, A.M. and S.I.G.F.; project administration, S.I.G.F., P.M.G.N. and F.A.D.R.A.G.; funding acquisition, S.I.G.F., P.M.G.N., A.J.M.V. and A.C.F.R. All authors have read and agreed to the published version of the manuscript.

Funding

CEMMPRE is sponsored by national funds through FCT—Fundação para a Ciência e a Tecnologia, under the project UIDB/00285/2020 and CQC is supported by the Fundação para a Ciência e a Tecnologia (FCT), Portuguese Agency for Scientific Research, through the projects UIDB/QUI/UI0313/2020 and COMPETE Programme (Operational Programme for Competitiveness).

Data Availability Statement

The original contributions presented in the study are included in the article, further inquiries can be directed to the corresponding author.

Acknowledgments

The authors are grateful to LED&MAT (Laboratory for Wear, Testing & Materials) from the Pedro Nunes Institute for all the support given in the development of this study. The authors are also grateful to the Coimbra Chemistry Centre (CQC); Center for Innovation and Research in Oral Sciences (CIROS), Faculty of Medicine, University of Coimbra, Coimbra, Portugal and to CEMMPRE—Center of Mechanical Engineering Materials and Processes (CEMMPRE), Departamento de Engenharia Mecânica, University of Coimbra, 3030-788 Coimbra, Portugal.

Conflicts of Interest

The authors declare no conflicts of interest.

References

  1. Toumelin-Chemla, F.; Rouelle, F.; Burdairon, G. Corrosive Properties of Fluoride-Containing Odontologic Gels against Titanium. J. Dent. 1996, 24, 109–115. [Google Scholar] [CrossRef]
  2. Starosvetsky, D.; Gotman, I. Corrosion Behavior of Titanium Nitride Coated Ni–Ti Shape Memory Surgical Alloy. Biomaterials 2001, 22, 1853–1859. [Google Scholar] [CrossRef]
  3. Licausi, M.P.; Igual Muñoz, A.; Amigó Borrás, V. Influence of the Fabrication Process and Fluoride Content on the Tribocorrosion Behaviour of Ti6Al4V Biomedical Alloy in Artificial Saliva. J. Mech. Behav. Biomed. Mater. 2013, 20, 137–148. [Google Scholar] [CrossRef]
  4. Rocha, L.A.; Oliveira, F.; Cruz, H.V.; Sukotjo, C.; Mathew, M.T. Bio-Tribocorrosion in Dental Applications. In Bio-Tribocorrosion in Biomaterials and Medical Implants; Elsevier Inc.: Amsterdam, The Netherlands, 2013; pp. 223–249. ISBN 9780857095404. [Google Scholar]
  5. Dini, C.; Costa, R.C.; Sukotjo, C.; Takoudis, C.G.; Mathew, M.T.; Barão, V.A.R. Progression of Bio-Tribocorrosion in Implant Dentistry. Front. Mech. Eng. 2020, 6, 1. [Google Scholar] [CrossRef]
  6. Mathew, M.T.; Srinivasa Pai, P.; Pourzal, R.; Fischer, A.; Wimmer, M.A. Significance of Tribocorrosion in Biomedical Applications: Overview and Current Status. Adv. Tribol. 2009, 2009, 250986. [Google Scholar] [CrossRef]
  7. Heintze, S. How to Qualify and Validate Wear Simulation Devices and Methods. Dent. Mater. 2006, 22, 712–734. [Google Scholar] [CrossRef]
  8. Donachie, M. Titanium: A Technical Guide, 2nd ed.; Donachie, M., Ed.; ASM International: Almere, The Netherlands, 2000; ISBN 978-0-87170-686-7. [Google Scholar]
  9. Nakagawa, M.; Matsuya, S.; Shiraishi, T.; Ohta, M. Effect of Fluoride Concentration and PH on Corrosion Behavior of Titanium for Dental Use. J. Dent. Res. 1999, 78, 1568–1572. [Google Scholar] [CrossRef]
  10. Huang, H.-H. Effects of Fluoride Concentration and Elastic Tensile Strain on the Corrosion Resistance of Commercially Pure Titanium. Biomaterials 2002, 23, 59–63. [Google Scholar] [CrossRef]
  11. Könönen, M.H.O.; Lavonius, E.T.; Kivilahti, J.K. SEM Observations on Stress Corrosion Cracking of Commercially Pure Titanium in a Topical Fluoride Solution. Dent. Mater. 1995, 11, 269–272. [Google Scholar] [CrossRef] [PubMed]
  12. Kaneko, K.; Yokoyama, K.; Moriyama, K.; Asaoka, K.; Sakai, J.; Nagumo, M. Delayed Fracture of Beta Titanium Orthodontic Wire in Fluoride Aqueous Solutions. Biomaterials 2003, 24, 2113–2120. [Google Scholar] [CrossRef] [PubMed]
  13. Rugg-Gunn, A. Fluoride Toothpastes and Fluoride Mouthrinses. Acta Med. Acad. 2013, 42, 168–178. [Google Scholar] [CrossRef]
  14. Johansson, B.I.; Bergman, B. Corrosion of Titanium and Amalgam Couples: Effect of Fluoride, Area Size, Surface Preparation and Fabrication Procedures. Dent. Mater. 1995, 11, 41–46. [Google Scholar] [CrossRef]
  15. Reclaru, L.; Meyer, J.-M. Effects of Fluorides on Titanium and Other Dental Alloys in Dentistry. Biomaterials 1998, 19, 85–92. [Google Scholar] [CrossRef]
  16. Huang, H.-H. Effect of Fluoride and Albumin Concentration on the Corrosion Behavior of Ti–6Al–4V Alloy. Biomaterials 2003, 24, 275–282. [Google Scholar] [CrossRef]
  17. Vieira, A.C.; Ribeiro, A.R.; Rocha, L.A.; Celis, J.P. Influence of PH and Corrosion Inhibitors on the Tribocorrosion of Titanium in Artificial Saliva. Wear 2006, 261, 994–1001. [Google Scholar] [CrossRef]
  18. Sivakumar, B.; Kumar, S.; Sankara Narayanan, T.S.N. Fretting Corrosion Behaviour of Ti–6Al–4V Alloy in Artificial Saliva Containing Varying Concentrations of Fluoride Ions. Wear 2011, 270, 317–324. [Google Scholar] [CrossRef]
  19. Zavanelli, R.A.; Guilherme, A.S.; Pessanha-Henriques, G.E.; Antonio de Arruda Nobilo, M.; Mesquita, M.F. Corrosion-Fatigue of Laser-Repaired Commercially Pure Titanium and Ti-6Al-4V Alloy under Different Test Environments. J. Oral Rehabil. 2004, 31, 1029–1034. [Google Scholar] [CrossRef]
  20. Correa, C.B.; Pires, J.R.; Fernandes-Filho, R.B.; Sartori, R.; Vaz, L.G. Fatigue and Fluoride Corrosion on Streptococcus mutans Adherence to Titanium-Based Implant/Component Surfaces. J. Prosthodont. 2009, 18, 382–387. [Google Scholar] [CrossRef]
  21. Guilherme, A.S.; Henriques, G.E.P.; Zavanelli, R.A.; Mesquita, M.F. Surface Roughness and Fatigue Performance of Commercially Pure Titanium and Ti-6Al-4V Alloy after Different Polishing Protocols. J. Prosthet. Dent. 2005, 93, 378–385. [Google Scholar] [CrossRef] [PubMed]
  22. Okazaki, Y.; Rao, S.; Ito, Y.; Tateishi, T. Corrosion Resistance, Mechanical Properties, Corrosion Fatigue Strength and Cytocompatibility of New Ti Alloys without Al and V. Biomaterials 1998, 19, 1197–1215. [Google Scholar] [CrossRef] [PubMed]
  23. Gadelmawla, E.S.; Koura, M.M.; Maksoud, T.M.A.; Elewa, I.M.; Soliman, H.H. Roughness Parameters. J. Mater. Process Technol. 2002, 123, 133–145. [Google Scholar] [CrossRef]
  24. ISO 6507-1:2018; Metallic Materials—Vickers Hardness Test—Part 1: Test Method. International Standard Organization: Geneva, Switzerland, 2018.
  25. Lambrechts, P.; Debels, E.; Vanlanduyt, K.; Peumans, M.; Vanmeerbeek, B. How to Simulate Wear? Overview of Existing Methods. Dent. Mater. 2006, 22, 693–701. [Google Scholar] [CrossRef]
  26. Marek, M. Dissolution of Mercury from Dental Amalgam at Different PH Values. J. Dent. Res. 1997, 76, 1308–1315. [Google Scholar] [CrossRef] [PubMed]
  27. Horasawa, N.; Takahashi, S.; Marek, M. Galvanic Interaction between Titanium and Gallium Alloy or Dental Amalgam. Dent. Mater. 1999, 15, 318–322. [Google Scholar] [CrossRef] [PubMed]
  28. Taher, N. Galvanic Corrosion Behavior of Implant Suprastructure Dental Alloys. Dent. Mater. 2003, 19, 54–59. [Google Scholar] [CrossRef] [PubMed]
  29. Bumgardner, J.D.; Johansson, B.I. Galvanic Corrosion and Cytotoxic Effects of Amalgam and Gallium Alloys Coupled to Titanium. Eur. J. Oral Sci. 1996, 104, 300–308. [Google Scholar] [CrossRef] [PubMed]
  30. DeLong, R. Intra-Oral Restorative Materials Wear: Rethinking the Current Approaches: How to Measure Wear. Dent. Mater. 2006, 22, 702–711. [Google Scholar] [CrossRef] [PubMed]
  31. Atar, E. Sliding Wear Performances of 316 L, Ti6Al4V, and CoCrMo Alloys. Kov. Mater. 2013, 51, 183–188. [Google Scholar] [CrossRef]
  32. Meddah, S.; Chadli, H.; Bourebia, M.; Montagne, A.; Iost, A.; Labaiz, M.; Taleb, A. Dry Sliding Wear Performance of an Annealed TiNi Alloy with Different Nickel Contents. Mater. Res. Express 2020, 7, 036508. [Google Scholar] [CrossRef]
  33. Fu, Y.; Loh, N.L.; Wei, J.; Yan, B.; Hing, P. Friction and Wear Behaviour of Carbon Nitride Films Deposited on Plasma Nitrided Ti–6Al–4V. Wear 2000, 237, 12–19. [Google Scholar] [CrossRef]
  34. Lee, C.S.; Oh, J.C.; Lee, S. Improvement of Hardness and Wear Resistance of (TiC, TiB)/Ti-6Al-4V Surface-Alloyed Materials Fabricated by High-Energy Electron-Beam Irradiation. Metall. Mater. Trans. A 2003, 34, 1461–1471. [Google Scholar] [CrossRef]
  35. Landolt, D.; Mischler, S.; Stemp, M.; Barril, S. Third Body Effects and Material Fluxes in Tribocorrosion Systems Involving a Sliding Contact. Wear 2004, 256, 517–524. [Google Scholar] [CrossRef]
  36. El-Bediwi, A.; Saad, M.; El-Fallal, A.A.; El-Helaly, T. Electrochemical Corrosion Behavior, Vickers Microhardness, and Microstructure of Co–Cr and Ni–Cr Dental Alloys. Radiat. Eff. Defects Solids 2009, 164, 712–718. [Google Scholar] [CrossRef]
  37. Archard, J.F. Contact and Rubbing of Flat Surfaces. J. Appl. Phys. 1953, 24, 981–988. [Google Scholar] [CrossRef]
  38. Martin, É.; Azzi, M.; Salishchev, G.A.; Szpunar, J. Influence of Microstructure and Texture on the Corrosion and Tribocorrosion Behavior of Ti–6Al–4V. Tribol. Int. 2010, 43, 918–924. [Google Scholar] [CrossRef]
  39. Chan, K.S.; Koike, M.; Okabe, T. Modeling Wear of Cast Ti Alloys. Acta Biomater. 2007, 3, 383–389. [Google Scholar] [CrossRef]
  40. Hockin, X.; Giusepppetti, A.; Eichmiller, F.; Ives, L. Two-Body Sliding Wear of a Direct-Filling Silver Alternative to Amalgam. Quintessence Int. 1999, 30, 199–208. [Google Scholar]
  41. Sunnegårdh-Grönberg, K.; Peutzfeldt, A.; Van Dijken, J.W.V. Hardness and in Vitro Wear of a Novel Ceramic Restorative Cement. Eur. J. Oral Sci. 2002, 110, 175–178. [Google Scholar] [CrossRef]
  42. De Gee, A.J.; Pallav, P.; Davidson, C.L. Effect of Abrasion Medium on Wear of Stress-Bearing Composites and Amalgam in Vitro. J. Dent. Res. 1986, 65, 654–665. [Google Scholar] [CrossRef]
Figure 1. Schematic representation of the equipment customized to perform wear testing of the samples.
Figure 1. Schematic representation of the equipment customized to perform wear testing of the samples.
Processes 12 01283 g001
Figure 2. Mean mass loss (micrograms) of G1, G2, and G3 samples tested in artificial saliva and fluoridated saliva. Notice the cylinder combination and electrolyte solution interaction.
Figure 2. Mean mass loss (micrograms) of G1, G2, and G3 samples tested in artificial saliva and fluoridated saliva. Notice the cylinder combination and electrolyte solution interaction.
Processes 12 01283 g002
Figure 3. Distribution of Ra and Rz variation (micrometers) of all samples from T0 to T1 clustered by type of solution.
Figure 3. Distribution of Ra and Rz variation (micrometers) of all samples from T0 to T1 clustered by type of solution.
Processes 12 01283 g003
Figure 4. SEM images of an amalgam sample tested with Ti-6Al-4V in artificial saliva: (a) 350× and (b) 1500×.
Figure 4. SEM images of an amalgam sample tested with Ti-6Al-4V in artificial saliva: (a) 350× and (b) 1500×.
Processes 12 01283 g004
Figure 5. SEM images of an amalgam sample tested with Ti-6Al-4V in fluoridated saliva: (a) 75× and (b) 1000×.
Figure 5. SEM images of an amalgam sample tested with Ti-6Al-4V in fluoridated saliva: (a) 75× and (b) 1000×.
Processes 12 01283 g005
Figure 6. SEM images of a Ti-6Al-4V tested with amalgam in artificial saliva: (a) 750× and (b) 1000×.
Figure 6. SEM images of a Ti-6Al-4V tested with amalgam in artificial saliva: (a) 750× and (b) 1000×.
Processes 12 01283 g006
Figure 7. SEM image of a Ti-6Al-4V tested with amalgam in artificial saliva at 3500×. Notice the micro-crack identified in the sliding track.
Figure 7. SEM image of a Ti-6Al-4V tested with amalgam in artificial saliva at 3500×. Notice the micro-crack identified in the sliding track.
Processes 12 01283 g007
Figure 8. SEM images of a Ti-6Al-4V sample tested with amalgam in fluoridated saliva: (a) 500× and (b) 1500×.
Figure 8. SEM images of a Ti-6Al-4V sample tested with amalgam in fluoridated saliva: (a) 500× and (b) 1500×.
Processes 12 01283 g008
Figure 9. SEM images of amalgam samples tested simultaneously: (a) artificial saliva at 500× and (b) fluoridated saliva at 350×.
Figure 9. SEM images of amalgam samples tested simultaneously: (a) artificial saliva at 500× and (b) fluoridated saliva at 350×.
Processes 12 01283 g009
Figure 10. SEM images of Ti-6Al-4V samples tested simultaneously: (a) artificial saliva at 75× and (b) fluoridated saliva at 500×.
Figure 10. SEM images of Ti-6Al-4V samples tested simultaneously: (a) artificial saliva at 75× and (b) fluoridated saliva at 500×.
Processes 12 01283 g010
Table 1. Mean and median values of the registered initial (T0) and final (T1) mass of the Ti-6Al-4V and amalgam samples tested in artificial saliva (AS) and fluoridated artificial saliva (FS). Values in grams, except for the values of T0–T1 variation, which are in micrograms. p-value for the T0–T1 comparison (Wilcoxon test).
Table 1. Mean and median values of the registered initial (T0) and final (T1) mass of the Ti-6Al-4V and amalgam samples tested in artificial saliva (AS) and fluoridated artificial saliva (FS). Values in grams, except for the values of T0–T1 variation, which are in micrograms. p-value for the T0–T1 comparison (Wilcoxon test).
T0T1ΔT0–T1
NMean (SD), gMean (SD), gMean (SE), mgp-Value
ASG1 Ti-6Al-4V102.4 (0.2)2.4 (0.2)820 (300)0.025 *
G2 Amalgam103.6 (0.4)3.6 (0.4)3070 (300)<0.001 *
G3Ti-6Al-4V53.5 (0.5)3.5 (0.5)3040 (400)0.002 *
Amalgam53.3 (0.6)3.3 (0.6)2440 (800)0.035 *
FSG1 Ti-6Al-4V102.4 (0.3)2.4 (0.3)3560 (200)<0.001 *
G2 Amalgam103.7 (0.4)3.7 (0.4)5620 (300)<0.001 *
G3Ti-6Al-4V52.4 (0.3)2.4 (0.3)7200 (2500)0.045 *
Amalgam53.4 (0.6)3.4 (0.6)1200 (3200)0.732
* is the usual indication of what is statistically significant.
Table 2. Mean values of the registered initial (T0) and final (T1) Ra and Rz to Ti-6Al-4V and amalgam samples tested in artificial saliva and fluoridated artificial saliva. Values in micrometers.  p-value for the T0–T1 comparison within each group (Wilcoxon test); †† p-value for the group comparison at T0 and T1 (Kruskal–Wallis test).
Table 2. Mean values of the registered initial (T0) and final (T1) Ra and Rz to Ti-6Al-4V and amalgam samples tested in artificial saliva and fluoridated artificial saliva. Values in micrometers.  p-value for the T0–T1 comparison within each group (Wilcoxon test); †† p-value for the group comparison at T0 and T1 (Kruskal–Wallis test).
Ra
T0T1
NMean (SD), µmMean (SD), µmp-Value
Artificial salivaG1 (Ti-6Al-4V) 100.43 (0.07)0.5 (0.1)0.431
G2 (Amalgam) 100.49 (0.05)0.59 (0.09)0.001 *
G3Ti-6Al-4V50.47 (0.05)0.5 (0.2)0.689
Amalgam50.49 (0.05)0.53 (0.07)0.118
Fluoridated salivaG1 (Ti-6Al-4V) 100.43 (0.08)0.44 (0.09)0.659
G2 (Amalgam) 100.47 (0.05)0.5 (0.1)0.146
G3Ti-6Al-4V50.38 (0.06)0.43 (0.03)0.032 *
Amalgam50.50 (0.05)0.68 (0.07)0.010 *
p-value ††0.042 *0.024 *
Rz
T0T1
NMean (SD),µmMean (SD),µmp-Value
Artificial salivaG1 (Ti-6Al-4V) 104.6 ± 0.75 ± 10.679
G2 (Amalgam) 105.1 ± 0.46 ± 10.009 *
G3Ti-6Al-4V55.0 ± 0.46 ± 20.580
Amalgam55 ± 15.6 ± 0.60.261
Fluoridated salivaG1 (Ti-6Al-4V) 104.7 ± 0.94.8 ± 0.60.728
G2 (Amalgam) 105.0 ± 0.65.5 ± 0.80.118
G3Ti-6Al-4V54.7 ± 0.44.5 ± 0.50.345
Amalgam54.7 ± 0.47 ± 20.023 *
p-value ††0.4260.019 *
* is the usual indication of what is statistically significant.
Table 3. Volume loss associated with the wear tracks measured by profilometry, mm3.  p-value for the artificial saliva and fluoridated saliva comparison within each group (Mann–Whitney test); †† p-value for the group comparison in each of the immersion solutions (artificial saliva and fluoridated saliva (Kruskal–Wallis test and all pairwise post hoc comparisons with Bonferroni correction; similar superscript letters indicate groups that differ at the 0.05 level).
Table 3. Volume loss associated with the wear tracks measured by profilometry, mm3.  p-value for the artificial saliva and fluoridated saliva comparison within each group (Mann–Whitney test); †† p-value for the group comparison in each of the immersion solutions (artificial saliva and fluoridated saliva (Kruskal–Wallis test and all pairwise post hoc comparisons with Bonferroni correction; similar superscript letters indicate groups that differ at the 0.05 level).
GroupMaterialWear Volume in Artificial Saliva Mean ± SDWear Volume in Fluoridated Saliva Mean ± SDp-Value
G1TiAlV0.0753 ± 0.0152 a,c0.0655 ± 0.0233 a,c0.315
G2Amalgam0.0061 ± 0.0097 a,b0.0050 ± 0.0092 a,b0.190
G3TiAlV0.0740 ± 0.0116 b,d0.0730 ± 0.0097 b,d>0.05
G3Amalgam0.0100 ± 0.0081 c,d0.0020 ± 0.0011 c,d0.151
p-value ††<0.001<0.001
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content.

Share and Cite

MDPI and ACS Style

Fangaia, S.I.G.; Messias, A.; Guerra, F.A.D.R.A.; Ribeiro, A.C.F.; Valente, A.J.M.; Nicolau, P.M.G. Evaluation of the Tribocorrosion Behavior of Ti-6Al-4V Biomedical Alloy in Simulated Oral Environments. Processes 2024, 12, 1283. https://doi.org/10.3390/pr12071283

AMA Style

Fangaia SIG, Messias A, Guerra FADRA, Ribeiro ACF, Valente AJM, Nicolau PMG. Evaluation of the Tribocorrosion Behavior of Ti-6Al-4V Biomedical Alloy in Simulated Oral Environments. Processes. 2024; 12(7):1283. https://doi.org/10.3390/pr12071283

Chicago/Turabian Style

Fangaia, Sónia I. G., Ana Messias, Fernando A. D. R. A. Guerra, Ana C. F. Ribeiro, Artur J. M. Valente, and Pedro M. G. Nicolau. 2024. "Evaluation of the Tribocorrosion Behavior of Ti-6Al-4V Biomedical Alloy in Simulated Oral Environments" Processes 12, no. 7: 1283. https://doi.org/10.3390/pr12071283

Note that from the first issue of 2016, this journal uses article numbers instead of page numbers. See further details here.

Article Metrics

Back to TopTop